The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs). This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.Aims
Methods
To compare the gait of unicompartmental knee arthroplasty (UKA)
and total knee arthroplasty (TKA) patients with healthy controls,
using a machine-learning approach. 145 participants (121 healthy controls, 12 patients with cruciate-retaining
TKA, and 12 with mobile-bearing medial UKA) were recruited. The
TKA and UKA patients were a minimum of 12 months post-operative,
and matched for pattern and severity of arthrosis, age, and body
mass index. Participants walked on an instrumented treadmill until their
maximum walking speed was reached. Temporospatial gait parameters,
and vertical ground reaction force data, were captured at each speed.
Oxford knee scores (OKS) were also collected. An ensemble of trees
algorithm was used to analyse the data: 27 gait variables were used
to train classification trees for each speed, with a binary output
prediction of whether these variables were derived from a UKA or
TKA patient. Healthy control gait data was then tested by the decision
trees at each speed and a final classification (UKA or TKA) reached
for each subject in a majority voting manner over all gait cycles
and speeds. Top walking speed was also recorded.Aims
Patients and Methods
Total knee arthroplasty (TKA) is an established
and successful procedure. However, the design of prostheses continues
to be modified in an attempt to optimise the functional outcome
of the patient. The aim of this study was to determine if patient outcome after
TKA was influenced by the design of the prosthesis used. A total of 212 patients (mean age 69; 43 to 92; 131 female (62%),
81 male (32%)) were enrolled in a single centre double-blind trial
and randomised to receive either a Kinemax (group 1) or a Triathlon
(group 2) TKA. Patients were assessed pre-operatively, at six weeks, six months,
one year and three years after surgery. The outcome assessments
used were the Oxford Knee Score; range of movement; pain numerical
rating scales; lower limb power output; timed functional assessment
battery and a satisfaction survey. Data were assessed incorporating
change over all assessment time points, using repeated measures
analysis of variance longitudinal mixed models. Implant group 2
showed a significantly greater range of movement (p = 0.009), greater
lower limb power output (p = 0.026) and reduced report of ‘worst
daily pain’ (p = 0.003) over the three years of follow-up. Differences
in Oxford Knee Score (p = 0.09), report of ‘average daily pain’
(p = 0.57) and timed functional performance tasks (p = 0.23) did
not reach statistical significance. Satisfaction with outcome was
significantly better in group 2 (p = 0.001). These results suggest that patient outcome after TKA can be influenced
by the prosthesis used. Cite this article: