Advertisement for orthosearch.org.uk
Results 1 - 19 of 19
Results per page:

Aims

For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis.

Methods

We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.


Bone & Joint 360
Vol. 12, Issue 6 | Pages 24 - 27
1 Dec 2023

The December 2023 Foot & Ankle Roundup. 360. looks at: Subchondral bone cysts remodel after correction of varus deformity in ankle arthritis; 3D-printed modular endoprosthesis reconstruction following total calcanectomy; Percutaneous partial bone excision in the management of diabetic toe osteomyelitis; Hemiepiphysiodesis is a viable surgical option for Juvenile hallux valgus; Ankle arthroplasty vs arthrodesis: which comes out on top?; Patient-related risk factors for poorer outcome following total ankle arthroplasty; The Outcomes in Ankle Replacement Study


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1174 - 1179
1 Oct 2022
Jamshidi K Bagherifard A Mirzaei A

Aims

Osteoarticular reconstruction of the distal femur in childhood has the advantage of preserving the tibial physis. However, due to the small size of the distal femur, matching the host bone with an osteoarticular allograft is challenging. In this study, we compared the outcomes and complications of a resurfaced allograft-prosthesis composite (rAPC) with those of an osteoarticular allograft to reconstruct the distal femur in children.

Methods

A retrospective analysis of 33 skeletally immature children with a malignant tumour of the distal femur, who underwent resection and reconstruction with a rAPC (n = 15) or osteoarticular allograft (n = 18), was conducted. The median age of the patients was ten years (interquartile range (IQR) 9 to 11) in the osteoarticular allograft group and nine years (IQR 8 to 10) in the rAPC group (p = 0.781). The median follow-up of the patients was seven years (IQR 4 to 8) in the osteoarticular allograft group and six years (IQR 3 to 7) in the rAPC group (p = 0.483). Limb function was evaluated using the Musculoskeletal Tumor Society (MSTS) score.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 290 - 296
1 Feb 2022
Gosheger G Ahrens H Dreher P Schneider KN Deventer N Budny T Heitkötter B Schulze M Theil C

Aims

Iliosacral sarcoma resections have been shown to have high rates of local recurrence (LR) and poor overall survival. There is also no universal classification for the resection of pelvic sarcomas invading the sacrum. This study proposes a novel classification system and analyzes the survival and risk of recurrence, when using this system.

Methods

This is a retrospective analysis of 151 patients (with median follow-up in survivors of 44 months (interquartile range 12 to 77)) who underwent hemipelvectomy with iliosacral resection at a single centre between 2007 and 2019. The proposed classification differentiates the extent of iliosacral resection and defines types S1 to S6 (S1 resection medial and parallel to the sacroiliac joint, S2 resection through the ipsilateral sacral lateral mass to the neuroforamina, S3 resection through the ipsilateral neuroforamina, S4 resection through ipsilateral the spinal canal, and S5 and S6 contralateral sacral resections). Descriptive statistics and the chi-squared test were used for categorical variables, and the Kaplan-Meier survival analysis were performed.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 423 - 429
1 Mar 2021
Diez-Escudero A Hailer NP

Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for cementless fixation in host bone, but this option might be highly desirable since the modification of implant surfaces in order to improve osteoconductivity can also increase bacterial adhesion. Therefore, an optimal silver content that inhibits bacterial colonization while maintaining osteoconductivity is crucial if silver were to be applied as a coating on parts intended for bone contact. This review summarizes the different methods used to apply silver coatings to arthroplasty components, with a focus on the amount and duration of silver release from the different coatings; the available experience with silver-coated implants that are in clinical use today; and future strategies to balance the effects of silver on bacteria and eukaryotic cells, and to develop silver-coated titanium components suitable for bone ingrowth.

Cite this article: Bone Joint J 2021;103-B(3):423–429.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 382 - 390
1 Feb 2021
Wang H Tang X Ji T Yan T Yang R Guo W

Aims

There is an increased risk of dislocation of the hip after the resection of a periacetabular tumour and endoprosthetic reconstruction of the defect in the hemipelvis. The aim of this study was to determine the rate and timing of dislocation and to identify its risk factors.

Methods

To determine the dislocation rate, we conducted a retrospective single-institution study of 441 patients with a periacetabular tumour who had undergone a standard modular hemipelvic endoprosthetic reconstruction between 2003 and 2019. After excluding ineligible patients, 420 patients were enrolled. Patient-specific, resection-specific, and reconstruction-specific variables were studied using univariate and multivariate analyses.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 798 - 805
1 Jun 2018
Zhang Y Guo W Tang X Yang R Ji T Yang Y Wang Y Wei R

Aims

The sacrum is frequently invaded by a pelvic tumour. The aim of this study was to review our experience of treating this group of patients and to identify the feasibility of a new surgical classification in the management of these tumours.

Patients and Methods

We reviewed 141 patients who, between 2005 and 2014, had undergone surgical excision of a pelvic tumour with invasion of the sacrum.

In a new classification, pelvisacral (Ps) I, II, and III resections refer to a sagittal osteotomy through the ipsilateral wing of the sacrum, through the sacral midline, or lateral to the contralateral sacral foramina, respectively. A Ps a resection describes a pelvic osteotomy through the ilium and a Ps b resection describes a concurrent resection of the acetabulum with osteotomies performed through the pubis and ischium or the pubic symphysis. Within each type, surgical approaches were standardized to guide resection of the tumour.


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 22 - 30
1 Jan 2018
Brown TS Salib CG Rose PS Sim FH Lewallen DG Abdel MP

Aims

Reconstruction of the acetabulum after resection of a periacetabular malignancy is technically challenging and many different techniques have been used with varying success. Our aim was to prepare a systematic review of the literature dealing with these techniques in order to clarify the management, the rate of complications and the outcomes.

Patients and Methods

A search of PubMed and MEDLINE was conducted for English language articles published between January 1990 and February 2017 with combinations of key search terms to identify studies dealing with periacetabular resection with reconstruction in patients with a malignancy. Studies in English that reported radiographic or clinical outcomes were included. Data collected from each study included: the number and type of reconstructions, the pathological diagnosis of the lesions, the mean age and follow-up, gender distribution, implant survivorship, complications, functional outcome, and mortality. The results from individual studies were combined for the general analysis, and then grouped according to the type of reconstruction.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 267 - 275
1 Feb 2017
Liang H Ji T Zhang Y Wang Y Guo W

Aims

The aims of this retrospective study were to report the feasibility of using 3D-printing technology for patients with a pelvic tumour who underwent reconstruction.

Patients and Methods

A total of 35 patients underwent resection of a pelvic tumour and reconstruction using 3D-printed endoprostheses between September 2013 and December 2015. According to Enneking’s classification of bone defects, there were three Type I lesions, 12 Type II+III lesions, five Type I+II lesions, two Type I+II+III lesions, ten type I+II+IV lesions and three type I+II+III+IV lesions. A total of three patients underwent reconstruction using an iliac prosthesis, 12 using a standard hemipelvic prosthesis and 20 using a screw-rod connected hemipelvic prosthesis.


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1463 - 1470
1 Nov 2016
Grammatopoulos G Alvand A Martin H Whitwell D Taylor A Gibbons CLMH

Objectives. A possible solution for the management of proximal femoral bone loss is a modular femoral endoprosthesis (EPR). Although the outcome of EPRs in tumour surgery has been well described, the outcome of their use in revision hip surgery has received less attention. The aim of this study was to describe the outcome of using EPR for non-neoplastic indications. Methods. A retrospective review of 79 patients who underwent 80 EPRs for non-neoplastic indications was performed, including the rates of complication and survival and the mean Oxford Hip Scores (OHS), at a mean of five years post-operatively. The mean age at the time of surgery was 69 years (28 to 93) and the mean number of previous operations on the hip was 2.4 (0 to 17). The most common indications for EPR implantation were periprosthetic joint infection (PJI) (n = 40), periprosthetic fracture (n = 12) and failed osteosynthesis of a proximal femoral fracture or complex trauma (n = 11). Results. Salvage was achieved in all patients. A total of 25 patients (25 EPRs, 31.6%) had a complication, the most common being infection (n = 9) and dislocation (n = 3). Further surgery was required for 18 EPRs (22%), nine of which were revision procedures. The five year survival of the EPR was 87% (95%CI: 76% to 98%). The mean OHS was 28 (4 to 48). Inferior survival and outcomes were seen in EPRs which were performed for the treatment of infection. However, the eradication of infection was achieved in 33 of the 40 (82.5%) which were undertaken for this indication. Conclusion. We recommend the continued use of proximal femoral EPRs for non-neoplastic indications, including PJI. Cite this article: Bone Joint J 2016;98-B:1463–70


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 33 - 33
1 May 2016
Baxmann M Pfaff A Grupp T Morlock M
Full Access

Introduction. Dual modular hip prostheses were introduced to optimize the individual and intra-surgical adaptation of the implant design to the native anatomics und biomechanics of the hip. The downside of a modular implant design with an additional modular interface is the potential susceptibility to fretting, crevice corrosion and wear [1–2]. The purpose of this study was to characterize the metal ion release of a modular hip implant system with different modular junctions and material combinations in consideration of the corrosive physiological environment. Methods. One design of a dual modular hip prosthesis (Ti6Al4V, Metha®, Aesculap AG, Germany) with a high offset neck adapter (CoCrMo, CCD-angle of 130°, neutral antetorsion) and a monobloc prosthesis (stem size 4) of the same implant type were used to characterize the metal ion release of modular and non-modular hip implants. Stems were embedded in PMMA with 10° adduction and 9° flexion according to ISO 7206-6 and assembled with ceramic (Biolox® delta) or CoCrMo femoral heads (XL-offset) by three light impacts with a hammer. All implant options were tested in four different test fluids: Ringer's solution, bovine calf serum and iron chloride solution (FeCl3-concentration: 10 g/L and 114 g/L). Cyclic axial sinusoidal compressive load (Fmax = 3800 N, peak load level of walking based on in vivo force measurements [3]) was applied for 10 million cycles using a servohydraulic testing machine (MTS MiniBionix 370). The test frequency was continuously varied between 15 Hz (9900 cycles) followed by 1 Hz (100 cycles). The metal ion concentration (cobalt, chromium and titanium) of the test fluids were analysed using ICP-OES and ICP-MS at intervals of 0, 5·105, 2·106 and 10·106 cycles (measuring sensitivity < 1 µg/L). Results. Due to the additional modular interface between stem and neck adapter the total metal ion release of the modular hip endoprosthesis system increased significantly and is comparable to the coupling of a monobloc stem and a CoCrMo femoral head (Fig. 1). The application of ceramic femoral heads reduced the total cobalt and chromium release in the stem-head taper interface of non-modular and modular stems. In comparison between the four test fluids could be observed that lower pH-values and higher FeCl3-concentrations increased the metal ion release (Fig 2). In contrast, the use of bovine calf serum decreased the metal ion release of modular junctions due to the presence of proteins and other organic components. Discussion. For testing hip implants with proximal femoral modularity according to ISO and ASTM standards, sodium chloride solutions are frequently used to determine the fatigue strength and durability of the stem-neck connection. The present study illustrate that the expansion of standard requirements of biomechanical testing and the use of alternative test fluids is necessary to simulate metal ion release by electro-chemical processes. A promising approach is the use of adapted iron-chloride solutions (10 g/L FeCl3, pH 2) to evaluate the susceptibility of modular hip junctions to fretting, crevice and contact corrosion


Bone & Joint 360
Vol. 4, Issue 4 | Pages 30 - 31
1 Aug 2015

The August 2015 Oncology Roundup360 looks at: Glasgow prognostic score in soft-tissue sarcoma; Denosumab in giant cell tumour; Timing, complications and radiotherapy; Pigmented villonodular synovitis and arthroscopy; PATHFx: estimating survival in pathological cancer; Prosthetic lengthening of short stumps; Chondrosarcoma and pathological fracture


Bone & Joint 360
Vol. 2, Issue 6 | Pages 28 - 31
1 Dec 2013

The December 2013 Oncology Roundup360 looks at: Peri-articular resection fraught with complications; Navigated margins; Skeletal tumours and thromboembolism; Conditional survival in Ewing’s sarcoma; Reverse shoulders and tumour; For how long should we follow up sarcoma patients?; and already metastasised?


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1104 - 1110
1 Aug 2011
Ji T Guo W Yang R Tang S Sun X

We set out to determine the impact of surgery on quality of life and function in patients who had undergone surgery for symptomatic peri-acetabular metastases. From a prospective database we retrospectively reviewed 46 consecutive patients who had been treated operatively between June 2003 and June 2009. The mean age of the patients was 56.4 years (20 to 73) and the mean post-operative follow-up was 19.2 months (4 to 70). Functional evaluation and quality-of-life assessments were performed. At the most recent follow-up, 26 patients (56.5%) were alive. Their median survival time was 25.0 months. Ten major postoperative complications had occurred in eight patients (17.4%). The mean post-operative Musculoskeletal Tumor Society score (MSTS 93) was 56.3% (6.7% to 90.0%). Improvement in the Eastern Cooperative Oncology Group (ECOG) performance status was seen in 32 patients (69.6%). On the European Organisation for Research and Treatment of Cancer core quality-of-life questionnaire (QLQ-C30) measure of global health status there was a statistically significant improvement from the patients’ pre-operative status (42.8 (sd 13.7)) to that found at the latest follow-up (58.0 (sd 12.5)) (p = 0.001). The only statistically significant change in the nine symptom domains of the QLQ-C30 was a reduction in the mean level of pain (from 59.1 to 29.5 (out of 100)) (p < 0.001).

Surgery for patients with peri-acetabular metastases reduces pain and improves their quality of life, and has a low rate of surgical complications.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 314 - 314
1 Jul 2011
Ahmed I Gray A Aderinto J Howie C Patton J
Full Access

Background: Chronic prosthetic joint infection is a cause of patient morbidity and can be challenging to treat. Surgeons performing revision arthroplasty of the hip and knee are confronted with a growing number of patients with extensive loss of bone stock. The use of a modular endoprosthesis is a possible method of treatment in such patients. Aim: The purpose of this study was to assess the functional outcomes and the success of a single and two stage revision procedure in eradicating chronic prosthetic joint infection using a femoral endoprosthesis. Methods: A prospective database was reviewed of 20 patients who underwent a proximal, total or distal femoral endoprosthetic replacement after chronic prosthetic infection. Radiographs performed at the time of latest follow up were evaluated for signs of loosening, osteomyelitis and implant failure. The functional status was assessed using the Short Form (SF)-36 health survey score, Toronto extremity salvage score (TESS) and the Enneking score. Results: Thirteen patients underwent a single stage revision procedure and seven had been treated with a staged revision. At the latest follow up none of the 13 patients treated with a single stage procedure had evidence of ongoing infection. Of the seven patients who had a staged revision, 3 patients had evidence of ongoing infection. The mean pre operative Enneking score for the entire group was 17.1 points and this improved to 47.5 points post operatively (p< 0.0002). The mean pre operative TESS score for the entire group was 42% and this improved to 59% post operatively (p< 0.005). There was also a statistically significant improvement in all of the components of the SF-36 score. Conclusion: We believe that the use of a modular endo-prosthesis in the treatment of chronic prosthetic joint infection is a successful and viable option in eradicating infection, preserving the limb and providing a good functional result


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 118 - 118
1 May 2011
Puchner S Hofstaetter J Hipfl C Funovics P Kotz R Dominkus M
Full Access

Background: Endoprosthetic reconstruction has become the gold standard of treatment after the resection of tumors around the proximal femur, however, the rate of complications linked to megaprostheses is clearly higher than with standard implants. Aim of this study was to investigate the incidence and type of complications related to modular proximal femur prostheses. Patients and Methods: By retrospective database analysis of the Vienna Tumor Registry, we evaluated the incidence of complications in 170 consecutive patients who have received a proximal femur KMFTR/GMRS at our institution between 1982 and 2007. 71 patients with an average age of 41.7 years (range 18.2–79.9 years) received the implant following the resection of a malignant bone tumor, 95 patients with an average age of 61.7 years (range 5.9–84.2 years) due to metastatic disease. The average time of follow-up was 3.5 years (SD ±4.9 years). Results: Overall patient survival at five years was 32%. For patients with metastatic disease the overall survival was 10% at five years. Patients being treated for a primary bone tumor had an overall survival of 55% at 5 years. The overall survival of the prosthesis was 90% at two years and 72% at five years. Twenty-one patients (12.65%) suffered from dislocation after a mean time of 6.5 month (range 0.3–33 months) after surgery. Out of these, nine had to be treated by open reduction. Nine patients suffered recurrent dislocation after their first event. Patients who underwent extensive pelvic reconstruction had a significantly higher dislocation rate (33.3%) compared to patients with no or standard acetabular components (11.2%). Deep infection occurred in twelve patients (7.3%) after a mean of 39 months (range 1–166 months) after surgery. Treatment of infection was one-stage revision in eight and hip disarticulation in one patient. Two patients were successfully treated by local wound revision. One patient died of septic shock four days after surgery. Re-infection occurred in three patients. Aseptic loosing occurred in 13 patients (12.8%) after a mean time of 75.6 months (range 1–223 months) after surgery. Revision surgery was necessary in 27 patients (15.8%) with a mean time to first revision of 32 months (range 0.3–116 months). Prosthetic fracture was found in two patients. Local recurrence occurred in 14 patients (8.4%). In seven patients (4.2%) amputation was necessary. Conclusion: Modular endoprosthesis allow excellent reconstruction of the proximal femur following tumor resection. However, the main complications, dislocation and infection, still remain considerable drawbacks


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 467 - 467
1 Jul 2010
Aliev M Nisichenko D Saravanan S Orekhov M Sergeev P Babalaev A Sokolovskiy V
Full Access

From 1992 on 2008, 615/515 patients underwent primary or revisional endoprosthetic replacement of major joints. In 51 patients (31 men & 20 women) modular system MUTARS (Implantcast, Germany) has been used. The median age was 23.3 years (15 to 52 years). MUTARS modular endoprosthesis has been used in 10 patients with deep infection of endoprosthetic bed as a revisional endoprosthetic replacement: 1 Total endoprosthetic replacement of femur, 5 Total knee joint replacement (2 for distal femoral defect and 3 for proximal tibial defect). In 3(27%) patients, we used newly patented silver ion coated MUTARS either after two stage treatment for infection of endoprosthetic bed or as a prophylaxis of endoprosthetic infection. In 1 patient (23 yrs), with 12cm limb length shortening, we used extensible MUTARS as a revisional endoprosthetic replacement. The following complications we observed: Instability of endoprosthesis – 3/51 (5.9%), deep endoprosthetic bed infection – 4/51 (7.8%). In comparison group, when using custom-made endoprosthesis, the frequency of infectious complications have made 60/574 (10.5 %), and instability of implants was observed in 79/574 (13.8 %) cases. Transition of using modular systems for primary and revisional endoprosthesis allows to reduce the level of instability from 13.8 % to 5.9 %. The quantity of infectious complications is also not great as in comparison with control group. For revisional endoprosthetic replacement, we think, the given modular system is optimal, for correcting limb length deficiency and restoration of basic function at patients. Use of silver ion coated modular implants is a promising method for treating deep endoprosthetic bed infection


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1632 - 1637
1 Dec 2007
Myers GJC Abudu AT Carter SR Tillman RM Grimer RJ

We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database.

A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 790 - 795
1 Jun 2006
Ahlmann ER Menendez LR Kermani C Gotha H

We reviewed retrospectively the results in 211 consecutive patients who had undergone limb salvage for bone neoplasia with endoprosthetic reconstruction of the proximal femur (96), distal femur (78), proximal tibia (30) and total femur (7). Their mean age was 50 years (11 to 86) and the mean follow-up period was 37.3 months (1 to 204). A total of 35 (16.6%) prostheses failed. Overall, implant survival was 78% (95% confidence interval (CI) 0.29 to 0.54) at five years, 60% (95% CI 0.93 to 2.35) at ten years and 60% (95% CI 1.27 to 3.88) at 15 years. Survivorship of the limb was 97.6% (95% CI 1.73 to 3.35) at ten years. The gender, age, diagnosis and location of the tumour were not prognostic variables for failure. Modular endoprosthetic replacement in the lower limb is a durable long-term reconstructive option, with the implants generally outlasting the patient.