The outcomes of patients with unexpected positive cultures (UPCs) during revision total hip arthroplasty (THA) and total knee arthroplasty (TKA) remain unknown. The objectives of this study were to establish the prevalence and infection-free implant survival in UPCs during presumed aseptic single-stage revision THA and TKA at mid-term follow-up. This study included 297 patients undergoing presumed aseptic single-stage revision THA or TKA at a single treatment centre. All patients with at least three UPCs obtained during revision surgery were treated with minimum three months of oral antibiotics following revision surgery. The prevalence of UPCs and causative microorganisms, the recurrence of periprosthetic joint infections (PJIs), and the infection-free implant survival were established at minimum five years’ follow-up (5.1 to 12.3).Aims
Methods
The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre. All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery.Aims
Methods
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods