The April 2024 Trauma Roundup360 looks at: The infra-acetabular screw in acetabular fracture surgery; Is skin traction helpful in patients with intertrochanteric hip fractures?; Reducing pain and improving function following hip fracture surgery; Are postoperative splints helpful following ankle fracture fixation?; Biomechanics of internal fixation in Hoffa fractures: a comparison of four different constructs; Dual-plate fixation of periprosthetic distal femur fractures; Do direct oral anticoagulants necessarily mean a delay to hip fracture surgery?; Plate or retrograde nail for low distal femur fractures?.
The October 2023 Trauma Roundup360 looks at: Intramedullary nailing versus sliding hip screw in trochanteric fracture management: the INSITE randomized clinical trial; Five-year outcomes for patients with a displaced fracture of the distal tibia; Direct anterior versus anterolateral approach in hip joint hemiarthroplasty; Proximal humerus fractures: treat them all nonoperatively?; Tranexamic acid administration by prehospital personnel; Locked plating versus nailing for proximal tibia fractures: a multicentre randomized controlled trial; A retrospective review of the rate of septic knee arthritis after retrograde femoral nailing for traumatic femoral fractures at a single academic institution.
The August 2023 Trauma Roundup360 looks at: A comparison of functional cast and volar-flexion ulnar deviation for dorsally displaced distal radius fractures; Give your stable ankle fractures some AIR!; Early stabilization of rib fractures – an effective thing to do?; Locked plating versus nailing for proximal tibia fractures: A multicentre randomized controlled trial; Time to flap coverage in open tibia fractures; Does tranexamic acid affect the incidence of heterotropic ossification around the elbow?; High BMI – good or bad in surgical fixation of hip fractures?
The June 2023 Trauma Roundup. 360. looks at: Aspirin or low-molecular-weight heparin for thromboprophylaxis?;
The objective of this study was to use patient-specific finite element modeling to measure the 3D interfragmentary strain environment in clinically realistic fractures. The hypothesis was that in the early post-operative period, the tissues in and around the fracture gap can tolerate a state of strain in excess of 10%, the classical limit proposed in the Perren strain theory. Eight patients (6 males, 2 females; ages 22–95 years) with distal femur fractures (OTA/AO 33-A/B/C) treated in a Level I trauma center were retrospectively identified. All were treated with
The April 2023 Trauma Roundup360 looks at: Displaced femoral neck fractures in patients aged 55 to 70 years: internal fixation or total hip arthroplasty?; Tibial plateau fractures: continuous passive motion approves range of motion; Lisfranc fractures: to fuse or not to fuse, that is the question; Is hardware removal after clavicle fracture plate fixation beneficial?; Fixation to coverage in Grade IIIB open fractures – what’s the time window?; Nonoperative versus locking plate fixation in the proximal humerus; Retrograde knee nailing or lateral plate for distal femur fractures?
Introduction and Objective. Distal femoral fractures are commonly treated with a straight plate fixed to the lateral aspects of both proximal and distal fragments. However, the lateral approach may not always be desirable due to persisting soft-tissue or additional vascular injury necessitating a medial approach. These problems may be overcome by pre-contouring the plate in helically shaped fashion, allowing its distal part to be fixed to the medial aspect of the femoral condyle. The objective of this study was to investigate the biomechanical competence of medial femoral helical plating versus conventional straight
Introduction and Objective. Plating of geriatric distal femoral fractures with Locking Compression Plate Distal Femur (LCP–DF) often requires augmentation with a supplemental medial plate to achieve sufficient stability allowing early mobilization. However, medial vital structures may be impaired by supplemental medial plating using a straight plate. Therefore, a helically shaped medial plate may be used to avoid damage of these structures. Aim of the current study was to investigate the biomechanical competence of augmented LCP–DF plating using a supplemental straight versus helically shaped medial plate. Materials and Methods. Ten pairs of human cadaveric femora with poor bone quality were assigned pairwise for instrumentation using a lateral anatomical 15-hole LCP–DF combined with a medial 14-hole LCP, the latter being either straight or manually pre-contoured to a 90-degree helical shape. An unstable distal femoral fracture AO/OTA 33–A3 was simulated by means of osteotomies. All specimens were biomechanically tested under non-destructive quasi-static and destructive progressively increasing combined cyclic axial and torsional loading in internal rotation, with monitoring by means of optical motion tracking. Results. Initial axial stiffness and torsional stiffness in internal and external rotation for straight double plating (548.1 ± 134.2 N/mm, 2.69 ± 0.52 Nm/° and 2.69 ± 0.50 Nm/°) was significantly higher versus helical double plating (442.9 ± 133.7 N/mm, 2.07 ± 0.32 Nm/° and 2.16 ± 0.22 Nm/°), p≤0.04. Initial interfragmentary axial displacement and flexural rotation under 500 N static loading were significantly smaller for straight plating (0.11 ± 0.14 mm and 0.21 ± 0.10°) versus helical plating (0.31 ± 0.14 mm and 0.68 ± 0.16°), p<0.01. However, initial varus deformation under this loading remained not significantly different between the two fixation methods (straight: 0.57 ± 0.23°, helical: 0.75 ± 0.34°), p=0.08. During dynamic loading, within the course of the first 4000 cycles the movements of the distal fragment in flexion were significantly bigger for helical over straight plating (1.03 ± 0.33° versus 0.40 ± 0.20°), p<0.01. However, no significant differences were observed between the two fixation methods in terms of varus, internal rotation, axial and shear displacements at the fracture site, and number of cycles to failure. Conclusions. Augmented
Dual plating of the medial and lateral distal femur has been proposed to reduce angular malunion and hardware failure secondary to delayed union or nonunion. This strategy improves the strength and alignment of the construct, but it may compromise the vascularity of the distal femur paradoxically impairing healing. This study investigates the effect of dual plating versus single plating on the perfusion of the distal femur. Ten matched pairs of fresh-frozen cadaveric lower extremities were assigned to either isolated
Tibial plateau fractures are serious injuries about the knee that have the potential to affect patients’ long-term function. To our knowledge, this is the first study to use patient-reported outcomes (PROs) with a musculoskeletal focus to assess the long-term outcome, as compared to a short-term outcome baseline, of tibial plateau fractures treated using modern techniques. In total, 102 patients who sustained a displaced tibial plateau fracture and underwent operative repair by one of three orthopaedic traumatologists at a large, academic medical centre and had a minimum of five-year follow-up were identified. Breakdown of patients by Schatzker classification is as follows: two (1.9%) Schatzker I, 54 (50.9%) Schatzker II, two (1.9%) Schatzker III, 13 (12.3%) Schatzker IV, nine (8.5%) Schatzker V, and 26 (24.5%) Schatzker VI. Follow-up data obtained included: Visual Analogue Scale (VAS) or Numeric Rating Scale (NRS) pain scores, Short Musculoskeletal Functional Assessment (SMFA), and knee range of movement (ROM). Data at latest follow-up were then compared to 12-month data using a paired Aims
Methods
Aims. Dual plating of distal femoral fractures with medial and lateral implants has been performed to improve construct mechanics and alignment, in cases where isolated
Peri-prosthetic fractures above a TKA are becoming increasingly more common, and typically occur at the junction of the anterior flange of the femoral component and the osteopenic metaphyseal distal femur. In the vast majority of cases, the TKA is well fixed and has been functioning well prior to fracture. For fractures above well-fixed components, internal fixation is preferred. Fixation options include retrograde nailing or
Peri-prosthetic fractures above a TKA are becoming increasingly more common, and typically occur at the junction of the anterior flange of the femoral component and the osteopenic metaphyseal distal femur. In the vast majority of cases the TKA is well fixed and has been functioning well prior to fracture. For loose components, revision is typically indicated. Typically a megaprosthesis is required. Well-fixed components, internal fixation is preferred. Fixation options include retrograde nailing or
Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. The medical records and radiographs of all patients who underwent
CHAOS for femoral deformity at our institution between 2005 and
2011 were retrospectively reviewed. Records were available for all
55 consecutive procedures undertaken in 49 patients with a mean
age of 35.6 years (10.9 to 75.3) at the time of surgery.Aims
Patients and Methods
Peri-prosthetic fractures above a TKA are becoming increasingly more common, and typically occur at the junction of the anterior flange of the femoral component and the osteopenic metaphyseal distal femur. In the vast majority of cases the TKA is well fixed and has been functioning well prior to fracture. For loose components, revision is typically indicated. Often, distal femoral mega prostheses are required to deal with metaphyseal bone loss. Good results have been reported in small series, however, complications, including infection remain concerning, and these implants are incredibly expensive. Although performing a mega prosthesis in the setting of a well fixed TKA is not unreasonable due to immediate full weight bearing, in my opinion, prosthetic replacement should be limited to cases of failed ORIF (rare), or in cases where fixation is likely to fail (i.e., severe osteolysis distally). For the majority of fractures above well fixed components, internal fixation is preferred for the main reason that the overwhelming majority of these fractures will heal. Fixation options include retrograde nailing or
Periprosthetic femoral shaft fractures are a significant complication of total hip arthroplasty. Plate osteosynthesis with or without onlay strut allograft has been the mainstay of treatment around well-fixed stems. Nonunions are a rare, challenging complication of this fixation method. The number of published treatment strategies for periprosthetic femoral nonunions are limited. In this series, we report the outcomes of a novel orthogonal plating surgical technique for addressing nonunions in the setting of Vancouver B1 and C-type periprosthetic fractures that previously failed open reduction internal fixation (ORIF). A retrospective chart review of all patients from 2010 to 2014 with Vancouver B1/C total hip arthroplasty periprosthetic femoral nonunions was performed. All patients were treated primarily with ORIF. Nonunion was defined as no radiographic signs of fracture healing nine months post-operatively, with or without hardware failure. Exclusion criteria included open fractures and periprosthetic infections. The technique utilised a mechanobiologic strategy of atraumatic exposure, resection of necrotic tissue, bone grafting with adjuvant recombinant growth factor and revision open reduction internal fixation. Initially, compression was achieved using an articulated tensioning device and application of an anterior plate. This was followed by locked
Peri-prosthetic fractures above a total knee arthroplasty (TKA) are becoming increasingly more common, and typically occur at the junction of the anterior flange of the femoral component and the osteopenic metaphyseal distal femur. In the vast majority of cases the TKA is well-fixed and has been functioning well prior to fracture. For loose components, revision is typically indicated. Typically a megaprosthesis is required. For well-fixed components, internal fixation is preferred. Fixation options include retrograde nailing or
This study tests the biomechanical properties of adjacent locked
plate constructs in a femur model using Sawbones. Previous studies
have described biomechanical behaviour related to inter-device distances.
We hypothesise that a smaller lateral inter-plate distance will
result in a biomechanically stronger construct, and that addition
of an anterior plate will increase the overall strength of the construct. Sawbones were plated laterally with two large-fragment locking
compression plates with inter-plate distances of 10 mm or 1 mm.
Small-fragment locking compression plates of 7-hole, 9-hole, and
11-hole sizes were placed anteriorly to span the inter-plate distance.
Four-point bend loading was applied, and the moment required to
displace the constructs by 10 mm was recorded.Objectives
Methods