This study aimed to investigate the optimal sagittal positioning of the uncemented femoral component in total knee arthroplasty to minimize the risk of aseptic loosening and periprosthetic fracture. Ten different sagittal placements of the femoral component, ranging from -5 mm (causing anterior notch) to +4 mm (causing anterior gap), were analyzed using finite element analysis. Both gait and squat loading conditions were simulated, and Von Mises stress and interface micromotion were evaluated to assess fracture and loosening risk.Aims
Methods
Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected
Sagittal plane imbalance (SPI), or asymmetry between extension and flexion gaps, is an important issue in total knee arthroplasty (TKA). The purpose of this study was to compare SPI between kinematic alignment (KA), mechanical alignment (MA), and functional alignment (FA) strategies. In 137 robotic-assisted TKAs, extension and flexion stressed gap laxities and bone resections were measured. The primary outcome was the proportion and magnitude of medial and lateral SPI (gap differential > 2.0 mm) for KA, MA, and FA. Secondary outcomes were the proportion of knees with severe (> 4.0 mm) SPI, and resection thicknesses for each technique, with KA as reference.Aims
Methods
Bacteriophages infect, replicate inside bacteria, and are released from the host through lysis. Here, we evaluate the effects of repetitive doses of the For the haematogenous infection, Aims
Methods
This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height. Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.Aims
Methods
Aims. Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium. Methods. Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia. Results. Median metal concentrations were as follows: cobalt: 0.69 μg/g (interquartile range (IQR) 0.10 to 6.10); chromium: 1.1 μg/g (IQR 0.27 to 4.10); and titanium: 1.6 μg/g (IQR 0.90 to 4.07). Moderate ALVAL scores were found in 30% (n = 39) of the revised knees. There were ten patients with an ALVAL score of 6 or more who were revised for suspected periprosthetic joint infection (PJI), aseptic loosening, or osteolysis. R2 varied between 0.269 and 0.369 for the ordinal regression models. The most important variables were model type, indication for revision, and cobalt and chromium in the ordinal regression models. Conclusion. We found that metal particles released from the
While mechanical alignment (MA) is the traditional technique in total knee arthroplasty (TKA), its potential for altering constitutional alignment remains poorly understood. This study aimed to quantify unintentional changes to constitutional coronal alignment and joint line obliquity (JLO) resulting from MA. A retrospective cohort study was undertaken of 700 primary MA TKAs (643 patients) performed between 2014 and 2017. Lateral distal femoral and medial proximal tibial angles were measured pre- and postoperatively to calculate the arithmetic hip-knee-ankle angle (aHKA), JLO, and Coronal Plane Alignment of the Knee (CPAK) phenotypes. The primary outcome was the magnitude and direction of aHKA, JLO, and CPAK alterations.Aims
Methods
To investigate the efficacy of ethylenediaminetetraacetic acid-normal saline (EDTA-NS) in dispersing biofilms and reducing bacterial infections. EDTA-NS solutions were irrigated at different durations (1, 5, 10, and 30 minutes) and concentrations (1, 2, 5, 10, and 50 mM) to disrupt Aims
Methods
Numerous papers present in-vivo knee kinematics data following total knee arthroplasty (TKA) from fluoroscopic testing. Comparing data is challenging given the large number of factors that potentially affect the reported kinematics. This paper aims at understanding the effect of following three different factors: implant geometry, performed activity and analysis method. A total of 30 patients who underwent TKA were included in this study. This group was subdivided in three equal groups: each group receiving a different type of posterior stabilized total
This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
To evaluate mid-to long-term patient-reported outcome measures (PROMs) of endoprosthetic reconstruction after resection of malignant tumours arising around the knee, and to investigate the risk factors for unfavourable PROMs. The medical records of 75 patients who underwent surgery between 2000 and 2020 were retrospectively reviewed, and 44 patients who were alive and available for follow-up (at a mean of 9.7 years postoperatively) were included in the study. Leg length discrepancy was measured on whole-leg radiographs, and functional assessment was performed with PROMs (Toronto Extremity Salvage Score (TESS) and Comprehensive Outcome Measure for Musculoskeletal Oncology Lower Extremity (COMMON-LE)) with two different aspects. The thresholds for unfavourable PROMs were determined using anchor questions regarding satisfaction, and the risk factors for unfavourable PROMs were investigated.Aims
Methods
Aim. In 10% of the presumed aseptic hip or knee revisions, a low-grade infection is unexpectedly diagnosed based on the tissue samples taken during revision. Extended antimicrobial prophylaxis can possibly reduce the failure rate in cases of unexpected PJI, because the prophylaxis can be considered as early empiric treatment. In this randomized controlled study we analysed whether extended antimicrobial prophylaxis compared to a single dose is beneficial to improve the outcome of treatment in unexpected PJI in revision arthroplasty. Method. This study was nested in a randomized clinical trial comparing single-dose cefazolin with prolonged prophylaxis (15 doses of cefazolin over 5 days) for revision arthroplasty of the hip or knee. For this analysis, patients were included if an unsuspected PJI (defined as ≥2 positive intraoperative tissue samples with the same microorganism) was diagnosed. PJI treatment consisted of 12 weeks of a rifampicin-based regimen in Staphylococcal PJI, without removal of the prosthesis. We examined Infection characteristics and success of treatment after one year, defined as the absence of signs or treatment for PJI during follow-up. Results. After randomization of 662 patients, 68 unexpected PJI were diagnosed. In 5 cases no antimicrobial treatment was started. The success rate after one year follow-up for those who received PJI treatment was 96% (28/29) in the single dose group and 91% (31/34) in the extended prophylaxis group (p=1.00). The most frequently identified pathogens in unexpected PJI were Cutibacterium acnes (n=50) and Staphylococcus epidermidis (n=14). The causatives were susceptible for the cefazolin prophylaxis in 61 of the 63 cases. The interval between the stopped prophylaxis and the re-start of antimicrobial treatment was on average 10 days (SD 4) for the single dose and 5 days (SD 4) for the extended group. The mean duration of antimicrobial treatment was 83 days (SD 12) and did not differ between both groups (p=0.16). Conclusions. This is the first randomized controlled trial in which extended prophylaxis showed no benefit on the prosthesis survival for patients with an unexpected PJI after assumed aseptic revision of the hip or
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
Achievement of accurate microbiological diagnosis prior to revision is key to reducing the high rates of persistent infection after revision knee surgery. The effect of change in the microorganism between the first- and second-stage revision of total knee arthroplasty for periprosthetic joint infection (PJI) on the success of management is not clear. A two-centre retrospective cohort study was conducted to review the outcome of patients who have undergone two-stage revision for treatment of knee arthroplasty PJI, focusing specifically on isolated micro-organisms at both the first- and second-stage procedure. Patient demographics, medical, and orthopaedic history data, including postoperative outcomes and subsequent treatment, were obtained from the electronic records and medical notes.Aims
Methods
Several previously identified patient-, injury-, and treatment-related factors are associated with the development of nonunion in distal femur fractures. However, the predictive value of these factors is not well defined. We aimed to assess the predictive ability of previously identified risk factors in the development of nonunion leading to secondary surgery in distal femur fractures. We conducted a retrospective cohort study of adult patients with traumatic distal femur fracture treated with lateral locking plate between 2009 and 2018. The patients who underwent secondary surgery due to fracture healing problem or plate failure were considered having nonunion. Background knowledge of risk factors of distal femur fracture nonunion based on previous literature was used to form an initial set of variables. A logistic regression model was used with previously identified patient- and injury-related variables (age, sex, BMI, diabetes, smoking, periprosthetic fracture, open fracture, trauma energy, fracture zone length, fracture comminution, medial side comminution) in the first analysis and with treatment-related variables (different surgeon-controlled factors, e.g. plate length, screw placement, and proximal fixation) in the second analysis to predict the nonunion leading to secondary surgery in distal femur fractures.Aims
Methods
Abstract. INTRODUCTION. The anatomic distal femoral locking plate (DF-LCP) has simplified the management of supracondylar femoral fractures with stable
Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model. We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.Aims
Methods
The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.Aims
Methods
Total Knee Arthroplasty (TKA) is an effective surgery performed for knee osteoarthritis. Despite this success up to 20% of patients are less than fully satisfied. Rotating platform (RP) bearings have demonstrated significant reduction in rates of wear in laboratory studies but thus far have failed to demonstrate a clinical difference compared to fixed bearing (FB) arthroplasty in patients. This may be due to studies limited by small sample size and single centre design. Additionally, no studies have utilised modern patient-reported outcome measures (PROMs) such as PKIP to investigate differences between these bearings. A non-randomised, prospective, multi-centre study was conducted across centres in Australia, United Kingdom and the United States from November 2012 to May 2015.
Constrained implants with intra-medullary fixation are expedient for complex TKA. Constraint is associated with loosening, but can correction of deformity mitigate risk of loosening?. Primary TKA's with a non-linked constrained prosthesis from 2010-2018 were identified. Indications were ligamentous instability or intra-medullary fixation to bypass stress risers. All included fully cemented 30mm stem extensions on tibia and femur. If soft tissue stability was achieved, a posterior stabilized (PS) tibial insert was selected. Pre and post TKA full length radiographs showed. i. hip-knee-ankle angles (HKAA). ii. Kennedy Zone (KZ) where hip to ankle vector crosses knee joint. 77 TKA's in 68 patients, average age 69.3 years (41-89.5) with OA (65%) post-trauma (24.5%) and inflammatory arthropathy (10.5%). Pre-op radiographs (62 knees) showed varus in 37.0%. (HKAA: 4. o. -29. o. ), valgus in 59.6% (HKAA range 8. o. -41. o. ) and 2 knees in neutral. 13 cases deceased within 2 years were excluded. Six with 2 year follow up pending have not been revised. Mean follow-up is 6.1 yrs (2.4-11.9yrs). Long post-op radiographs showed 34 (57.6%) in central KZ (HKKA 180. o. +/- 2. o. ). . Thirteen (22.0%) were in mechanical varus (HKAA 3. o. -15. o. ) and 12 (20.3%) in mechanical valgus: HKAA (171. o. -178. o. ). Three failed with infection; 2 after ORIF and one with BMI>50. The greatest post op varus suffered peri-prosthetic fracture. There was no aseptic loosening or instability. Only full-length radiographs accurately measure alignment and very few similar studies exist. No cases failed by loosening or instability, but PPF followed persistent malalignment. Infection complicated prior ORIF and elevated BMI. This does not endorse indiscriminate use of mechanically constrained