Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 35 - 35
1 Mar 2021
Ng G Bankes M Daou HE Beaulé P Cobb J Jeffers J
Full Access

Abstract. OBJECTIVES. Although surgical periacetabular osteotomy (PAO) for hip dysplasia aims to optimise acetabular coverage and restore hip function, it is unclear how surgery affects capsular mechanics and joint stability. The purpose was to examine how the reoriented acetabular coverage affects capsular mechanics and joint stability in dysplastic hips. METHODS. Twelve cadaveric dysplastic hips (n = 12) were denuded to the capsule and mounted onto a robotic tester. The robot positioned each hip in multiple flexion angles (Extension, Neutral 0°, Flexion 30°, Flexion 60°, Flexion 90°) and performed internal-external rotations and abduction-adduction to 5 Nm in each rotational or planar direction. Each hip underwent a PAO, preserving the capsule, and was retested postoperatively in the robot. Paired sample t-tests compared the range of motion before and after PAO surgery (CI = 95%). RESULTS. Pre-operatively, the dysplastic hips demonstrated large ranges of internal-external rotations and abduction-adduction motions throughout all flexion positions. Post-operatively, the PAO slackenend the anterosuperior capsule and tightened the inferior capsule. This increased external rotation in Flexion 60° and Flexion 90° (∆. ER. = +16 and +23%) but provided lateral coverage to decrease internal rotation at Flexion 90° (∆. IR. = –15%). The PAO also reduced abduction throughout, but increased adduction in Neutral 0°, Flexion 30°, and Flexion 60° (∆. ADD. = +34, +30%, +29% respectively). CONCLUSIONS. The PAO provided crucial osseous structural coverage to the femoral head, decreasing hypermobility and adverse loading at extreme hip flexion-extension. However, it also slackened the anterosuperior capsule and increased adduction and external rotation, which may lead to ischiofemoral impingement and adductor irritations. Capsular instability may be secondary to acetabular undercoverage, thus capsular alteration may be warranted for larger corrections or rotational osteotomies. To preserve native hip and delay joint degeneration, it is crucial to preserve capsule and elucidate amount of reorientation needed without causing iatrogenic instability. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 1012 - 1018
1 Jul 2005
Beck M Kalhor M Leunig M Ganz R

Recently, femoroacetabular impingement has been recognised as a cause of early osteoarthritis. There are two mechanisms of impingement: 1) cam impingement caused by a non-spherical head and 2) pincer impingement caused by excessive acetabular cover. We hypothesised that both mechanisms result in different patterns of articular damage. Of 302 analysed hips only 26 had an isolated cam and 16 an isolated pincer impingement. Cam impingement caused damage to the anterosuperior acetabular cartilage with separation between the labrum and cartilage. During flexion, the cartilage was sheared off the bone by the non-spherical femoral head while the labrum remained untouched. In pincer impingement, the cartilage damage was located circumferentially and included only a narrow strip. During movement the labrum is crushed between the acetabular rim and the femoral neck causing degeneration and ossification. Both cam and pincer impingement lead to osteoarthritis of the hip. Labral damage indicates ongoing impingement and rarely occurs alone


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_VIII | Pages 50 - 50
1 Mar 2012
Yamamoto T Iwasaki K Motomura G Mawatari T Nakashima Y Iwamoto Y
Full Access

Introduction. Subchondral insufficiency fracture of the femoral head (SIF) is a recently proposed concept. It is generally seen in elderly people, however, some young adults have also been reported to suffer from this fracture. The purpose of this study was to investigate the clinical results of a transtrochanteric anterior rotational osteotomy (ARO) performed in young adults with SIF. Methods. This study focused on young SIF cases (age range from 10 to 29 years). Five patients were diagnosed to have SIF at our institution and conservative treatment was initially performed. The symptoms resolved in 1 case while the other 4 cases showed progression of a collapse. In these 4 cases, ARO was performed, since the fractured area was located in the anterosuperior portion of the femoral head. The average age of the patients was 22 years (range, 16 to 29 years) at the time of surgery, consisting of 2 men and 2 women. Results. All cases were followed-up and the average follow-up period was 3.4 years (range, 1.8 to 6.8 years). The average preoperative Harris hip score was 67 points, which improved to an average of 93 points at the latest follow-up. Radiographically, the fractured lesion healed and no progression of a collapse was observed in any of the 4 cases. Conclusion. In the majority of SIF cases, only the anterosuperior portion of the femoral head is involved by the fracture. When considering the optimal surgical treatments for SIF, ARO may be one of the useful options, especially for the treatment of young patients


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 971 - 976
1 Jul 2007
Kampa RJ Prasthofer A Lawrence-Watt DJ Pattison RM

In order to determine the potential for an internervous safe zone, 20 hips from human cadavers were dissected to map out the precise pattern of innervation of the hip capsule. The results were illustrated in the form of a clock face. The reference point for measurement was the inferior acetabular notch, representing six o’clock. Capsular branches from between five and seven nerves contributed to each hip joint, and were found to innervate the capsule in a relatively constant pattern. An internervous safe zone was identified anterosuperiorly in an arc of 45° between the positions of one o’clock and half past two. Our study shows that there is an internervous zone that could be safely used in a capsule-retaining anterior, anterolateral or lateral approach to the hip, or during portal placement in hip arthroscopy


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 98 - 98
1 Jul 2014
Palmer A Fernquest S Hamish L Pollard T McNally E Wilson D Wilson D Madler B Carr A Glyn-Jones S
Full Access

Summary. The dGEMRIC index correlates more strongly with the pattern of radiographic joint space narrowing in hip osteoarthritis at five year follow-up than morphological measurements of the proximal femur. It therefore offers potential to refine predictive models of hip osteoarthritis progression. Introduction. Longitudinal general population studies have shown that femoroacetabular impingement increases the risk of developing hip osteoarthritis, however, morphological parameters have a low positive predictive value. Arthroscopic debridement of impingement lesions has been proposed as a potential strategy for the prevention of osteoarthritis, however, the development of such strategies requires the identification of individuals at high risk of disease progression. We investigated whether delayed Gadolinium-Enhanced MRI of Cartilage (dGEMRIC) predicts disease progression. This imaging modality is an indirect measure of cartilage glycosaminoglycan content. Patients and Methods. 34 asymptomatic individuals from a longitudinal cohort study (sibkids) were assessed at baseline with the collection of Patient Reported Outcome Measures (PROMs), anteroposterior and cross-table lateral radiographs, 3D morphological MRI, and dGEMRIC at 3T of their index hip. A dGEMRIC index was calculated as a ratio of the anterosuperior acetabular cartilage T1 relaxation time and the total femoral and acetabular cartilage T1 relaxation time. 29 individuals were followed up at 5 years for repeat assessment (average age 51 years and range 36 to 67). Radiological measurements were made by a single observer using in house Hipmorf software. Radiographic disease progression was assessed using minimum joint space width (JSW), lateral sourcil JSW, and medial sourcil JSW. These were measured on baseline and five year follow-up anteroposterior radiographs with an intra-observer ICC of 0.916. Alpha angle measurements were made by the same observer on radiographs and MRI radial slices with an intra-observer ICC of 0.926. Results. Mean minimum JSW for the cohort fell by 0.16mm over five years (p=0.024). Baseline dGEMRIC index did not correlate with change in minimum JSW (r=0.031 p=0.873). There was a moderate correlation between baseline dGEMRIC and the direction of JSW loss (change in JSW at the lateral sourcil minus change in JSW at the medial sourcil) (r=0.561 p=0.002). There was a weak correlation between the change in Non-Arthritic Hip Score and baseline dGEMRIC (r=0.256 P=0.180). Maximum alpha angle measured on baseline MRI radial slices did not correlate with change in minimum JSW and weakly correlated with the direction of JSW narrowing (r=0.273 p=0.160). Conclusion. A low dGEMRIC index indicates reduced glycosaminoglycan concentration in the anterosuperior acetabular cartilage compared with the total femoral and acetabular cartilage. This correlates with lateral JSW narrowing relative to medial JSW narrowing as osteoarthritis progresses. The dGEMRIC index correlates better with osteoarthritis progression than alpha angle measurements and offers the potential to refine a predictive model for osteoarthritis progression to aid patient selection for clinical trials


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 181 - 181
1 Jul 2014
Speirs A Frei H Lamontagne M Beaule P
Full Access

Summary. The cartilage layer from cam-type femoroacetabular impingement deformities had lower stiffness and increased permeability compared to normal cartilage. This is consistent with osteoarthritis and supports the hypothesis of abnormal contact stresses. Introduction. Femoroacetabular impingement (FAI) has recently been associated with osteoarthritic (OA) degeneration of the hip and may be responsible for up to 90% of adult idiopathic OA cases. FAI results from deformities in the hip joint which may lead to abnormal contact stresses and degeneration. The more common cam-type deformity consists of a convex anterior femoral head-neck junction which impinges the anterosuperior acetabular rim during flexion and internal rotation of the hip. Increased subchondral bone density has been reported in this region which may be a bone remodelling response to increased contact stress. The abnormal contact is expected to cause degeneration of the cartilage layer. The goal of this study was to assess the mechanical properties of cartilage retrieved from the cam deformity and to compare this with normal articular cartilage from the femoral head. It is hypothesised that the cartilage will have a lower elastic modulus and higher permeability than normal cartilage. Patients & Methods. Osteochondral biopsies were retrieved from nine patients undergoing surgical correction of a symptomatic cam deformity as well as 10 fresh cadaveric specimens (10 hips, 6 donors). An indentation stress relaxation test was performed on each specimen to 10% of the estimated cartilage thickness. A needle penetration test was performed to accurately measure the thickness. The equilibrium modulus was calculated per Hayes et al. A specimen-specific axisymmetric finite element model was used in a non-linear optimization to obtain the fibril-reinforced poroelastic properties of the cartilage that best fit the experimental data. The material properties were non-fibrillar modulus (E. s. ), Poisson's ratio (ν. s. ) and permeability (k) and strain-independent and –dependent moduli (E. 0. , E. ε. )[4]. Results. The equilibrium modulus was 0.14 MPa and 0.63 from surgical and cadaver specimens, respectively (p=0.002). Compared to cadaver specimens, E. s. in surgical specimens was 73% lower (p=0.01), ν. s. was 43% lower (p=0.01) and k was an order of magnitude higher (p=0.02). Fibril moduli were not significantly different (p>0.35). Discussion/Conclusions. This study showed decreased elastic modulus and increased permeability in cartilage from cam deformities compared to cadaver controls. These differences are consistent with changes expected in osteoarthritic cartilage degeneration. Fibril moduli were 14% to 57% lower in surgical specimens consistent with fibrillation, however results were not significant due to high variability. Altered cellular activity and proteoglycan depletion has been reported in cartilage of cam deformities, which are similar to changes expected in osteoarthritis. The altered mechanical and biochemical properties of this cartilage therefore support the hypothesis that osteoarthritis is secondary to cam FAI deformities and is a result of abnormal contact stresses between the deformity and acetabular rim


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 2 | Pages 300 - 304
1 Mar 2002
Nötzli HP Siebenrock KA Hempfing A Ramseier LE Ganz R

We used laser Doppler flowmetry (LDF) with a high energy (20 mW) laser to measure perfusion of the femoral head intraoperatively in 32 hips. The surgical procedure was joint debridement requiring dislocation or subluxation of the hip. The laser probe was placed within the anterosuperior quadrant of the femoral head. Blood flow was monitored in specific positions of the hip before and after dislocation or subluxation. With the femoral head reduced, external rotation, both in extension and flexion, caused a reduction of blood flow. During subluxation or dislocation, it was impaired when the posterosuperior femoral neck was allowed to rest on the posterior acetabular rim. A pulsatile signal returned when the hip was reduced, or was taken out of extreme positions when dislocated. After the final reduction, the signal amplitudes were first slightly lower (12%) compared with the initial value but tended to be restored to the initial levels within 30 minutes. Most of the changes in the signal can be explained by compromise of the extraosseous branches of the medial femoral circumflex artery and are reversible. Our study shows that LDF provides proof for the clinical observation that perfusion of the femoral head is maintained after dislocation if specific surgical precautions are followed


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 258 - 263
1 Feb 2008
Parratte S Kilian P Pauly V Champsaur P Argenson JA

We have evaluated in vitro the accuracy of percutaneous and ultrasound registration as measured in terms of errors in rotation and version relative to the bony anterior pelvic plane in computer-assisted total hip replacement, and analysed the intra- and inter-observer reliability of manual or ultrasound registration.

Four clinicians were asked to perform registration of the landmarks of the anterior pelvic plane on two cadavers. Registration was performed under four different conditions of acquisition. Errors in rotation were not significant. Version errors were significant with percutaneous methods (16.2°; p < 0.001 and 19.25° with surgical draping; p < 0.001), but not with the ultrasound acquisition (6.2°, p = 0.13). Intra-observer repeatability was achieved for all the methods. Inter-observer analysis showed acceptable agreement in the sagittal but not in the frontal plane.

Ultrasound acquisition of the anterior pelvic plane was more reliable in vitro than the cutaneous digitisation currently used.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1317 - 1324
1 Sep 2010
Solomon LB Lee YC Callary SA Beck M Howie DW

We dissected 20 cadaver hips in order to investigate the anatomy and excursion of the trochanteric muscles in relation to the posterior approach for total hip replacement. String models of each muscle were created and their excursion measured while the femur was moved between its anatomical position and the dislocated position. The position of the hip was determined by computer navigation.

In contrast to previous studies which showed a separate insertion of piriformis and obturator internus, our findings indicated that piriformis inserted onto the superior and anterior margins of the greater trochanter through a conjoint tendon with obturator internus, and had connections to gluteus medius posteriorly. Division of these connections allowed lateral mobilisation of gluteus medius with minimal retraction. Analysis of the excursion of these muscles revealed that positioning the thigh for preparation of the femur through this approach elongated piriformis to a maximum of 182%, obturator internus to 185% and obturator externus to 220% of their resting lengths, which are above the thresholds for rupture of these muscles.

Our findings suggested that gluteus medius may be protected from overstretching by release of its connection with the conjoint tendon. In addition, failure to detach piriformis or the obturators during a posterior approach for total hip replacement could potentially produce damage to these muscles because of over-stretching, obturator externus being the most vulnerable.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1105 - 1109
1 Aug 2006
Kandemir U Allaire RB Jolly JT Debski RE McMahon PJ

Our aim was to determine the most repeatable three-dimensional measurement of glenoid orientation and to compare it between shoulders with intact and torn rotator cuffs. Our null hypothesis was that glenoid orientation in the scapulae of shoulders with a full-thickness tear of the rotator cuff was the same as that in shoulders with an intact rotator cuff.

We studied 24 shoulders in cadavers, 12 with an intact rotator cuff and 12 with a full-thickness tear. Two different observers used a three-dimensional digitising system to measure glenoid orientation in the scapular plane (ie glenoid inclination) using six different techniques. Glenoid version was also measured. The overall precision of the measurements revealed an error of less than 0.6°.

Intraobserver reliability (correlation coefficients of 0.990 and 0.984 for each observer) and interobserver reliability (correlation coefficient of 0.985) were highest for measurement of glenoid inclination based on the angle obtained from a line connecting the superior and inferior points of the glenoid and that connecting the most superior point of the glenoid and the most superior point on the body of the scapula. There were no differences in glenoid inclination (p = 0.34) or glenoid version (p = 0.12) in scapulae from shoulders with an intact rotator cuff and those with a full-thickness tear. Abnormal glenoid orientation was not present in shoulders with a torn rotator cuff.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 549 - 556
1 Apr 2007
Udofia I Liu F Jin Z Roberts P Grigoris P

Finite element analysis was used to examine the initial stability after hip resurfacing and the effect of the procedure on the contact mechanics at the articulating surfaces. Models were created with the components positioned anatomically and loaded physiologically through major muscle forces. Total micromovement of less than 10 μm was predicted for the press-fit acetabular components models, much below the 50 μm limit required to encourage osseointegration. Relatively high compressive acetabular and contact stresses were observed in these models. The press-fit procedure showed a moderate influence on the contact mechanics at the bearing surfaces, but produced marked deformation of the acetabular components. No edge contact was predicted for the acetabular components studied.

It is concluded that the frictional compressive stresses generated by the 1 mm to 2 mm interference-fit acetabular components, together with the minimal micromovement, would provide adequate stability for the implant, at least in the immediate post-operative situation.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1110 - 1115
1 Aug 2006
Ong KL Kurtz SM Manley MT Rushton N Mohammed NA Field RE

The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented versus uncemented fixation showed no advantage of either with regard to bone loading. When the Birmingham hip resurfacing femoral component was fixed to bone, proximal femoral stresses and strains were non-physiological. Bone resorption was predicted in the inferomedial and superolateral bone within the Birmingham hip resurfacing shell. Resorption was limited to the superolateral region when the stem was not fixed. The increased bone strain observed adjacent to the distal stem should stimulate an increase in bone density at that location. The remodelling of bone seen during revision of failed Birmingham hip resurfacing implants appears to be consistent with the predictions of our finite element analysis.