Despite many claims of good wear properties following
total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral
component, there are conflicting clinical results. We hypothesised
that there would be no difference in either the mid-term clinical
and radiological outcomes or the characteristics of the polyethylene
wear particles (weight, size and shape) in patients using an OxZr
or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent
bilateral TKR, receiving an OxZr femoral component in one knee and
a
Aims. The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components. Methods. Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five
Introduction. Total knee arthroplasty (TKA) femoral components are known to wear and roughen with clinical use, and reaction to metal is a well-documented complication of TKA. Ceramic materials are resistant to wear and corrosion, but the surface wear of ceramic femoral components in TKA has not been reported. This study measured the changes in roughness and mass lost from cobalt-chromium (CoCr) femoral components tested in a knee simulator, and compared them to those observed in ceramic components. Methods. Six cast
Objectives. Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation. Methods. A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry. Results. There was no significant difference (p > 0.05) between the wear rate of implants challenged with CS-BVF (3.3 mm. 3. /MC (95% confidence interval (CI) 1.8 to 4.8)) and the wear rate of those not challenged (2.8 mm. 3. /MC (95% CI 1.3 to 4.3)). However, scratching the cobalt-chrome (CoCr) significantly (p < 0.001) increased the wear rate (20.6 mm. 3. /MC (95% CI 15.5 to 25.7)). The mean surface roughness of implants challenged with CS-BVF was equivalent to negative controls both after damage simulation (p = 0.98) and at the conclusion of the study (p = 0.28). Conclusion. When used close to articulating surfaces, a low-hardness, high-purity CS-BVF had no influence on wear. When trapped between the articulating surfaces of a total knee arthroplasty, CS-BVF did not scratch the surface of
Purpose. The success of total knee replacement (TKR) surgery can be attributed to improvements in TKR design, instrumentation, and surgical technique. Over a decade ago oxidized zirconium (OxZr) femoral components were introduced as an alternative bearing surface to cobalt-chromium (CoCr), based on strong in-vitro evidence, to improve the longevity of TKR implants. Early reports have demonstrated the clinical success of this material however no long-term comparative studies have demonstrated the superiority of OxZr implants compared to a more traditional CoCr implant. This study aims to compare long-term survivorship and outcomes in OxZr and
Introduction. Previous studies of CoCr alloy femoral components for total knee arthroplasty (TKA) have identified 3. rd. body abrasive wear, and apparent inflammatory cell induced corrosion (ICIC) [1] as potential damage mechanisms. The association between observed surface damage on the femoral condyle and metal ion release into the surrounding tissues is currently unclear. The purpose of this study was to investigate the damage on the bearing surface in TKA femoral components recovered at autopsy and compare the damage to the metal ion concentrations in the synovial fluid. Methods. 12 autopsy TKA
INTRODUCTION. The development of new bearing surfaces for total joint replacement is constantly evolving. Oxidized zirconium (Oxinium) has been introduced for use in both total hip arthroplasty (THA) and total knee arthroplasty (TKA). The aetiology of wear is multifactorial and includes adhesive, abrasive, third-body and fatigue wear mechanisms. Oxinium femoral components have demonstrated clear improvements in wear characteristics in-vitro. The purpose of this prospective study was to evaluate the mid-term (minimum 5 year) clinical and radiographic results and survivorship of the Genesis II™ knee implant system using an Oxinium femoral component. METHODS. Between January 2001 and December 2008, 382 Genesis II Oxinium (Smith & Nephew) primary total knee arthroplasties (TKA) (313 patients) were implanted at our institution. A comparison with a cohort of 317 patients (382 knees) who received a Genesis II knee implant using a ‘conventional’ cobalt-chrome (Co-Cr) femoral component was performed during the same time period. Prospective data was collected on all patients including demographics (age, BMI, diagnosis) as well as pre and postoperative clinical outcome scores (SF-12, WOMAC, and knee society clinical rating scores (KSCRS). Radiological analysis for evidence of osteolysis and loosening was performed in all patients. Comparisons were performed to determine differences between the Oxinium and Cobalt Chrome cohorts. Kaplan-Meier survival analysis was performed to show cumulative survival over time. Failure was defined as femoral component revision due to any cause. RESULTS. The mean age at time of surgery was 54.0 and 60.6 years respectively for the Oxinium and Co-Cr groups. The mean time from surgery was 8.13 +/− 2.03 years and 8.59 +/− 2.01 years for the Oxinium and Co-Cr groups respectively. Both groups demonstrated a statistically significant improvement in clinical outcome measures between pre, and post operative SF-12, WOMAC, and KSCRS scoring systems with no difference between two groups. There were a total of 15 revisions (infection: 8, stiffness: 2, aseptic loosening of patella component: 2, Aseptic loosening of the tibial component: 1, instability: 1, and one malaligned Femoral Component) in the Oxinium group. In the CoCr group there were a total of 9 revisions (Infection:5, aseptic loosening:1, instability: 1, stiffness: 1 and one patient revised for unexplained pain). Radiographically, no loosening was noted in the Oxinium group and all implants appeared well fixed. There was one case of loosening seen in the cobalt chrome group. The overall survival was 99.7 for both Oxinium and
Introduction. PEEK-OPTIMA™ has been considered as an alternative to cobalt chrome in the femoral component of total knee replacements. Whole joint wear simulation studies of both the tibiofemoral and patellofemoral joints carried out to date have shown an equivalent wear rate of UHMWPE tibial and patella components against PEEK and cobalt chrome (CoCr) femoral components. In this study, the influence of third body wear on UHMWPE-on-PEEK was investigated, tests on UHMWPE-on-CoCr were carried out in parallel to compare PEEK to a conventional femoral component material. Methods. Wear simulation was carried out in simple geometry using a 6-station multi-directional pin-on-plate simulator. 5 scratches were created on each PEEK and CoCr plate perpendicular to the direction of the wear test using a diamond stylus to produce scratches with a geometry similar to that observed in retrieved
INTRODUCTION. Tibiofemoral contact at the base of the articular surface spine in posterior-stabilized total knee arthroplasty (TKA) implants can lead to spine fracture [1]. Revision TKA implants also have an articular surface spine to provide sufficient constraint when soft tissues are compromised. While some revision TKA designs have metal reinforcement in the articular surface spine, others rely solely on a polyethylene spine. This study used finite element analysis (FEA) to study the effect of metal reinforcement on stresses in the spine when subjected to posteriorly directed loading. METHODS. Two clinically successful Zimmer Biomet revision TKA designs were selected; NexGen LCCK with metal reinforcement and all-poly Vanguard SSK. The largest sizes were selected. FEA models consisted of the polyethylene articular surface and a
Introduction. Infection remains as one of the major challenges of total joint surgery. One-stage irrigation, debridement and reimplantation, or two-stage revision surgery with a temporary implantation of antibiotic eluting bone cement spacer followed by reimplantation are two methods often used to treat infected patients with mixed outcomes. Like bone cement, ultra-high molecular weight polyethylene (UHMWPE) can also be used as a carrier for antibiotics. Recently, we demonstrated that vancomycin and rifampin can be delivered from UHMWPE implants at therapeutic levels to eradicate Staphylococcus aureus biofilm in a lupine animal model. There are regulatory challenges in translating these types of combination devices to clinical use. Last year, at this meeting, we presented the preliminary pre-clinical testing for a temporary UHMWPE spacer containing gentamicin sulfate as a first step towards clinical use. Since then, we carried out a survey among the Knee Society membership about their preference for spacer use in two-stage revision surgery and found that 43% prefer to use a
Introduction. Previous studies of long-term CoCr alloy femoral components for TKA have identified 3rd body abrasive wear and inflammatory cell induced corrosion (ICIC). The extent of femoral condyle surface damage in contemporary
As allergic reactions to implant wear are gaining more attention [4], the incorporation of ceramic materials to device design appears to be a promising development. In particular, ceramic femoral components of total knee replacements have been designed to produce less wear under standard [1] and adverse [5] implant conditions. Whereas the wear reduction effect of ceramics is generally accepted for hip implants, the corresponding effect for knee implants is not proven. Ezzet et al. reported a wear reduction of 42% for standard wear conditions [2] and of 55% for adverse wear conditions [3] when compared to a geometrically identical
Demand for TKR surgery is rising, including a more diverse patient demographic with increasing expectations [1]. Therefore, greater efforts are being devoted to laboratory testing. As a result, laboratory testing may set a clinical performance presumption for surgeons and patients. For example, oxidized ZrNB (Oxinium) femoral components have been projected to show 85% less wear than
Suggestions for improved wear performance of total knee replacements have included replacement of standard
The aim of this study was to compare the outcome of cemented TKR using either oxidized zirconium (oxinium) or cobalt chrome (CoCr) femoral components in patients undergoing simultaneous bilateral TKR. Patients involved in the study received one of each prosthesis, thereby acting as their own control. The hypothesis was that there would be no difference in the clinical and radiographic outcome between the two prosthetic materials. Forty consecutive patients who were undergoing bilateral Genesis ll TKR consented to participate in the study. Patients were assessed preoperatively, at five days, six weeks and one, two and five years, postoperatively. The outcome measures included the KOOS, Knee Society Score, BOA Patient Satisfaction Scale, and radiographs at six weeks and one, two and five years. In two patients polyethlylene exchange was performed at 56 months from surgery during patellofemoral resurfacing. The four retrieved polyethylene liners were studied for wear with the aid of a stereo zoom microscope and an environmental scanning electron microscope (ESEM). Both the patients and the all examiners were blinded as to the prosthesis type throughout the study. Forty patients (80 knees) were included in the study. At five years, three patients were deceased and two had developed senile dementia. No patients were lost to follow up. At five years from surgery the CoCr knee was preferred by 41% of patients compared to 13% who preferred the Oxinium knee (p=0.009). There was no significant difference in range of motion between the two prosthesis at five days, six weeks or one, two and five years. There were also no significant differences between the two prostheses in any of the other variables assessed. The four retrieved polyethylene inserts showed similar patterns of wear in terms of both wear types and patterns under examination with both the stereo zoom and scanning electron microscope with no clear differences between CoCr and Oxinium bearing against the polyethlylene. There was no difference in the grade or incidence of radiographic lucencies between the two prosthesis at five years. At five years after surgery the only significant difference between the Genesis II Oxinium prosthesis and the CoCr prosthesis was a subjective preference for the CoCr prosthesis by a higher proportion of patients. There were no unexpected complications associated with the use the Oxinium femoral implants. In the four retrieved polyethylene liners, no significant differences were identified between the two prosthesis materials in terms of detectable wear type and patterns. Continued follow up of this cohort is planned to establish whether Oxinium femoral implants have an improved survivorship compared to
We propose a state-of-the-art temporary spacer, consisting of a cobalt-chrome (CoCr) femoral component and a gentamicin-eluting ultra-high molecular weight polyethylene (UHMWPE) tibial insert, which can provide therapeutic delivery of gentamicin, while retaining excellent mechanical properties. The proposed implant is designed to replace conventional spacers made from bone cement. Gentamicin-loaded UHMWPE was prepared using phase-separated compression moulding, and its drug elution kinetics, antibacterial, mechanical, and wear properties were compared with those of conventional gentamicin-loaded bone cement.Aims
Methods
Introduction. Malrotation of the femoral component is a cause of patellofemoral maltracking after TKA. Its precise effect on the patellofemoral (PF) mechanics has not been well quantified. The aim of this study was to investigate the effect of malrotation of the femoral component on PF initial contact area, initial contact pressure and wear after 4 million full gait cycles in TKA using a knee simulator. Moreover, the influence of the counterface material (CoCr or OxZr) on PF wear was also investigated. Materials & Methods. Femoral components (FCs) were cemented onto specially designed fixtures, allowing positioning of the FC in different angles of axial rotation. Patellar buttons and FCs were then mounted in a Prosim knee simulator. Patellofemoral contact mechanics. Seven axial rotation configurations were tested: neutral (FC parallel to the epicondylar axis), 2.5° endo- and exorotation, 5° endo- and exorotation and 7.5° endo- and exorotation. Patellar contact location, contact area and contact pressure were measured dynamically during 20 gait cycles with a Tekscan sensor covering the patella collecting data at a rate of 100 frames per second. Patellofemoral wear. For three alignments (neutral, 5° endo- and exorotation), a PF wear test of 4 million cycles in bovine serum (diluted to 40%) was done with three CoCr and three OxZr components on conventional ultra-high molecular weight polyethylene (UHMWPE, density: 0.93mg/mm. 3. ). Every 0.5 million cycles the test lubricant was replaced, the patellar samples were cleaned and dried and polyethylene wear was measured gravimetrically. A linear regression model was used to calculate the wear rate of each patellar sample. Aggregate wear rates were determined for each test condition by pooling the measurements of all three patellar samples. Results. For all six endorotation and exorotation configurations, the contact area was significantly lower and the contact pressure significantly higher than the neutral position (p < 0.001, Figs 1 and 2). In the patellofemoral wear test, the highest average wear rate was found in the group of endorotated
INTRODUCTION. Many studies have looked at the effects of titanium tibial baseplates compared to cobalt chrome baseplates on backside wear. However, the surface finish of the materials is usually different (polished/unpolished) [1,2]. Backside wear may be a function not only of tray material but also of the locking mechanism. The purpose of this study was to evaluate the wear performance of conventional polyethylene inserts when mated with titanium tibial trays or cobalt chrome tibial trays that both have non-polished topside surfaces. MATERIALS AND METHODS. Three titanium (Ti) trays were used along with three cobalt chrome (CoCr) trays. The Ti trays underwent Type II anodization prior to testing. All trays were Triathlon® design (Stryker Orthopaedics, Mahwah, NJ). Tibial inserts were manufactured from GUR 1020 conventional polyethylene then vacuum/flush packaged and sterilized in nitrogen (30 kGy). Appropriate sized
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
We sought to establish whether an oxidised zirconium (OxZr) femoral
component causes less loss of polyethylene volume than a cobalt
alloy (CoCr) femoral component in total knee arthroplasty. A total of 20 retrieved tibial inserts that had articulated with
OxZr components were matched with 20 inserts from CoCr articulations
for patient age, body mass index, length of implantation, and revision
diagnosis. Changes in dimensions of the articular surfaces were compared
with those of pristine inserts using laser scanning. The differences
in volume between the retrieved and pristine surfaces of the two
groups were calculated and compared.Aims
Materials and Methods