Aims. Retrospective review of a consecutive series of 1,168 total ankle replacements (TAR) performed at Wrightington, to analyse modes of failure and clinical outcomes following
One assumed function of Total Ankle Replacement (TAR) is that by maintaining ankle joint motion we can protect the other hind foot joints from further degredation. 1. However, there is no work to our knowledge that compares hindfoot outcomes between
Background. Patients who undergo either primary or revision total ankle replacement (TAR) expect improvements in pain, function and quality of life. The goal of this study was to measure the functional outcome improvements and the difference in patient-reported outcomes in patients undergoing primary total ankle replacements compared to revision
Background. Total Ankle Replacement (TAR) is an established treatment option for end-stage ankle arthritis. We analysed at minimum, 10-year patient-reported and clinical outcomes of 156 TARs from a single centre. We specifically compared outcomes between patients under 60 and over 60 at time of surgery. Methods. Data was collected retrospectively from our departmental patient database. It included all patients who underwent a
Despite advancements, revision rates following total ankle replacement (TAR) are high in comparison to other total joint replacements. This explant analysis study aimed to investigate whether there was appreciable metal particulate debris release from various contemporary TARs by describing patterns of material loss. Twenty-eight explanted TARs (9 designs: 3 fixed and 6 mobile bearing), revised for any reason, were studied. The articulating surfaces of the metal tibial and talar components as well as the polyethylene insert were assessed for damage features using light microscopy. Based on the results of the microscopic analysis, scanning electron microscopy with energy dispersive X-ray spectroscopy was performed to determine the composition of embedded debris identified, as well as non-contacting 3D profilometry. Pitting, indicative of material loss, was identified on the articulating surfaces of 54% of tibial components and 96% of talar components. Bearing constraint was not found to be a factor, with similar proportions of fixed and mobile bearing metal components showing pitting. More cobalt-chromium than titanium alloy tibial components exhibited pitting (63% versus 20%). Significantly higher average surface roughness (Sa) values were measured for pitted areas in comparison to unpitted areas of these metal components (p<0.05). Additionally, metallic embedded debris (cobalt-chromium likely due to pitting of the tibial and talar components or titanium likely from loss of their porous coatings) was identified in 18% of polyethylene inserts. The presence of hard 3. rd. body particles was also indicated by macroscopically visible sliding plane scratching, identified on 79% of talar components. This explant analysis study demonstrates that metal debris is released from the articulating surfaces and the coatings of various contemporary TARs, both fixed and mobile bearing. These findings suggest that metal debris release in TARs may be an under-recognised issue that should be considered in the study of painful or failed
Introduction. Primary ankle arthroplasty (TAR) is increasingly used to treat end-stage ankle arthritis. Reported revision rates of
Ankle fusion (AF), a durable intervention for ankle arthritis, has been the management of choice but restricts mobility. Recently, total ankle replacement (TAR) has been offered to patients looking to maintain mobility. The aim was to compare the biomechanics of AF and
Introduction. We report the functional outcome and survivorship of the Hintegra Total Ankle Replacement (TAR), in consecutive cases by multiple surgeons in a single UK institution. Between 2010–2014 the Hintegra
Introduction. Day Case Surgery (defined as same day discharge) is a priority within the National Health Service and has been shown to provide beneficial outcomes for patients and hospitals. We report our experience developing a Day Case Programme for Total Ankle Replacement (TAR). Methods. Prior to the introduction of a Day Case Programme, average length of stay following
Introduction. Total ankle replacement (TAR) is performed for post-traumatic arthritis, inflammatory arthropathy, osteoarthritis and other indications. The Scottish Arthroplasty Project (SAP) began collection of data on
Background. The importance of total ankle replacement (TAR) implant orientation in the axial plane is poorly understood with major variation in surgical technique of implants on the market. Our aims were to better understand the axial rotational profile of patients undergoing
Introduction. Comparative studies examining Fixed-Bearing (FB) and Mobile-Bearing (MB) Total Ankle Replacement (TAR) designs have demonstrated similar results and successful long-term outcomes for both. To date there has been no study directly comparing FB and MB designs of the same prosthesis. We present the first prospective randomised trial comparing patient satisfaction, functional outcomes and radiographic results of the Salto Talaris Fixed-Bearing and the Salto Mobile-Bearing Total Ankle Replacement in the treatment of end-stage ankle arthritis. Methods. A total of 108 adult patients with end-stage ankle arthritis were enrolled in the study between November 2014 and October 2021 with similar demographic comparison. Prospective patient-reported outcomes and standardised weightbearing ankle radiographs were performed preoperatively, at 6 weeks, 6 months and 12 months post-operatively, followed by yearly intervals. All surgeries were performed by a single non-design orthopaedic foot and ankle specialist with experience in over 200 Salto and Salto Talaris
Introduction. Total ankle arthroplasty (TAA) is an increasingly popular treatment option for patients with end-stage ankle arthritis. However, for most implant systems, failure rates of 10–20% have been reported within the first 10 years after primary TAA. Pain is the primary symptom that indicates failure of TAA but cause of it can be difficult to establish. Methods. All patients who underwent a primary TAA at our center were included in the study. The clinical outcomes were studied for patients requiring a further revision procedure following primary TAA. The reasons for revision surgery and outcomes of surgery were analyzed using appropriate inferential statistical tests. Results. Between 2007 and 2018, 42 primary TAA required revisions in 40 patients. There were 25 men (59.5%) and 15 women (35.7%) with mean age of 57.5 years the time of primary TAA. All patients had undergone primary procedure at a mean duration of 3.5 years previously (range: 3 months to 10 years). Of the total revision procedures, 12/40 (30%) of revisions were carried out due to malalignment, 10/42 (23.8%) due to loosening of the implants or bone subsidence, 5/42 procedures (11.9%) following infection, 4/42 (9.5%) due to polyethylene migration, 1/42 (2.3%) due to fracture and 1/42 (2.3%) due to Charcot arthropathy. In 9/42 (21.4%) cases, imaging showed no objective reason for pain. 50% of patients who underwent revision TAA reported 78.5% satisfaction with results of surgery at 2 years follow up post-operatively. Conclusion. Major reasons for revising primary
Muller-Weiss disease is an uncommon condition with unclear etiology and no gold standard treatment. The question arises; which joints to fuse? Although no consensuses prevail, one must postulate fusion should include those affected. Consequently, to establish an algorithm for its surgical management we set out to study clinical and radiographic features with use of SPECT-CT and a literature review. 57 consecutive feet presenting with Muller-Weiss disease analysed; 15 men, 25 women, age 22–84. Condition bilateral in 17, left side 16, right in 7 patients. Specific history and examination by senior author. Radiographic series and SPECT-CT obtained with surgery performed on significantly symptomatic feet. Measurements of Meary-Tomeno angles, anteroposterior thickness of navicular at the midpoint of each naviculo-cuneiform, alongside the medial extrusion distance and percentage of compression in each case performed. Poor correlation between Meary's angle and 1) degree of compression at naviculo-cuneiform joints, 2) degree of extrusion 3) compression vs extrusion using R. 2. coefficient of determination (invalidating Maceira et al. classification). In unilateral cases, extrusion significantly greater on affected side 94.7% (P< 0.001 Fisher exact test). Degree of extrusion significantly greater in bilateral than unilateral cases (p=0.004 unpaired T test). Valgus hindfoot and Meary's negative most common pattern with no correlation between heel alignment and Meary's R. 2. = 0.003. SPECT-CT useful to determine subtalar involvement in ‘stage 2 disease.’. Following review of cases and published literature we propose the following classification for Muller-Weiss disease with treatment algorithm. 3 Stage delineation; Stage 1 (Normal hindfoot alignment); 1A. Talonavicular disease only - Isolated Talonavicular arthrodesis 1B. Talonavicular + Subtalar; double medial or triple arthrodesis. Stage 2. Talonavicular + Naviculocuneiform; 2A. Adequate bone stock - Talo-naviculo-cuneiform arthrodesis, 2B. Inadequate bone stock +- subtalar disease; Talo-naviculo-cuneiform arthrodesis with tricortical bone graft (Mayich). Stage 3; Asymmetric ankle varus. Pantalar arthrodesis Double/triple/TNC/
Introduction. Total ankle replacement as a valid treatment for end stage ankle arthritis, is gaining popularity and every year there is an increasing number of procedures. With revision rates as high as 21% at 5 years and 43% at 10 years there is a need for understanding and reporting the outcome of revision ankle replacement. Our aim was to study the patient reported outcomes following revision
Background. Total ankle replacements (TARs) are becoming increasingly more common in the treatment of end stage ankle arthritis. As a consequence, more patients are presenting with the complex situation of the failing
Aim. Surgical options for management of a failed ankle arthroplasty are currently limited; typically conversion to fusion is recommended with only a few patients being considered for revision replacement surgery. This paper presents our experience of revision ankle replacements in a cohort of patients with failed primary replacements. Method. A total of 18 revision
Varus ankle osteoarthritis (OA) is typically associated with peritalar instability, which may result in altered subtalar joint position. This study aimed to determine the extent to which total ankle replacement (TAR) in varus ankle OA can restore the subtalar position alignment using 3-dimensional semi-automated measurements on WBCT. Fourteen patients (15 ankles, mean age 61) who underwent
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
An increasingly used treatment for end-stage ankle osteoarthritis is total ankle replacement (TAR). However, implant loosening and subsidence are commonly reported complications, leading to relatively high