Lengthening osteotomies of the calcaneus in children are in general
grafted with bone from the iliac crest. Artificial bone grafts have
been introduced, however, their structural and clinical durability
has not been documented. Radiostereometric analysis (RSA) is a very
accurate and precise method for measurements of rigid body movements including
the evaluation of joint implant and fracture stability, however,
RSA has not previously been used in clinical studies of calcaneal
osteotomies. We assessed the precision of RSA as a measurement tool
in a lateral calcaneal lengthening osteotomy (LCLO). LCLO was performed in six fixed adult cadaver feet. Tantalum
markers were inserted on each side of the osteotomy and in the cuboideum.
Lengthening was done with a plexiglas wedge. A total of 24 radiological
double examinations were obtained. Two feet were excluded due to
loose and poorly dispersed markers. Precision was assessed as systematic
bias and 95% repeatability limits.Objectives
Methods
The accuracy and precision of two new methods of model-based
radiostereometric analysis (RSA) were hypothesised to be superior
to a plain radiograph method in the assessment of polyethylene (PE)
wear. A phantom device was constructed to simulate three-dimensional
(3D) PE wear. Images were obtained consecutively for each simulated
wear position for each modality. Three commercially available packages
were evaluated: model-based RSA using laser-scanned cup models (MB-RSA),
model-based RSA using computer-generated elementary geometrical
shape models (EGS-RSA), and PolyWare. Precision (95% repeatability
limits) and accuracy (Root Mean Square Errors) for two-dimensional
(2D) and 3D wear measurements were assessed.Objectives
Methods