Advertisement for orthosearch.org.uk
Results 1 - 20 of 2394
Results per page:

Aims

For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis.

Methods

We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1206 - 1215
1 Nov 2024
Fontalis A Buchalter D Mancino F Shen T Sculco PK Mayman D Haddad FS Vigdorchik J

Understanding spinopelvic mechanics is important for the success of total hip arthroplasty (THA). Despite significant advancements in appreciating spinopelvic balance, numerous challenges remain. It is crucial to recognize the individual variability and postoperative changes in spinopelvic parameters and their consequential impact on prosthetic component positioning to mitigate the risk of dislocation and enhance postoperative outcomes. This review describes the integration of advanced diagnostic approaches, enhanced technology, implant considerations, and surgical planning, all tailored to the unique anatomy and biomechanics of each patient. It underscores the importance of accurately predicting postoperative spinopelvic mechanics, selecting suitable imaging techniques, establishing a consistent nomenclature for spinopelvic stiffness, and considering implant-specific strategies. Furthermore, it highlights the potential of artificial intelligence to personalize care.

Cite this article: Bone Joint J 2024;106-B(11):1206–1215.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 832 - 836
4 Oct 2024
Kayani B Mancino F Baawa-Ameyaw J Roussot MA Haddad FS

Aims

The outcomes of patients with unexpected positive cultures (UPCs) during revision total hip arthroplasty (THA) and total knee arthroplasty (TKA) remain unknown. The objectives of this study were to establish the prevalence and infection-free implant survival in UPCs during presumed aseptic single-stage revision THA and TKA at mid-term follow-up.

Methods

This study included 297 patients undergoing presumed aseptic single-stage revision THA or TKA at a single treatment centre. All patients with at least three UPCs obtained during revision surgery were treated with minimum three months of oral antibiotics following revision surgery. The prevalence of UPCs and causative microorganisms, the recurrence of periprosthetic joint infections (PJIs), and the infection-free implant survival were established at minimum five years’ follow-up (5.1 to 12.3).


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims. Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up. Methods. This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up. Results. Clinical assessment showed persistent major improvement in all scores. On radiological assessment, only one case showed a lucent line (without symptoms). At last follow-up, wear and migration were below the critical thresholds. No cup-related revisions were needed, indicating an outstanding survival rate of 100%. Conclusion. Isoelastic VEPE cups offer high success rates and may prevent osteolysis, aseptic loosening, and the need for revision surgeries in the long term. However, longer follow-up is needed to validate our findings and confirm the advantages offered by this cup. Cite this article: Bone Jt Open 2024;5(10):825–831


Bone & Joint 360
Vol. 13, Issue 5 | Pages 21 - 23
1 Oct 2024

The October 2024 Hip & Pelvis Roundup360 looks at: Does the primary surgical approach matter when choosing the approach for revision total hip arthroplasty?; Time to achieve the minimal clinically important difference in primary total hip arthroplasty: comparison of anterior and posterior surgical approaches; To scope or not to scope: arthroscopy as an adjunct to PAO does not provide better clinical outcomes at one year than PAO alone; Re-exploring horizons in hip resurfacing: two-year results of a ceramic-on-ceramic hip resurfacing; Association between tranexamic acid and decreased periprosthetic joint infection risk in patients undergoing total hip and knee arthroplasty; Octogenarians fare well: in revision for infection age is not a bar.


Bone & Joint 360
Vol. 13, Issue 5 | Pages 42 - 44
1 Oct 2024

The October 2024 Oncology Roundup360 looks at: Composite reconstruction: is it the answer for pelvic resections?; Can the cartilaginous thickness determine the risk of malignancy in pelvic cartilaginous tumours, and how accurate is the preoperative biopsy of these tumours?; Incidence and survival outcomes of patients with high-grade appendicular bone sarcoma and isolated regional lymph node metastasis; Improved metastatic-free survival after systematic re-excision following complete macroscopic unplanned excision of limb or trunk soft-tissue sarcoma; UK guidelines for the management of soft-tissue sarcomas; Current management of desmoid tumours: a review.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1021 - 1030
1 Sep 2024
Oto J Herranz R Fuertes M Plana E Verger P Baixauli F Amaya JV Medina P

Aims

Bacterial infection activates neutrophils to release neutrophil extracellular traps (NETs) in bacterial biofilms of periprosthetic joint infections (PJIs). The aim of this study was to evaluate the increase in NET activation and release (NETosis) and haemostasis markers in the plasma of patients with PJI, to evaluate whether such plasma induces the activation of neutrophils, to ascertain whether increased NETosis is also mediated by reduced DNaseI activity, to explore novel therapeutic interventions for NETosis in PJI in vitro, and to evaluate the potential diagnostic use of these markers.

Methods

We prospectively recruited 107 patients in the preoperative period of prosthetic surgery, 71 with a suspicion of PJI and 36 who underwent arthroplasty for non-septic indications as controls, and obtained citrated plasma. PJI was confirmed in 50 patients. We measured NET markers, inflammation markers, DNaseI activity, haemostatic markers, and the thrombin generation test (TGT). We analyzed the ability of plasma from confirmed PJI and controls to induce NETosis and to degrade in vitro-generated NETs, and explored the therapeutic restoration of the impairment to degrade NETs of PJI plasma with recombinant human DNaseI. Finally, we assessed the contribution of these markers to the diagnosis of PJI.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 898 - 906
1 Sep 2024
Kayani B Wazir MUK Mancino F Plastow R Haddad FS

Aims. The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. Methods. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed. Results. The BOne trauma and Soft-Tissue Injury classification system in total Hip arthroplasty (BOSTI Hip) grades osseous acetabular trauma and periarticular muscle damage during THA. The classification system has an interclass correlation coefficient of 0.90 (95% CI 0.86 to 0.93) for interobserver agreement and 0.89 (95% CI 0.84 to 0.93) for intraobserver agreement. RO THA was associated with improved BOSTI Hip scores (p = 0.002) and more pristine osseous surfaces in the anterior superior (p = 0.001) and posterior superior (p < 0.001) acetabular quadrants compared with CO THA. There were no differences between the groups in relation to injury to the gluteus medius (p = 0.084), obturator internus (p = 0.241), piriformis (p = 0.081), superior gamellus (p = 0.116), inferior gamellus (p = 0.132), quadratus femoris (p = 0.208), and vastus lateralis (p = 0.135), but overall combined muscle injury was reduced in RO THA compared with CO THA (p = 0.023). Discussion. The proposed BOSTI Hip classification provides a reproducible grading system for stratifying iatrogenic bone trauma and soft-tissue injury during THA. RO THA was associated with improved BOSTI Hip scores, more pristine osseous acetabular surfaces, and reduced combined periarticular muscle injury compared with CO THA. Further research is required to understand if these intraoperative findings translate to differences in clinical outcomes between the treatment groups. Cite this article: Bone Joint J 2024;106-B(9):898–906


Bone & Joint Open
Vol. 5, Issue 8 | Pages 715 - 720
23 Aug 2024
Shen TS Cheng R Chiu Y McLawhorn AS Figgie MP Westrich GH

Aims

Implant waste during total hip arthroplasty (THA) represents a significant cost to the USA healthcare system. While studies have explored methods to improve THA cost-effectiveness, the literature comparing the proportions of implant waste by intraoperative technology used during THA is limited. The aims of this study were to: 1) examine whether the use of enabling technologies during THA results in a smaller proportion of wasted implants compared to navigation-guided and conventional manual THA; 2) determine the proportion of wasted implants by implant type; and 3) examine the effects of surgeon experience on rates of implant waste by technology used.

Methods

We identified 104,420 implants either implanted or wasted during 18,329 primary THAs performed on 16,724 patients between January 2018 and June 2022 at our institution. THAs were separated by technology used: robotic-assisted (n = 4,171), imageless navigation (n = 6,887), and manual (n = 7,721). The primary outcome of interest was the rate of implant waste during primary THA.


Highly cross-linked polyethylene (HXLPE) has decreased wear and revision rates in total hip replacement (THR) at a long-term. However, the effect of HXLPE manufacturing characteristics on femoral head penetration has not been clearly defined yet. We report this single-institution study to investigate the clinical and radiological results of different HXLPE liners in THR. In this retrospective cohort analysis of our prospective database, we identified 904 THRs performed between 2000 and 2013. Seven different HXLPE liner types were assessed: remelted (3), annealed (2), sequentially annealed (1) and vitamin E-infused (1). The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Roman Software v1.70 package. Thirty hips were revised for the following reasons: aseptic loosening (11), dislocation (12), periprosthetic femoral fracture (6), and infection (1). No hip was revised for wear or osteolysis. 741 THRs were evaluated for a mean follow-up of 15 years (range, 10 to 20). The mean total penetration 15 years after THR was 0.17 mm in the sequentially annealed, and 0.16 mm in the vitamin E-infused groups, whereas it was 0.26 mm in the melted 95 kGy, 0.27 mm in the melted 5 Mrad and 0.25 mm in the melted 100 kGy groups (p=0.001). From one to 15 years after surgery, the mean wear was 0.206 mm in cups with an acetabular inclination greater than 50º and 0.105 mm in those placed between 35º and 50º (p<0.001). Although HXLPE characteristics can result in a different wear performance in patients undergoing THR at a mean follow-up of 15 years, the clinical results are excellent. The position of the acetabular component can influence on the femoral head penetration of modern HXLPEs. Nevertheless, these patients should continue to be monitored to detect future problems


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 13 - 13
19 Aug 2024
DeBenedetti A Weintraub MT Valle CJD Jacobs JJ Nam D
Full Access

The purpose of this randomized controlled trial was to evaluate serum metal levels in patients undergoing total hip arthroplasty (THA) with a conventional compared to a modular dual-mobility bearing. Patients undergoing primary THA for osteoarthritis were randomized to receive either a modular dual-mobility or conventional polyethylene bearing. All patients received the same titanium acetabular and femoral component and a ceramic femoral head. Serum metal levels were drawn pre-operatively then annually for a minimum of two years postoperatively. An a priori power analysis determined that 40 patients (20 per cohort) were needed to identify a clinically relevant difference in serum cobalt of 0.35 ng/ml (ppb) at 90% power. Forty-six patients were randomized to a modular dual-mobility (n=25) or conventional bearing (n=21) with 40 at a minimum follow-up of two years. No differences in serum cobalt (mean 0.14 ppb [range, 0.075–0.29] vs. 0.20 ppb [range, 0.075–0.57], p=0.39) or chromium levels (mean 0.14 ppb [range, 0.05–0.50] vs. 0.12 ppb [range, 0.05–0.35], p=0.65) were identified between the modular dual-mobility and conventional cohorts, respectively. There was no statistically significant difference in serum Co or Cr at two years postoperatively in subjects implanted with a ceramic head and this particular dual mobility bearing in comparison to a ceramic head and a conventional acetabular component. While modest expected elevations in serum Co and Cr were observed in the dual mobility group, in no case did the Co level exceed the laboratory reference range nor the threshold of one part per billion that has been associated with adverse local tissue reactions to mechanically-assisted crevice corrosion


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 65 - 65
19 Aug 2024
Walter W Lin D Weinrauch P de Smet K Beaule P Young D Xu J Manktelow A
Full Access

Hip resurfacing arthroplasty (HRA) is a bone conserving alternative to total hip arthroplasty. We present the early 2-year clinical and radiographic follow-up of a novel ceramic-on-ceramic (CoC) HRA in an international multi-centric cohort. Patients undergoing HRA between September 2018 and January 2021 were prospectively included. Patient-reported outcome measures (PROMS) in the form of the Forgotten Joint Score (FJS), HOOS Jr, WOMAC, Oxford Hip Score (OHS) and UCLA Activity Score were collected preoperatively and at 1- and 2-years post-operation. Serial radiographs were assessed for migration, component alignment, evidence of osteolysis/loosening and heterotopic ossification formation. 200 patients were identified to have reached 2-year follow-up. Of these, 185 completed PROMS follow-up at 2 years. There was significant improvement in HOOS (p< 0.001) and OHS (p< 0.001) and FJS (p< 0.001) between the pre-operative and 2-year outcomes. Patients reported improved pain (p<0.001), function (p<0.001) and reduced stiffness (p<0.001) as measured by the WOMAC score. Patients had improved activity scores on the UCLA Active Score (P<0.001) with 53% reporting return to impact activity at 2 years. There was no osteolysis and the mean acetabular cup inclination angle was 41deg and the femoral component shaft angle was 137deg. No fractures were reported over but there was one sciatic nerve palsy with partial recovery. Two patients were revised; one at 3 months for pain due to a misdiagnosed back problem and another at 33 months for loosening of the acetabular component with delamination of the titanium ingrowth surface. CoC resurfacing at 2-years post-operation demonstrate promising results with satisfactory PROMS


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 20 - 20
19 Aug 2024
Tikhilov RM Bilyk SS Dzhavadov AA Shubnyakov II
Full Access

Literature data show varying rates of aseptic loosening of standard hemispherical cups after primary total hip arthroplasty (THA) in patients with Crowe IV developmental dysplasia of the hip (DDH). In order to improve the results regarding the frequency of aseptic loosening of the acetabular component, we decided to use custom-made acetabular implants (CMAI) in this category of patients. The aim of our study was to report outcomes after primary THA using CMAI in patients with Crowe IV DDH. We retrospectively analyzed the results of primary THA using the CMAI in 58 hips. The mean follow-up period was 5.2 years (range 4.2 – 6.3). Patients demonstrated improved functional outcomes according to the Oxford Hip Score: preoperatively 16.9 and postoperatively 38.5 (P < 0.05). Revisions were performed in 2 cases due to dislocations, in 1 case due to nonunion of the greater trochanter and in 1 case due to chronic pain syndrome caused by contact of the CMAI flange with the iliopsoas muscle. In a patient with chronic pain syndrome, during the revision, the pubic flange of the CMAI, which was in contact with the iliopsoas muscle, was removed. There were no radiological signs of loosening of the CMAI. The use of CMAI shows good results in patients with DDH. Extended capability for supplementary screw fixation is an excellent feature of such implants, allowing them to obtain reliable primary fixation. Further observation is required to assess CMAI in the long-term follow-up


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 53 - 53
19 Aug 2024
Hellemondt van GG Colo E Faraj SSA Hooff van M Spruit M
Full Access

The objective of this study was to assess the clinical and radiological results of patients who were revised using a new generation custom-made triflange acetabular component (CTAC) for component loosening and large acetabular defect (Paprosky 3A and 3B) after previous total hip arthroplasty (THA). New generation CTACs involve the use of patient-specific drill guides and incorporate three-dimensional printed bone models, enhancing precision during surgical implantation. Data were extracted from a single centre prospective database of patients with large acetabular defects who were treated with a new generation CTAC. Patients were included if they had a minimum follow-up of five years. The modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at two- and five-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance (p<0.05). A total of 49 (70%) of 70 patients with a mean age of 73.5 years (SD 7.7) had a complete follow-up of 5 years. A significant improvement was found in HOOS, mOHS, EQ-5D-3L utility and NRS, VAS pain rest and activity between baseline and final follow-up. Complications included 8 cases with loosening screws, 4 with bony fractures, 4 periprosthetic infections and 2 cases with dislocation. One patient with bilateral pelvic discontinuity had revision surgery due to recurrent dislocations. No revision surgery was performed for screw failure or implant breakage. New generation CTAC in patients with THA acetabular loosening and massive acetabular bone loss (Paprosky 3A and 3B) can result in stable constructs and significant improvement in functioning and health-related quality of life at five years’ follow-up


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 61 - 61
19 Aug 2024
Whitehouse MR Patel R French J Beswick A Navvuga P Marques E Blom A Lenguerrand E
Full Access

We investigated the revision rates of primary total hip replacement (THR) reported in the National Joint Registry (NJR) by types of bearing surfaces used. We analysed THR procedures across all orthopaedic units in England and Wales. Our analyses estimated all-cause and cause-specific revision rates. We identified primary THRs with heads and monobloc cups or modular acetabular component THRs with head and shell/liner combinations. We used flexible parametric survival models to estimate adjusted hazard ratios (HR). A total of 1,026,481 primary THRs performed between 2003–2019 are included in the primary analysis (Monobloc: n=378,979 and Modular: n=647,502) with 20,869 (2%) of these primary THRs subsequently undergoing a revision episode (Monobloc: n=7,381 and Modular: n=13,488). Compared to implants with a cobalt chrome head and highly crosslinked polyethylene (HCLPE) cup, the all-cause risk of revision for monobloc acetabular implant was higher for patients with cobalt chrome or stainless steel head and non-HCLPE cup. The risk of revision was lower for patients with a delta ceramic head and HCLPE cup implant, at any post-operative period. Compared to patients with a cobalt chrome head and HCLPE liner primary THR, the all-cause risk of revision for modular acetabular implant varied non-constantly. THRs with a delta ceramic or oxidised zirconium head and HCLPE liner had a lower risk of revision throughout the entire post-operative period. The all-cause and indication-specific risk of prosthesis revision, at different time points following the initial implantation, is lower for implants with a delta ceramic or oxidised zirconium head and a HCLPE liner/cup than commonly used alternatives such as cobalt chrome heads and HCLPE liner/cup


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 59 - 59
19 Aug 2024
Bakircioglu S Bulut MA Oral M Caglar O Atilla B Tokgozoglu AM
Full Access

Extensive and severe bone loss of the femur may be a result of a failed total hip arthroplasty (THA) or total knee arthroplasty (TKA) with multiple revision surgeries which may be caused by factors such as infection, periprosthetic fracture or osteolysis. The aim of this study was to assess outcomes of using the “Push-Through Total Femoral Prosthesis” (PTTF) for revision of a total hip replacement with extreme bone loss. Fourteen patients who had extensive bone defects of the femur due to failed THR's and were treated with PTTF between 2012 and 2020 were included in this study. Primary functional outcomes were assessed using Harris Hip Score (HHS), Toronto Extremity Salvage Score (TESS) and Musculoskeletal Tumor Society (MSTS) scores. Range of motion, complications, and ambulatory status were also recorded to assess secondary outcomes. Two of 14 patients underwent surgery with PTTF for both knee and hip arthroplasty revision. The mean time between index surgery and PTTF was 15 years (3 to 32 yrs.). Acetabular components were revised in six of 14 patients. After a mean follow-up of 5.9 years, hip dislocations occurred in 3 patients. All dislocated hips were in patients with retained non-constrained acetabular bearings. Patient satisfaction was high (MSTS: 67%, HHS: 61.2%, TESS 64.6%) despite a high re-operation rate and minor postoperative problems. PTTF is a unique alternative that may be considered for a failed THA revision procedure in patients with an extreme femoral bone defect. Patients are able to ambulate pain free relatively well. Routine usage of constrained liners should be considered to avoid hip dislocation which was our main problem following the procedure


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 54 - 54
19 Aug 2024
AlFayyadh F Neufeld ME Howard LC Masri BA Greidanus NV Garbuz D
Full Access

There remains concern with the use of constrained liners (CL) implanted at the time of acetabular cup revision in revision total hip replacement (rTHA). The aim of this study was to determine the implant survival in rTHA when a CL was implanted at the same time as acetabular cup revision. We reviewed our institutional database to identify all consecutive rTHAs where a CL was implanted simultaneously at the time acetabular cup revision from 2001 to 2021. One-hundred and seventy-four revisions (173 patients) were included in the study. Mean follow-up of 8.7 years (range two – 21.7). The most common indications for rTHA were instability (35%), second-stage periprosthetic joint infection (26.4%), and aseptic loosening (17.2%). Kaplan Meier Analysis was used to determine survival with all-cause re-revision and revision for cup aseptic loosening (fixation failure) as the endpoints. A total of 32 (18.3%) patients underwent re-revision at a mean time of 2.9 years (range 0.1 – 14.1). The most common reasons for re-revision were instability (14), periprosthetic joint infection (seven), and loosening of the femoral component (four). Three (1.7%) required re-revision due to aseptic loosening of the acetabular component (fixation failure) at a mean of two years (0.1 – 5.1). Acetabular component survival free from re-revision due to aseptic loosening was 98.9% (95% CI 97.3 – 100) at five-years and 98.1% (95% CI 95.8 – 100) at 10-years. There were no acetabular component fixation failures in modern highly porous shells. CLs implanted at the time acetabular cup revision in rTHA have a 98.1% 10-year survival free from acetabular cup aseptic loosening (fixation failure). There were no cup fixation failures in modern highly porous shells. Thus, when necessary, implanting a CL during revision of an acetabular component with stable screw fixation is safe with an extremely low risk of cup fixation failure


The Paprosky acetabular bone defect classification system and related algorithms for acetabular reconstruction cannot properly guide cementless acetabular reconstruction in the presence of porous metal augments. We aimed to introduce a rim, points, and column (RPC)-oriented cementless acetabular reconstruction algorithm and its clinical and radiographic outcomes. A total of 123 patients (128 hips) were enrolled. A minimum 5-year radiographic follow-up was available for 96 (75.8%) hips. The mean clinical and radiographic follow-up durations were 6.8±0.9 (range: 5.2–9.2) and 6.3±1.9 (range: 5.0–9.2) years, respectively. Harris hip score (HHS) improved significantly from 35.39±9.91 preoperatively to 85.98±12.81 postoperatively (P<0.001). Among the fixation modes, 42 (32.8%) hips were reconstructed with rim fixation, 42 (32.8%) with three-point fixation without point reconstruction, 40 (31.3%) with three-point fixation combined with point reconstruction, and 4 (3.1%) with three-point fixation combined with pelvic distraction. Complementary medial wall reconstruction was performed in 20 (15.6%) patients. All acetabular components were radiographically stable. Nine-year cumulative Kaplan–Meier survival rates for 123 patients with the endpoint defined as periprosthetic joint infection, any reoperation, and dissatisfaction were 96.91% (confidence interval [CI]: 86.26%, 99.34%), 97.66% (CI: 92.91%, 99.24%), and 96.06% (CI: 86.4%, 98.89%), respectively. Cup stability in cementless acetabular reconstruction depends on rim or three-point fixation. The continuity of the anterior and posterior columns determines whether the points provide adequate stability to the cup. Medial wall reconstruction is an important complementary fixation method for rim or three-point fixation. The patients who underwent cementless acetabular reconstruction guided by the RPC decision-making algorithm demonstrated satisfactory mid-term clinical function, satisfaction levels, radiographic results, and complication rates


Bone & Joint Open
Vol. 5, Issue 8 | Pages 671 - 680
14 Aug 2024
Fontalis A Zhao B Putzeys P Mancino F Zhang S Vanspauwen T Glod F Plastow R Mazomenos E Haddad FS

Aims

Precise implant positioning, tailored to individual spinopelvic biomechanics and phenotype, is paramount for stability in total hip arthroplasty (THA). Despite a few studies on instability prediction, there is a notable gap in research utilizing artificial intelligence (AI). The objective of our pilot study was to evaluate the feasibility of developing an AI algorithm tailored to individual spinopelvic mechanics and patient phenotype for predicting impingement.

Methods

This international, multicentre prospective cohort study across two centres encompassed 157 adults undergoing primary robotic arm-assisted THA. Impingement during specific flexion and extension stances was identified using the virtual range of motion (ROM) tool of the robotic software. The primary AI model, the Light Gradient-Boosting Machine (LGBM), used tabular data to predict impingement presence, direction (flexion or extension), and type. A secondary model integrating tabular data with plain anteroposterior pelvis radiographs was evaluated to assess for any potential enhancement in prediction accuracy.


Bone & Joint 360
Vol. 13, Issue 4 | Pages 13 - 16
2 Aug 2024

The August 2024 Hip & Pelvis Roundup360 looks at: Understanding perceived leg length discrepancy post-total hip arthroplasty: the role of pelvic obliquity; Influence of femoral stem design on revision rates in total hip arthroplasty; Outcomes of arthroscopic labral treatment of femoroacetabular impingement in adolescents; Characteristics and quality of online searches for direct anterior versus posterior approach for total hip arthroplasty; Rapid return to braking after anterior and posterior approach total hip arthroplasty; How much protection does a collar provide?; Timing matters: reducing infection risk in total hip arthroplasty with corticosteroid injection intervals; Identifying pain recovery patterns in total hip arthroplasty using PROMIS data.