To assess the extent of osteointegration in two designs of shoulder
resurfacing implants. Bony integration to the Copeland cylindrical
central stem design and the Epoca RH conical-crown design were compared. Implants retrieved from six patients in each group were pair-matched.
Mean time to revision surgery of Copeland implants was 37 months
(standard deviation (Aims
Patients and Methods
Resurfacing of the humeral head is commonly used
within the UK to treat osteoarthritis (OA) of the shoulder. We present
the results of a small prospective randomised study of this procedure
using the Global CAP prosthesis with two different coatings, Porocoat
and DuoFix hydroxyapatite (HA). We followed two groups of ten patients
with OA of the shoulder for two years after insertion of the prosthesis
with tantalum marker beads, recording pain, Constant–Murley and
American Shoulder and Elbow Surgeons (ASES) outcome scores, and
using radiostereometric analysis to assess migration. The outcomes
were similar to those of other series, with significant reductions
in pain (p = 0.003) and an improvement in the Constant (p = 0.001)
and ASES scores (p = 0.006). The mean migration of the prosthesis
three months post-operatively was 0.78 mm (0.51 to 1.69) and 0.72
mm (0.33 to 1.45) for the Porocoat and DuoFix groups, respectively.
Analysis of variance indicated that the rate of migration reached
a plateau after three months post-operatively in both groups. At
follow-up of two years the mean migration was 1 mm ( The addition of a coating of HA to the sintered surface does
not improve fixation of this prosthesis. Cite this article: