Majority of ultra-high molecular weight polyethylene (UHMWPE) medical devices used in total joint arthroplasty are crosslinked using gamma radiation to improve wear resistance. Alternative methods of crosslinking are urgently needed to replace gamma radiation due to rapid decline in its supply. Peroxide crosslinking is a candidate method with widespread industrial applications. Oxidative stability and biocompatibility, which are critical requirements for medical device applications, can be achieved using vitamin-E as an additive and by removing peroxide by-products through high temperature melting, respectively. We investigated compression molded UHMWPE/vitamin-E/di-cumyl peroxide blends followed by high-temperature melting in inert gas as a material candidate for tibial knee inserts. Wear resistance increased and mechanical properties remained largely unchanged. Oxidation induction time was higher than most of the other clinically available formulations. The material passed the local-end point biocompatibility tests per ISO 10993. Compounds found in exhaustive extraction were of no concern with margin-of-safety values well above the accepted level, indicating a desirable toxicological risk profile. Peroxide crosslinked, vitamin-E stabilized, and high temperature melted UHMWPE has recently been cleared for clinical use in tibial knee inserts. With all the salient characteristics needed in a material that can provide superior long-term performance in total joint patients, peroxide crosslinking can replace gamma radiation crosslinking of UHMWPE for use in all total joint replacement implant including acetabular liners.
Isolated liner exchange with highly crosslinked polyethylene (HXLPE) is an option to address polyethylene wear and osteolysis after total hip arthroplasty (THA). The liner can be fixed with either the original locking mechanism or cemented into the acetabular cup. Whether the method used for fixation of HXLPE liner has any bearing on the outcomes 10 to 15 years after surgery is still unclear. Data for all patients who had undergone isolated liner exchange surgery with HXLPE in our institution between August 2000 and January 2015 were retrieved. Patients were classified according to the locking mechanism used (original locking mechanism or cemented). Survivorship and revision rates were compared among different subgroups. A total of 86 isolated liner exchanges were performed and patients had a mean duration of follow-up of 13 years (range, 5 to 22). 50 patients had the new liner cemented and the other 36 patients used the original locking mechanism. 20 patients (23.3%) had complications, with dislocation being the most common (7). 10 of them (12.6%) required re-revision surgery. Overall estimated mean survivorship was 19.2 years. There was no significant difference in the re-revision rate between original locking mechanism (11.1%) and cementation (12.0%) for the fixation of HXLPE (p=0.899). Using the Kaplan-Meier survival analysis, the revision-free survival of HXLPE fixed with original locking mechanism and cementation were 94.1% and 93.2%, respectively, at 10 years, and were 84.7% and 81.3%, respectively, at 20 years (p=0.840). The re-revision rate and the revision-free survival of liner exchange surgery using HXLPE liner were not affected by the fixation technique used. Both cementing a HXLPE liner or using the original locking mechanism were associated with good survivorship at 13 years follow up.
Our aim in this study was to describe a continuing review of
11 total hip arthroplasties using 22.225 mm Alumina ceramic femoral
heads on a Charnley flanged femoral component, articulating against
a silane crosslinked polyethylene. Nine patients (11 THAs) were reviewed at a mean of 27.5 years
(26 to 28) post-operatively. Outcome was assessed using the d’Aubigne
and Postel, and Charnley scores and penetration was recorded on
radiographs. In addition, the oxidation of a 29-year-old shelf-aged
acetabular component was analysed.Aims
Patients and Methods
To compare complications, survivorship and results in 2 groups of Furlong-HAP Active, one with ceramic-ceramic and the other with metal-XLPE friction pair Prospective, non-randomized, comparison of 2 series of JRI uncemented prosthesis, implanted with identical protocol by 1 surgeon in 1 hospital from 2006 to 2014. Friction pair was ceramic (Biolox Forte or Delta) in 35 patients of 53.7+/−10.6 years (25–69) (21 males, 60%), and CrCo-XLPE in 65 cases of 69.0+/−8.9 years (42–81) (36 males, 55%); there were significant differences in age (p<0.00001) but not in sex (p=0.6565). Head diameter: Ce-Ce with 19 of 28mm, 9 of 32 and 7 of 36mm; Me-PE with 63 of 28mm, 1 of 32 and 1 of 36. Follow-up averaged 10.5+/−3.1 years (1–15) in ceramic and 9.8+/−3.8 years (2–15) in XLPE group. Pearson, Fisher, Kolmogorov-Smirnov, Student, Mann-Whitney, calculated with the informatic tools Microsoft Excel 2007 and Complications in ceramic joints: 2 late infections (Fisher exact test=0.6101) (1 DAIR, 1 one-stage exchange); 1 dislocation (Fisher exact test=0.2549) (closed reduction); 1 Vancouver C fracture (ORIF) (Fisher exact test=0.6548). Complications in Me-XLPE joints: 2 late infections (Fisher=0.6101) (1 DAIR, 1 two-stage exchange); 7 dislocations (Fisher=0.2549) (2 early, open reduction) (5 late: 3 closed reduction, 1 cup revision, 1 constrained cup); 4 Vancouver B fractures (Fisher=0.6548) (2 intraoperative, cerclages; 2 late, exchange). Final follow-up: Harris Hip Score averaged 93.2+/−13.7 (23–100) in ceramic and 94.3+/−8.7 (65–100) in XLPE joints (p=0.64552). Wear: 0.06+/−0.38mm (0–1.5) in ceramic and 0.16+/−0.5mm (0–2) in Me-PE THAs (p=0.30302). Osteolysis in Charnley-De Lee zones: 8 zones (6 patients) (17%) in ceramic cups, 25 zones (15 patients) (23%) in XLPE cups (p=0.980127). Survivorship without any surgery or closed reduction after 15 years: 91.0% in ceramic joints, 83.8% in Me-XLPE joints. Survivorship without component exchange after 15 years: 93.9% in ceramic joints, 93.6% in Me-XLPE joints. At least after 10 years follow-up of Furlong-HAP Active, metal-XLPE and ceramic-ceramic joints present no significant differences in complications, clinical score, wear, acetabular osteolysis, or survivorship without component exchange. On the contrary, survivorship without any surgery or closed reduction is different because of the high rate of dislocation in 28mm metal-poly joints.
Highly crosslinked polyethylene (HXLPE) has been used with great clinical success in total hip arthroplasty (THA) since its debut in the late 1990's. However, reports regarding this bearing couple in its second decade of service are still scant. The aim of this study was to 1. Determine the long term clinical and radiological results and 2. Investigate what factors affect wear rates using a metal-on-HXLPE bearing articulation. 55 THA's using a single brand of HXLPE liner, cementless cup and 28mm hip ball were performed in 44 patients. Age, sex, and Charlson Comorbidity Index (CCI) and need for revision surgery were recorded. Linear and volumetric wear was determined using the Martell method. Mean age at operation was 51.2 (29–73 +/− 12.1) years. Mean duration of follow up was 16.9 years (range 15.0–20.1 +/− 1.1 years). Osteolysis was not present in the latest follow up radiographs. Median linear and volumetric wear rate was 0.035mm/year (95% CI 0.031–0.047) and 7.12mm3/year (95% CI 6.92–17.25) respectively. Acetabular component position was not found to be related to both linear and volumetric wear. No significant difference was found in the linear and volumetric wear rates of thinner and thicker liners (8mm or below and > 8mm) (p=0.827 and p=0.843 respectively). HXLPE is associated with very low linear and volumetric wear rates which has virtually obviated osteolysis and has translated to excellent survivorship even at long term follow up. In-vivo oxidation does not appear to be of clinical concern at this point in its service cycle.
Our primary aim was to describe migration of the Exeter stem with a 32 mm head on highly crosslinked polyethylene and whether this is influenced by age. Our secondary aims were to assess functional outcome, satisfaction, activity, and bone mineral density (BMD) according to age. A prospective cohort study was conducted. Patients were recruited into three age groups: less than 65 years (n = 65), 65 to 74 years (n = 68), and 75 years and older (n = 67). There were 200 patients enrolled in the study, of whom 115 were female and 85 were male, with a mean age of 69.9 years (sd 9.5, 42 to 92). They were assessed preoperatively, and at three, 12 and, 24 months postoperatively. Stem migration was assessed using Einzel-Bild-Röntgen-Analyse (EBRA). Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Harris Hip Score (HHS), Hip Disability and Osteoarthritis Outcome Score (HOOS), EuroQol-5 domains questionnaire (EQ-5D), short form-36 questionnaire (SF-36,) and patient satisfaction were used to assess outcome. The Lower Extremity Activity Scale (LEAS), Timed Up and Go (TUG) test, and activPAL monitor (energy expelled, time lying/standing/walking and step count) were used to assess activity. The BMD was assessed in Gruen and Charnley zones.Objectives
Patients and Methods
Trabecular metal (TM) augments are designed to support an uncemented socket in revision surgery when adequate rim fit is not possible. We have used TM augments in an alternative arrangement, to contain segmental defects to facilitate impaction bone grafting (IBG) and cementation of a cemented socket. However, there is a paucity of literature supporting the use of this technique. We present one of the largest studies to date, reporting early outcomes of patients from a tertiary centre. A single-centre retrospective analytical study of prospectively collected data was performed on patients who had undergone complex acetabular reconstruction using TM augments, IBG and a cemented cup. All patients operated between 2015 and 2019 were included. We identified 105 patients with a mean age of 74yrs. The mean follow-up was 2.3 years(1–5.5yrs). Our primary outcome measure was all-cause revision of the construct. The secondary outcome measures were, Oxford hip score (OHS), radiographic evidence of cup migration/loosening and post-op complications. Eighty-four out of 105 patients belonged to Paprosky grade IIb, IIc or IIIa. Kaplan-Meier survivorship for all-cause revision was 96.36% (CI, 90.58–100.00) at 2 years with 3 failures. Two were due to early infection which required two-stage re-revision. The third was due to post-operative acetabular fracture which was then re-revised with TM augment, bone graft and large uncemented cup. Pre-op and post-op matched OHS scores were available for 60 hips(57%) with a mean improvement of 13 points. Radiographic analysis showed graft incorporation in all cases with no evidence of cup loosening. The mean vertical cup migration was 0.5mm (Range −5 to 7mm). No other complications were recorded. This study shows that reconstruction of large acetabular defects during revision THA using a combination of TM augments to contain the acetabulum along with IBG to preserve the bone stock and a cemented socket is a reliable and safe technique with low revision rates and satisfactory clinical and radiographic results. Long term studies are needed to assess the possibility of preservation and regeneration of bone stock.
The purpose of this study was to compare oxinium versus metal-on-polyethylene wear in two consecutive prospective randomized series of low friction total hip arthroplasty at a minimum 10-year follow-up. A total of 100 patients with a median age of 60.9 years were randomized to receive either oxinium (50 hips) or metal (50 hips) femoral head. The polyethylene socket was EtO sterilized in the first 50 patients, whereas it was highly cross-linked and remelted (XLPE) in the following 50 patients. The primary criterion for evaluation was linear head penetration measurement using the Martell system by an investigator blinded to the material. Also, a survivorship analysis was performed using wear related loosening revised or not as the end point. Complete data were available for analysis in 40 hips at a median follow-up of 12.9 years (11 to 14), and in 36 hips at a median follow-up of 12.3 years (10 to 13) in the EtO sterilized and XLPE series, respectively. In the EtO sterilized series, the mean steady-state wear rate was 0.245 ± 0.080 mm/year in the oxinium group versus 0.186 ± 0.062 mm/year in the metal group (p = 0.009). In the XLPE series, the mean steady-state wear rate was 0.037±0.016 mm/year in the oxinium group versus 0.036±0.015 mm/year in the metal group (p = 0.94). The survival rate at 10 years was 100% in both XLPE series, whereas it was 82.9% (IC 95%, 65–100) and 70.5% (IC95%, 50.1–90.9) in the metal-EtO and oxinium-EtO series, respectively. This RCT demonstrated that up to 14-year follow-up, wear was significantly reduced when using XLPE, irrespective of the femoral head material. Also, no osteolysis related complication was observed in the XLPE series. In the current study, oxinium femoral heads showed no advantage over metal heads and therefore their continued used should be questioned related to their cost.
Lewinnek's safe zone recommendation to minimise dislocations was a target of 5–25° for anteversion angle and 30–50° for inclination angle. Subsequently, it was demonstrated that mal-positioning of the acetabular cup can also lead to edge loading, liner fracture, and greater conventional polyethylene wear. The purpose of this study was to measure the effect of acetabular cup position on highly crosslinked polyethylene wear in total hip arthroplasty (THA) at long-term follow-up. We identified all patients that underwent primary THA with a minimum of 10 years follow-up using an institutional database in London, Ontario, Canada. Patients with a single implant design consisting of a 28 mm cobalt chromium head and highly crosslinked polyethylene liner (ram extruded, GUR 1050, 100 kGy gamma irradiated, remelted, ethylene oxide sterilised) were selected for inclusion. In total, 85 hips from 79 recruited patients were analysed. Patients underwent a supine radiostereometric analysis (RSA) exam in which the x-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup anteversion angle, inclination angle, and 3D penetration rate (including wear and creep) were measured from the stereo radiograph pairs. At a mean follow-up of 13 years (range, 10–17 years) the mean penetration rate was 0.059 mm/year (95% CI: 0.045 to 0.073 mm/year). Mean anteversion angle was 18.2° (range, −14 to 40°) and mean inclination angle was 43.6° (range, 27 to 61°). With respect to the Lewinnek safe zone, 67% hips met the target for anteversion angle, 77% met the target for inclination angle, and 51% met the target for both. There was no correlation between anteversion angle and penetration rate (r = −0.14, p = 0.72) or between inclination angle and penetration rate (r = 0.11, p = 0.35). There was also no difference (p = 0.07) in penetration rate between hips located within the Lewinnek safe zone for both anteversion angle and inclination angle (mean 0.057 mm/year, 95% CI: 0.036 to 0.079 mm/year) and those outside the safe zone (mean 0.062 mm/year, 95% CI: 0.042 to 0.083 mm/year). Acetabular cup position had no effect on the wear rate of highly crosslinked polyethylene at long-term follow-up. Although care should still be taken to correctly position the acetabular cup for stability, highly crosslinked polyethylene is a forgiving bearing material that can withstand a wide range of cup positions without negatively impacting longevity due to wear.
Young patients are at increased risk of revision after primary THA (THA). The bearing surface may be of importance for the longevity of the joint. We aimed to compare the risk of revision of primary stemmed cementless THA with MoM and CoC with metal-on-highly-crosslinked-polyethylene (MoXLP) bearings in patients between 20–54 years. From NARA, we included 2,153 MoM, 4,120 CoC and 10,329 MoXLP THA operated between 1995 and 2017. Kaplan-Meier estimator was used for calculation of THA survivorship and Cox regression to estimate the hazard ratio (HR) of revision (95% CI) due to any and specific causes. MoXLP was reference. The median follow-up was 10.3 years for MoM, 6.6 years for CoC and 4.8 years for MoXLP. 15 years postoperatively the Kaplan-Meier survival estimates were 80% (78–83%) for MoM, 92% (91–93%) for CoC and 94% (93–95%) for MoXLP. The 0–2, 2–7 and 7–15 years adjusted HRs of revision by any cause were 1.4 (0.9–2.4), 3.2 (2.1–5.1) and 3.9 (1.9–7.9) for MoM and 1.1 (0.8–1.4), 1.0 (0.7–1.3) and 2.5 (1.3–4.8) for CoC bearings. After 7–15 years follow-up, the unadjusted HR of revision due to aseptic loosening was 5.4 (1.2–24) for MoM and 4.2 (0.9–20) for CoC THA. MoM and CoC had a 7–15 year adjusted HR of revision due to ‘other’ causes of 4.8 (1.6–14) and 2.1 (0.8–5.8). MoXLP bearings were associated with better survival than MoM and CoC bearings, mainly because of lower risk of revision due to aseptic loosening and ‘other’ causes.
Our previous work presented at BHS revealed a reduced risk of revision for all reasons in THAs using lipped (asymmetric) liners. Some audience members felt that this finding may be due to unaccounted confounders and the hip surgery community remains sceptical. A fully adjusted Cox model was built after exploratory Kaplan-Meier analyses. The following surgical approaches were included in the analysis: Posterior, Hardinge/anterolateral, Other. The variables included in the final Cox model included: Gender, liner asymmetry, age, head composition, stem fixation method, head diameter, indication for implantation and surgical approach. An additional analysis of the 3 most commonly used polyethylene liners with both a flat and asymmetric version was performed. In the fully adjusted Cox model, the use of a flat liner was associated with increased risk of revision for instability (HR: 1.79, 95%CI: 1.52–2.10) and increased risk of revision for all reasons (HR 1.195, 95%CI: 1.104 – 1.293) when compared to THAs utilising flat liners. This finding was upheld in the product specific analysis. When utilising flat liners, the Hardinge/anterolateral approach was associated with a reduced risk of revision for instability compared to the posterior approach (HR 0.56, 95%CI: 0.446 to 0.706). When an asymmetric liner was used, there was no significant difference in the risk of revision for instability between the Hardinge/anterolateral approach and the posterior approach (HR 0.838, 95%CI: 0.633 – 1.110). The use of polyethylene-based bearings gives the surgeon the option to use asymmetric liners. The reduced risk of revision in THAs with asymmetric liners was seen in the analysis of the complete dataset and in the product specific analysis. Our results demonstrate that the posterior approach, when used with asymmetric XLPE liners, is not associated with a higher risk of revision for instability as historical data suggested.
Patients under the age of 50 who undergo a total hip arthroplasty (THA) are at high risk for wear-related complications due to their higher activity level. Previous studies have shown that highly crosslinked polyethylene (HXLPE) is more durable with no evidence of loosening compared to conventional polyethylene due to its improved wear characteristics. Unfortunately, there are few studies with long term follow-up of HXLPE in this patient population. The purpose of this study was to evaluate two questions related to this population of patients undergoing THA. First, what were the clinical outcomes for HXLPE in patients 50 years or younger at an average follow-up of 15 years? Second, was osteolysis observed in any of these hips? Between November 1999 and April 2005, 105 THAs were performed for 90 patients 50 years of age or younger (mean, 42 years; range, 20–50 years). The mean body mass index (BMI) was 30 kg/m2 (range, 17–51 kg/m2). The mean follow-up was 15.25 years (range, 12–19 years). Eight patients (two bilateral) were lost to follow-up, which left 82 patients with 95 hips for analysis. HXLPE was the acetabular bearing for all hips. Harris Hip Scores were collected from all patients. Radiographs were obtained on all patients to evaluate for wear and osteolysis.Introduction
Methods
Total Hip Arthroplasty (THA) is a successful operation for the management of end stage hip osteoarthritis (OA) but long term success is limited by wear of the polyethylene bearing surface. The aim of this study was to compare the rate of revision at 15 years in patients <55 who had a THA for OA, and received either cross-linked (XLPE) or conventional non cross-linked polyethylene (non-XLPE). The study population was all patients with primary THAs undertaken for OA from 1999 to 31 December 2016. Outcomes were determined for all procedures, comparing THA performed with non-XLPE and XLPE and including the effect of age, sex, and reason for revision. The principal outcome measure was time to first revision using Kaplan-Meier estimates of survivorship. There were 17,869 procedures recorded for younger patients <55 years of age undergoing THA for OA and using either non XLPE or XLPE. There was a fivefold increase in the rate of revision for procedures using non-XLPE after seven years. The 15 year cumulative percent revision of primary THA performed for OA in patients <55 with non XLPE was 17.4% (95% CI 15.5,19.5) and for XLPE was 6.6% (95%CI 5.5,7.8) HR >7 years =5.3, p<0.001. Non-XLPE and XLPE were combined with three different femoral head bearing surfaces: ceramic, metal and ceramicised metal. Within each bearing surface, XLPE had a lower rate of revision than non-XLPE. For the most common head size of 28mm XLPE had a lower rate of revision. The use of XLPE has resulted in a significant reduction in the rate of revision for younger patients undergoing THA for OA at 15 years. This evidence suggests that longevity of THA is likely to be improved and may enable younger patients to undergo surgery, confident of a reduced need for revision in the long term.
Osteolysis secondary to ultra-high molecular weight polyethylene (UHMWPE) wear is a leading cause of late-term implant failure via aseptic loosening in patients treated with total hip arthroplasty (THA). Radiation crosslinking of UHMWPE has been shown to decrease wear. However, the resulting polymer (crosslinked-PE) has a high free radical content. Two different methods that have been used to reduce the remaining free radicals are mechanical annealing and chemical stabilization using Vitamin E, a free radical scavenger. The primary purpose of the current study was to evaluate and compare the wear properties of vitamin E-doped crosslinked-PE (VEPE) and one formulation of mechanically annealed crosslinked-PE using radiostereometric analysis (RSA) in patients five years after primary THA. We also sought to understand the association between polyethylene wear and patient-reported outcome measures (PROMs). Three-hundred and five patients from six international centers were enrolled. Seventy-six percent were treated with highly-crosslinked (95 kGy) VEPE liners, and the rest received moderately-crosslinked (50 kGy) (ModXL), mechanically annealed liners. Data was collected prospectively at one-, two-, and five-year intervals. At the 5-year follow-up, proximal femoral head penetration into the VEPE liners (median = 0.05mm (range, −0.03–1.20)) was significantly lower than the penetration into the ModXL liners (median = 0.15mm (range, −0.22–1.04)) (p<0.001). In the VEPE cohort the median proximal penetration did not increase from one- to five-year follow-up (p=0.209). In contrast, there was a significant increase in femoral head penetration for the ModXL group (p<0.001) during that same time. Multivariable regression showed that the only variable predictive of increased wear was ModXL liner type (B=0.12, p<0.001). There were no differences in PROMs between the liner groups, and there was no correlation between polyethylene wear and PROMs for the cohort as a whole. The current study is the largest analysis of polyethylene wear at five-year follow-up using the RSA technique. We observed similar bedding in through the two-year interval between the two liner types, however, there was significantly more wear in the ModXL cohort at five-years. Currently, the wear rates for both liner groups are below the osteolysis threshold and have not led to any implant failures via aseptic loosening. Continued follow-up will provide a better understanding of the association between wear rate and clinical outcomes.
Previous reports on the outcomes of isolated head and liner exchange in revision total hip arthroplasty have found high rates of instability following these surgeries. Most reports have studied constructs using ≤28mm femoral heads. The purpose of this study was to determine if modern techniques with the use of larger head sizes can improve the rate of instability after head and liner exchange. We identified 138 hips in 132 patients who underwent isolated head and liner exchange for polyethylene wear/osteolysis (57%), acute infection (27%), metallosis (13%), or other (2%). All patients underwent revision with either 32mm (23%), 36mm (62%), or 40mm (15%) diameter heads.
Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).Aims
Methods
Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.Aims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation.Aims
Methods