This study aimed to investigate the optimal sagittal positioning of the uncemented femoral component in total knee arthroplasty to minimize the risk of aseptic loosening and periprosthetic fracture. Ten different sagittal placements of the femoral component, ranging from -5 mm (causing anterior notch) to +4 mm (causing anterior gap), were analyzed using finite element analysis. Both gait and squat loading conditions were simulated, and Von Mises stress and interface micromotion were evaluated to assess fracture and loosening risk.Aims
Methods
Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.Aims
Methods
Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres. We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up.Aims
Methods
INTRODUCTION. Component impingement in total hip arthroplasty (THA) can cause implant damage or dislocation. Dual mobility (DM) implants are thought to reduce dislocation risk, but impingement on metal acetabular bearings may cause femoral component notching. We studied the prevalence of (and risk factors for) femoral notching with DM across two institutions. METHODS. We identified 37 patients with minimum 1-year radiographic follow-up after primary (19), revision (16), or conversion (2) THA with 3 distinct DM devices between 2012 and 2017. Indications for DM included osteonecrosis, femoral neck fracture, concomitant spinal or neurologic pathology, revision or conversion surgery, and history of prosthetic hip dislocation. Most recent radiographs were reviewed and assessed for notching. Acetabular anteversion and abduction were calculated as per Widmer (2004). Records were reviewed for dislocations and reoperations. RESULTS. 2/37 of cases demonstrated femoral component notching, best seen on Dunn views (available in 7/37 cases).
Reverse shoulder arthroplasty is becoming a frequent treatment of choice for patients with shoulder disorders. Complication rates after reverse shoulder arthroplasty may be three-fold that of conventional total shoulder arthroplasty especially in high risk patient populations and diagnoses like revision arthroplasty, fracture sequelae, and severe glenoid bone loss. Complications include component malposition, stiffness, neurological injury, infection, dislocation or instability, acromial or scapular spine fractures, scapular notching, and loosening of implants. Recognition of preoperative risk factors and appropriate 3D planning are essential in optimizing patient outcome and intraoperative success. Failure of reverse shoulder arthroplasty is a significant challenge requiring appropriate diagnosis of the failure mode. The most common neurological injuries involve the brachial plexus and the axillary nerve due to traction, manipulation of the arm, aberrant retractor placement, or relative lengthening of the arm. Intraoperative fractures are relatively uncommon but include the greater tuberosity, acromion, and glenoid. Tuberosity fracture can be repaired intraoperatively with suture techniques, glenoid fractures may be insignificant rim fractures or jeopardise baseplate fixation and require abandoning RSA until glenoid fracture ORIF heals and then a second stage RSA. Periprosthetic infection after RSA ranges from 1 to 10% and may be higher in revision cases and frequently is Propionibacterium acnes and Staphylococcus epidermidis. Dislocation was one of the most common complications after RSA approximately 5% but with increased surgeon experience and prosthetic design, dislocation rates are approaching 1–2%. An anterosuperior deltoid splitting approach has been associated with increased stability as well as subscapularis repair after RSA. Scapular notching is the most common complication after RSA.
Aims
Patients and Methods
Scapular notching is a frequently observed radiographic phenomenon
in reverse shoulder arthroplasty (RSA), signifying impingement of
components. The purposes of this study were to evaluate the effect
of glenoid component size and glenosphere type on impingement-free
range of movement (ROM) for extension and internal and external
rotation in a virtual RSA model, and to determine the optimal configuration
to reduce the incidence of friction-type scapular notching. Preoperative CT scans obtained in 21 patients (three male, 18
female) with primary osteoarthritis were analyzed using modelling
software. Two concurrent factors were tested for impingement-free
ROM and translation of the centre of rotation: glenosphere diameter
(36 mm Aims
Materials and Methods
The Walch Type C dysplastic glenoid is characterized by excessive
retroversion. This anatomical study describes its morphology. A total of 29 shoulders with a dysplastic glenoid were analyzed.
CT was used to measure retroversion, inclination, height, width,
radius-of-curvature, surface area, depth, subluxation of the humeral
head and the Goutallier classification of fatty infiltration. The
severity of dysplasia and deficiency of the posterior rim of the
glenoid were recorded.Aims
Patients and Methods
Background/Purpose. Total hip arthroplasties (THAs) with ceramic bearings are widely performed in young, active patients and thus, long-term outcome in these population is important. Moreover, clinical implication of noise, in which most studies focused on ‘squeaking’, remains controversial and one of concerns unsolved associated with the use of ceramic bearings. However, there is little literature regarding the long-term outcomes after THAs using these contemporary ceramic bearings in young patients. Therefore, we performed a long-term study with a minimum follow-up of . 1. 5 years after THAs using contemporary ceramic bearings in young patients with osteonecrosis of the femoral head (ONFH) less than fifty. Materials and Methods. Among sixty patients (71 hips) with a mean age of 39.1 years, 7 patients (7 hips) died and 4 patients (4 hips) were lost before 15-year follow-up. The remaining 60 hips were included in this study with an average follow-up period of 16.3 years (range, 15 to 18). All patients underwent cementless THA using a prosthesis of identical design and a 28-mm third-generation alumina head by single surgeon. The clinical evaluations included the modified Harris hip score (HHS), history of dislocation and noise around the hip joint: Noise was classified into squeaking, clicking, grinding and popping and evaluated at each follow-up. Snapping was excluded through physical examination or ultrasonography. Radiographic analysis was performed regarding notching on the neck of femoral component, loosening and osteolysis. Ceramic fracture and survivorship free from revision were also evaluated. Results. The mean Harris hip score improved from 55.3 to 95.5 points (range, 83 to 100) at the time of the final follow-up. Seventeen patients (34.7%) reported noise around the hip joint: “squeaking” in one and “clicking” in 16 patients.
Heterotopic ossification (HO) occurs after arthroplasty, especially
total hip arthroplasty. In this study we describe the incidence,
evolution, morphology and clinical consequences of HO following
reverse shoulder arthroplasty. This is a single-centre retrospective study of 132 consecutive
patients who received a Delta III or Delta Xtend reverse total shoulder
arthroplasty between 2006 and 2013 for the treatment of cuff tear
arthropathy. There were 96 women and 36 men. Their mean age at the
time of surgery was 69 years (49 to 89) and the mean follow-up was
36 months (12 to 84). The incidence, evolution, morphology and clinical
consequences of HO using the Constant-Murley score (CS) were analysed.
A modified Brooker classification of HO of the hip was used.Aims
Patients and Methods
We present the medium-term clinical results of a reverse total
shoulder arthroplasty with a trabecular metal glenoid base plate. We reviewed 125 consecutive primary reverse total shoulder arthroplasties
(RTSA) implanted in 124 patients for rotator cuff arthropathy. There
were 100 women and 24 men in the study group with a mean age of
76 years (58 to 89). The mean follow-up was 32 months (24 to 60).
No patient was lost to follow-up.Aim
Patients and Methods
Despite the high success rates of Reverse Shoulder replacements, complications of instability & scapular notching are a concern. Factors reducing relative motion of implant to underlying bone which include lateral offset to centre of rotation, screw & central peg insertion angle and early osteo-integration are maximized in the Trabecular Metal Reverse total shoulder system. We present clinico-radiological outcomes over 72 months. Analysis of a single surgeon series of 140 Reverse total shoulder replacements in 135 patients was done. Mean age was 72(range 58– 87 yrs); 81 females: 54 males. Indications were Rotator cuff arthropathy {n= 88} (63%); Osteo-arthritis with dysfunctional cuff {n= 22}(15%); post-trauma{n=23} (15%); revision from hemiarthroplasty {n=3} (2.4%) and from surface replacement {n=4} (2.8%) All patients were assessed using pre-operative Constants and Oxford scores and clinical & radiographic reviews with standard X-Rays at 6 weeks, 3, 6,12 months and yearly thereafter. X rays included an AP view in 45 degrees of external rotation and modified axillary view. Inferior Scapular notching using the Nerot-Sirveaux grades and Peg Glenoid Rim Distance were looked into by a consultant musculoskeletal radiologist/ Orthopaedic surgeon/ Senior Fellow (post CCT) or a specialist Trainee (ST4 and above). Pain on the visual analogue scale decreased by 98% (9.1 to 0.8) (p<0.01). Constant score improved by 81.8% (12.4 to 68.1) (p<0.05), Oxford shoulder score by 76.7% (56 to 13) (p<0.05). 95.6% of Humeral stems had no radiolucent lines and 4.4% had < 2mm of lucency. Scapular notching was calculated using Sirveaux grades with Peg scapular base angle distance (PSBA) measurements on PACS with Siemens calibration (grade 1= 4 (2.8%); grade 2 =1; grade 3 =0; grade 4=0). 3.57% showed radiographic signs of scapular notching at 72 months. Range of Peg Glenoid Rim Distance was 1.66 to 2.31 cm. Power analysis showed 65 patients were needed to have an 80% power to detect relation of Peg Glenoid Rim Distance to Scapular notching. A likelihood ratio test from Logistic regression model to check correlation of Peg Glenoid Rim Distance to Scapular notching gave a p value of 0.0005. A likelihood ratio from Logistic regression gave a p value of 0.0004 for Infraglenoid Scapular spurs. Highest incidence of spurring was seen in Reverse Total Shoulder Replacements done for Trauma and lowest in patients who got the procedure for Osteoarthritis. Complications included two glenosphere revisions; two stitch abscesses and two Acromial fractures in patients who had a fall two years after the procedure. Improved surgical outcomes can be attributed to surgical technique and implant characteristics. Trabacular metal promotes early osteointegration which resists shearing action of Deltoid on Glenoid component. This allows early mobilisation. Deltoid split approach preserves integrity of Subscapularis and Acromial osteotomy and lateral clavicle excision improve exposure and prevent Acromion fracture. Positioning the Glenoid component inferiorly on the Glenoid decreases incidence of Scapular notching. Our mid-term validated outcomes are promising with only 3.57% Grade I/II radiographic signs of scapular notching. Long term studies (10 year follow ups) are necessary to confirm its efficacy.
Background. Scapular notching causes glenoid bone loss after a reverse total shoulder arthroplasty (rTSA). The goal of this study was to assess the influence of prosthesis design on notching. Methods. Prospective, single surgeon cohort. Two different rTSA designs were consecutively implanted and compared: 25 Delta III rTSAs and 57 Delta Xtend rTSAs in 80 patients.
The clinical impact of scapular notching is controversial. Some reports suggest it has no impact while others have demonstrated it does negatively impact clinical outcomes. The goal of this clinical study is to analyze the pre- and post-operative outcomes of 415 patients who received rTSA with one specific prosthesis (Equinoxe; Exactech, Inc). 415 patients (mean age: 72.2yrs) with 2 years minimum follow-up were treated with rTSA for CTA, RCT, and OA by 8 fellowship trained orthopaedic surgeons. 363 patients were deemed to not have a scapular notch by the implanting surgeon at latest follow-up (72.1 yrs; 221F/131M) whereas 52 patients were deemed to have a scapular notch at latest follow-up (73.3 yrs; 33F/19M). Outcomes were scored using SST, UCLA, ASES, Constant, and SPADI metrics; active abduction, forward flexion, and internal/external rotation were also measured to quantify function. Average follow-up was 38.1 months (No Notch: 37.2; Notch: 44.4). A two-tailed, unpaired t-test identified differences (p<0.05) in pre-operative, post-operative, and pre-to-post improvements.Introduction
Methods
Introduction. Reverse shoulder prosthesis has been developed to treat the clinical and pathological condition noted as cuff tear arthropathy (CTA). The current models of reverse shoulder arthroplasty (RSA) expose the procedure to the risk of scapular notching, possibly leading to loosening of the glenoid. Aim. The purpose of this study was to report updated results at a minimum follow-up of four years of 25 patients underwent reverse shoulder arthroplasty between 2006 and 2010 with an eccentric 36-mm glenoid component (SMR Lima). Methods. Clinical and radiographic evaluation was performed preoperatively and at every year of follow-up. All patients were evaluated with MRI or CT scan preoperatively and with X-ray examinations postoperatively to evaluate the presence of inferior scapular notching. To value the clinical outcome the Constant score and VAS score have been evaluated preoperatively and every year of follow up. Results. Since the last report the mean Constant remain stable (63.11±8.92) and the mean VAS score decreased to 1.75±2.38 (p<0.005). Radiographs showed one case of grade 1 inferior scapular notching at 7-year follow-up.
Patients with pain and loss of shoulder function
due to nonunion of a fracture of the proximal third of the humerus may
benefit from reverse total shoulder replacement. This paper reports
a prospective, multicentre study, involving three hospitals and
three surgeons, of 35 patients (28 women, seven men) with a mean
age of 69 years (46 to 83) who underwent a reverse total shoulder
replacement for the treatment of nonunion of a fracture of the proximal humerus.
Using Checchia’s classification, nine nonunions were type I, eight
as type II, 12 as type III and six as type IV. The mean follow-up
was 51 months (24 to 99). Post-operatively, the patients had a significant
decrease in pain (p <
0.001), and a significant improvement in
flexion, abduction, external rotation and Constant score (p <
0.001), but not in internal rotation. A total of nine complications
were recorded in seven patients: six dislocations, one glenoid loosening
in a patient who had previously suffered dislocation, one transitory
paresis of the axillary nerve and one infection. Reverse total shoulder replacement may lead to a significant
reduction in pain, improvement in function and a high degree of
satisfaction. However, the rate of complications, particularly dislocation,
was high. Cite this article:
Metal-on-metal (MoM) hip resurfacing was developed in the 1990s by surgeons in Birmingham, UK, as a surgical solution to the problem of osteoarthritis in younger, more active patients. Early results were promising and the procedure gained in popularity. However, adverse reports of soft-tissue reaction and failure started to appear from 2008 onwards. Surgeons may be asked to write medico-legal reports on the surgical aspects of an individual case for claimant lawyers or in defence for the NHSLA or indemnity insurers. The purpose of this article is to cover some of the aspects of the operation that may be considered in such medico-legal reports.
There has been increased focus on understanding the risk factors associated with scapular notching in reverse shoulder arthroplasty (RSA). The purpose of this study is to evaluate the effect of scapular morphology and surgical technique on the occurrence of scapular notching using the notching index as a comprehensive predictive tool. Ninety-one patients treated with a primary RSA were followed for a minimum of 24 months. Using a previously published notching index formula ((PSNA × 0.13) + (PGRD)), a notching index value for all patients was calculated. Radiographic assessment of patients were grouped by Nerot grade of scapular notching, group mean differences for prosthetic scapular neck angle (PSNA), peg glenoid rim distance (PGRD), preoperative scapular neck angle (SNA), notching index and clinical outcomes were compared.Background
Methods
This study provides recommendations on the position
of the implant in reverse shoulder replacement in order to minimise
scapular notching and osteophyte formation. Radiographs from 151
patients who underwent primary reverse shoulder replacement with
a single prosthesis were analysed at a mean follow-up of 28.3 months
(24 to 44) for notching, osteophytes, the position of the glenoid
baseplate, the overhang of the glenosphere, and the prosthesis scapular
neck angle (PSNA). A total of 20 patients (13.2%) had a notch (16 Grade 1 and four
Grade 2) and 47 (31.1%) had an osteophyte. In patients without either
notching or an osteophyte the baseplate was found to be positioned
lower on the glenoid, with greater overhang of the glenosphere and
a lower PSNA than those with notching and an osteophyte. Female patients
had a higher rate of notching than males (13.3% Based on these findings we make recommendations on the placement
of the implant in both male and female patients to avoid notching
and osteophyte formation. Cite this article: