We reviewed all patients who sustained a fracture of the hip
and were treated in Northern Ireland over a period of 15 years to
identify trends in incidence, the demographics of the patients,
the rates of mortality, the configuration of the fracture and the
choice of implant. Since 01 January 2001 data about every fracture of the hip sustained
in an adult have been collected centrally in Northern Ireland. All
adults with such a fracture between 2000 and 2015 were included
in the study. Temporal changes in their demographics, the mode of
treatment, and outcomes including mortality were analysed.Aims
Patients and Methods
The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of proximal fixation was recorded from
intra-operative fluoroscopic images, and technical inadequacy for
each point was defined. All patients were followed to final follow-up
and mechanical failures were identified. A multivariate statistical
analysis was performed, adjusting for confounders. A total of 16
failures (7.2%) were identified. The position of the lag screw relative
to the lateral cortex was the most important point of proximal fixation,
and when inadequate the failure rate was 25.8% (eight of 31: odds
ratio 7.5 (95% confidence interval 2.5 to 22.7), p <
0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral
fractures is rare (<
1%) when three-point proximal fixation is
achieved. However, when proximal fixation is inadequate, failure
rates increase. The strongest predictor of failure is positioning
the lateral end of the lag screw short of the lateral cortex. Adherence
to simple technical points minimises the risk of fixation failure
in this vulnerable patient group. Cite this article:
Purpose of study. The addition of interlocking screws to intramedullary nails adds greatly to the stability of these constructs, yet the placement of distal screws accounts for a significant proportion of the total fluoroscopy and operative times. The Sureshot® (Smith and Nephew™) is a computerised system that allows placement of distal screws without fluoroscopy by using electromagnetic guided imagery. The purpose of this study is to compare traditional free-hand technique to the Sureshot® technique in terms of operating time, radiation dose and accuracy. Methods. Between September 2011 and March 2012 we prospectively randomised 66 consecutive patients presenting to us with femur shaft fractures requiring intramedullary nails to either free-hand (n=33) or Sureshot® assisted (n=28) distal locking. Fractures warranting only one distal locking screw, or those requiring retrograde or
Introduction. Corrective femoral osteotomy in adults, as a closed procedure with the use of an intramedullary saw, is an elegant, minimally invasive technique for the correction of lower limb length inequalities or problems of torsion. Stabilisation following the osteotomy was achieved with a
Introduction Patients sustaining fractures of the proximal femur, with co-morbid medical problems, have increased rates of morbidity and mortality. Chronic renal failure is one such co-morbidity. This study examines the outcome in patients with chronic renal failure who sustain fractures of the proximal femur. Patients and Methods All patients with a fractured neck of femur who presented to our department from September 1997 to March 2004 were retrospectively reviewed. Eighteen of these patients were found to have chronic renal failure requiring dialysis. Medical records were reviewed and information was collected and analysed. A full literature review was conducted. Results There were nine intra-capsular fractures and nine extra-capsular fractures. Four patients with intra-capsular fractures were treated by internal fixation and four by arthroplasty. One patient with an intra-capsular fracture was treated non-operatively. Eight patients with extra-capsular fractures were treated with a dynamic hip screw device and one was treated using a