The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 459 - 462
1 Apr 2009
Heinert G Hendricks J Loeffler MD

Digital templating in hip replacement is commonly performed with radiological markers to determine the magnification. The latter can also be determined by measuring the distance from the x-ray focal spot to the object and the distance from the x-ray focal spot to the radiological cassette or image receptor. We used post-operative radiographs of total hip replacements and hemiarthroplasties from 22 patients to calculate the magnification using both methods. The accuracy of each method was ascertained by measuring the size of the head of the implant projected on to the radiographs and comparing the result with the known size recorded in the medical records.

The accuracy was found to be similar with a mean absolute measurement error of 2.6% (sd 1.4; 0.0% to 5.2%) for the radiological marker and 2.8% (sd 2.2; 0.4% to 10.1%) for the distance method (p = 0.75). The mean radiation dose for templating radiographs (pelvis and lateral of the hip) was similar when taken using a radiological marker (328 mSv sd 142) (n = 51) or using the distance measurement method (324 mSv sd 39) (n = 39) (p = 0.90).

We conclude that the distance measuring method is as accurate as the radiological marker method, but may avoid some of the disadvantages such as misplacement of the marker or placement outside the radiological field. It may also be more acceptable to the patient and radiographer.