To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article: