The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 137 - 142
1 Mar 2024
van Veghel MHW van der Koelen RE Hannink G Schreurs BW Rijnen WHC

Aims

The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA).

Methods

We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with femoral revision for any indication, for septic loosening, for fracture of the femoral component and for aseptic loosening as endpoints. The indications for revision were evaluated. Fractures of the femoral component were described in detail.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 144 - 150
1 Feb 2024
Lynch Wong M Robinson M Bryce L Cassidy R Lamb JN Diamond O Beverland D

Aims

The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component.

Methods

We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 17 - 17
23 Jun 2023
Schreurs BW van Veghel MH van der Koelen RE Hannink G Rijnen WH
Full Access

Although data on uncemented short stems are available, studies on cemented short-stemmed THAs are limited. These cemented short stems may have inferior long-term outcomes and higher femoral component fracture rates. Hence, we examined the long-term follow-up of cemented short Exeter stems used in primary THA. Within the Exeter stem range, 7 stems have a stem length of 125 mm or less. These stems are often used in small patients, in young patients with a narrow femoral canal or patients with anatomical abnormalities. Based on our local database, we included 394 consecutive cemented stems used in primary THA (n=333 patients) with a stem length ≤125 mm implanted in our tertiary referral center between 1993 and December 2021. We used the Dutch Arthroplasty Registry (LROI) to complete and cross-check the data. Kaplan-Meier survival analyses were performed to determine 20-year survival rates with stem revision for any reason, for septic loosening, for aseptic loosening and for femoral component fracture as endpoints. The proportion of male patients was 21% (n=83). Median age at surgery was 42 years (interquartile range: 30–55). The main indication for primary THA was childhood hip diseases (51%). The 20-year stem survival rate of the short stem was 85.4% (95% CI: 73.9–92.0) for revision for any reason and 96.2% (95%CI: 90.5–98.5) for revision for septic loosening. No stems were revised for aseptic femoral loosening. However, there were 4 stem fractures at 6.6, 11.6, 16.5 and 18.2 years of follow-up. The stem survival with femoral component fracture as endpoint was 92.7% (CI: 78.5–97.6) at 20 years. Cemented short Exeter stems in primary THA show acceptable survival rates at long-term follow-up. Although femoral component fracture is a rare complication of a cemented short Exeter stem, orthopaedic surgeons should be aware of its incidence and possible risk factors


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1678 - 1685
1 Nov 2021
Abdelaziz H Schröder M Shum Tien C Ibrahim K Gehrke T Salber J Citak M

Aims

One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions.

Methods

In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 7 - 7
1 Aug 2021
Kennedy I Ng N Young D Kane N Marsh A Meek D
Full Access

Cement-in-cement revision of the femoral component represents a widely practiced technique for a variety of indications. In this study we compare the clinical and radiological outcomes of two polished tapered stems. From our prospectively collated database we identified all patients undergoing cement-in-cement revision from January 2005 – 2013 who had a minimum of two years follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-stem AMT high offset No 1. Patients were followed-up annually with clinical and radiological assessment. Ninety-seven patients matched the inclusion criteria. There were 50 Exeter and 47 C-stem AMT components. There were no significant differences between the patient demographics in either group. Mean follow-up was 9.7 years. A significant improvement in OHS, WOMAC and SF-12 scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4mm in this cohort compared to 0mm in the C-stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of femoral component fracture in the Exeter group. Our series shows promising long-term outcomes for the cement-in-cement revision technique using either the Exeter or C-stem AMT components. These results demonstrate that cement-in-cement revision using a double or triple taper-slip design is a safe and reliable technique when used for the correct indications


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy IW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims. Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. Methods. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment. Results. A total of 97 patients matched the inclusion criteria (50 Exeter and 47 C-Stem AMT components). There were no significant differences between the patient demographic data in either group. Mean follow-up was 9.7 years. A significant improvement in Oxford Hip Score (OHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 12-item Short-Form Survey (SF-12) scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4 mm in this cohort compared with 0 mm in the C-Stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-Stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of femoral component fracture in the Exeter group. Conclusion. Our series shows promising mid-term outcomes for the cement-in-cement revision technique using either the Exeter or C-Stem AMT components. These results demonstrate that cement-in-cement revision using a double or triple taper-slip design is a safe and reliable technique when used for the correct indications. Cite this article: Bone Joint J 2021;103-B(7):1215–1221


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes.

Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty.

This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice.

Cite this article: Bone Jt Open 2021;2(1):48–57.


The Bone & Joint Journal
Vol. 102-B, Issue 5 | Pages 573 - 579
1 May 2020
Krueger DR Guenther K Deml MC Perka C

Aims

We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed.

Methods

All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 293 - 300
1 Mar 2020
Zheng H Gu H Shao H Huang Y Yang D Tang H Zhou Y

Aims

Vancouver type B periprosthetic femoral fractures (PFF) are challenging complications after total hip arthroplasty (THA), and some treatment controversies remain. The objectives of this study were: to evaluate the short-to-mid-term clinical outcomes after treatment of Vancouver type B PFF and to compare postoperative outcome in subgroups according to classifications and treatments; to report the clinical outcomes after conservative treatment; and to identify risk factors for postoperative complications in Vancouver type B PFF.

Methods

A total of 97 consecutive PPFs (49 males and 48 females) were included with a mean age of 66 years (standard deviation (SD) 14.9). Of these, 86 patients were treated with surgery and 11 were treated conservatively. All living patients had a minimum two-year follow-up. Patient demographics details, fracture healing, functional scores, and complications were assessed. Clinical outcomes between internal fixation and revisions in patients with or without a stable femoral component were compared. Conservatively treated PPFs were evaluated in terms of mortality and healing status. A logistic regression analysis was performed to identify risk factors for complications.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1447 - 1458
1 Nov 2019
Chatziagorou G Lindahl H Kärrholm J

Aims

We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis.

Patients and Methods

The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1618 - 1625
1 Dec 2018
Gill JR Kiliyanpilakkill B Parker MJ

Aims

This study describes and compares the operative management and outcomes in a consecutive case series of patients with dislocated hemiarthroplasties of the hip, and compares outcomes with those of patients not sustaining a dislocation.

Patients and Methods

Of 3326 consecutive patients treated with hemiarthroplasty for fractured neck of femur, 46 (1.4%) sustained dislocations. Of the 46 dislocations, there were 37 female patients (80.4%) and nine male patients (19.6%) with a mean age of 83.8 years (66 to 100). Operative intervention for each, and subsequent dislocations, were recorded. The following outcome measures were recorded: dislocation; mortality up to one-year post-injury; additional surgery; residential status; mobility; and pain score at one year.


Bone & Joint 360
Vol. 6, Issue 3 | Pages 41 - 43
1 Jun 2017
Foy MA


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 754 - 760
1 Jun 2016
Malek IA Royce G Bhatti SU Whittaker JP Phillips SP Wilson IRB Wootton JR Starks I

Aims

We assessed the difference in hospital based and early clinical outcomes between the direct anterior approach and the posterior approach in patients who undergo total hip arthroplasty (THA).

Patients and Methods

The outcome was assessed in 448 (203 males, 245 females) consecutive patients undergoing unilateral primary THA after the implementation of an ‘Enhanced Recovery’ pathway. In all, 265 patients (mean age: 71 years (49 to 89); 117 males and 148 females) had surgery using the direct anterior approach (DAA) and 183 patients (mean age: 70 years (26 to 100); 86 males and 97 females) using a posterior approach. The groups were compared for age, gender, American Society of Anesthesiologists grade, body mass index, the side of the operation, pre-operative Oxford Hip Score (OHS) and attendance at ‘Joint school’. Mean follow-up was 18.1 months (one to 50).


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 461 - 467
1 Apr 2016
Abdel MP Watts CD Houdek MT Lewallen DG Berry DJ

Aim and Methods

The goals of this study were to define the risk factors, nature, chronology, and treatment strategies adopted for periprosthetic femoral fractures in 32 644 primary total hip arthroplasties (THAs).

Results

There were 564 intra-operative fractures (1.7%); 529 during uncemented stem placement (3.0%) and 35 during cemented stem placement (0.23%). Intra-operative fractures were more common in females and patients over 65 years (p < 0.001). The majority occurred during placement of the femoral component (60%), and involved the calcar (69%). There were 557 post-operative fractures (20-year probability: 3.5%; 95% confidence interval (CI) 3.2 to 3.9); 335 fractures after placement of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to 9.1) and 222 after placement of a cemented stem (20-year probability: 2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture within 30 days after an uncemented stem was ten times higher than a cemented stem. The most common post-operative fracture type was a Vancouver AG (32%; n = 135), with 67% occurring after a fall. In all, 36% (n = 152) were treated with revision arthroplasty.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 468 - 474
1 Apr 2016
Abdel MP Houdek MT Watts CD Lewallen DG Berry DJ

Aims

The goals of this study were to define the risk factors, characteristics, and chronology of fractures in 5417 revision total hip arthroplasties (THAs).

Patients and Methods

From our hospital’s prospectively collected database we identified all patients who had undergone a revision THA between 1969 and 2011 which involved the femoral stem. The patients’ medical records and radiographs were examined and the relevant data extracted. Post-operative periprosthetic fractures were classified using the Vancouver system. A total of 5417 revision THAs were identified.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 307 - 312
1 Mar 2016
Maggs JL Smeatham A Whitehouse SL Charity J Timperley AJ Gie GA

Aims

We report on the outcome of the Exeter Contemporary flanged cemented all-polyethylene acetabular component with a mean follow-up of 12 years (10 to 13.9). This study reviewed 203 hips in 194 patients. 129 hips in 122 patients are still in situ; 66 hips in 64 patients were in patients who died before ten years, and eight hips (eight patients) were revised. Clinical outcome scores were available for 108 hips (104 patients) and radiographs for 103 hips (100 patients).

Patients and Methods

A retrospective review was undertaken of a consecutive series of 203 routine primary cemented total hip arthroplasties (THA) in 194 patients.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1608 - 1614
1 Dec 2015
Tice A Kim P Dinh† L Ryu JJ Beaulé PE

The primary purpose of this study of metal-on-metal (MoM) hip resurfacing was to compare the effect of using a cementless or cemented femoral component on the subsequent bone mineral density (BMD) of the femoral neck.

This was a single-centre, prospective, double-blinded control trial which randomised 120 patients (105 men and 15 women) with a mean age of 49.4 years (21 to 68) to receive either a cemented or cementless femoral component. Follow-up was to two years. Outcome measures included total and six-point region-of-interest BMD of the femoral neck, radiological measurements of acetabular inclination, neck-shaft and stem-shaft angles, and functional outcome scores including the Harris hip score, the Western Ontario and McMaster Universities Osteoarthritis Index and the University of California at Los Angeles activity scale.

In total, 17 patients were lost to follow-up leaving 103 patients at two years. There were no revisions in the cementless group and three revisions (5%) in the cemented group (two because of hip pain and one for pseudotumour).

The total BMD was significantly higher in the cementless group at six months (p < 0.001) and one year (p = 0.01) than in the cemented group, although there was a loss of statistical significance in the difference at two years (p = 0.155).

All patient outcomes improved significantly: there were no significant differences between the two groups.

The results show better preservation of femoral neck BMD with a cementless femoral component after two years of follow-up. Further investigation is needed to establish whether this translates into improved survivorship.

Cite this article: Bone Joint J 2015;97-B:1608–14.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1192 - 1196
1 Sep 2015
Amanatullah DF Siman H Pallante GD Haber DB Sierra RJ Trousdale RT

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up.

The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (sd) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, sd 3.1). The mean diameter of the femoral component increased from 12.7 mm (sd 1.9) to 16.2 mm (sd 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion.

A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; sd 18.3) and after trephine removal and revision THA (81.2; sd 14.8, p = 0.918).

These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively.

Cite this article: Bone Joint J 2015;97-B:1192–6.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIV | Pages 8 - 8
1 Jul 2012
Sarraf K Abdul-Jabar H Wharton R Shah G Singer G
Full Access

Femoral component fracture is a rarely reported but devastating complication of total knee arthroplasty. It has occurred most frequently with Whiteside Ortholoc II replacements uncemented knee replacements. Presentation may be with acute pain, progressive pain or returning deformity. It occurs more commonly in the medial condyle of the femoral component. It is rarely seen in cemented replacements. All currently available literature describing fractures of condylar replacements, both cemented and uncemented. Predisposing factors include varus deformity either pre or post operatively. The mechanism of failure is thought to be failure of the infiltration of bone into the replacement. This is often due to polyethylene wear or metallosis causing abnormal tissue reaction with or without osteolysis. We present the case of a fractured Press Fit Condylar (PFC) cemented implant (DePuy, Johnson&Johnson, Raynham, Massachusettes, USA) affecting the medial condyle. To our knowledge this is only the third reported case of fracture in a PFC implant, and the first in a cemented PFC implant. Our patient was a 64 year old male who presented with unresolving knee pain post total knee arthroplasty, caused by fatigue fracture of the medial condyle of the femoral component. This was identified as loosening on plain radiographs and replaced with a revision prosthesis with a good post operative result. Given our aging population and with the increase of joint arthroplasty, this case sheds light on a potentially under recognised and increasingly important cause of knee pain following arthroplasty


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 5 - 5
1 Mar 2010
McIvor GWD Johnston GH
Full Access

Purpose: To compare the early results and complications associated with two different minimally invasive surgical techniques, the mini-posterior and G3 approaches. Method: The senior author prospectively allocated successive patients into either posterior or G3 groups for all primary total hip replacements from April 2005 to July 2006. Results: Of 173 hips, 116 hips were introduced via a mini-posterior approach, 57 via a G3. There were no significant differences in composition of the two groups in average age, gender, length of hospital stay, or early functional outcome. The posterior group patients were generally heavier, but OR time was significantly less than in the G3 group, as were blood transfusion rates. Technical issues were noted in 25% of cases, including calcar and periprosthetic fracture (9), malposition of the acetabular and femoral components (6), and leg length discrepancies (28). There was one death and one infection. There were 3 dislocations, from both groups, deemed to be more a function of improper component implantation, rather than of the surgical approach. Of 4 with neurologic symptoms, one patient from each group experienced anterior thigh numbness with activity; in the G3 group there was a lateral femoral cutaneous nerve deficit, and a severe sciatic nerve deficit. There were 7 (4%) patients who required additional surgery, 5 for periprosthetic fracture or implant malposition. Two re-operations were attributed to being likely specific to the surgical approach and its limited visualization, both G3s with femoral component malposition. Technically it was appreciated that there was a fine line between a too tight femoral component (heightened fracture risk) and one too loose (heightened subsidence risk). The strengths of this study are that it reflects the experience of one career arthroplasty surgeon performing a consecutive, nearly randomized to two MIS approaches, consecutive case series of primary total hip arthroplasties, using the same fiber-metal acetabulum and tapered fiber-metal HA coated femur prostheses in all cases. Conclusion: The G3 approach is technically difficult, with a long “learning curve”. This difficulty was acutely accentuated in larger patients. The mini-posterior incision was relatively easy, even in the larger patient, and became the preferred approach for the more challenging case