Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1455 - 1462
1 Nov 2016
Matharu GS Berryman F Brash L Pynsent PB Dunlop DJ Treacy RBC

Aims. We investigated whether blood metal ion levels could effectively identify patients with bilateral Birmingham Hip Resurfacing (BHR) implants who have adverse reactions to metal debris (ARMD). Patients and Methods. Metal ion levels in whole blood were measured in 185 patients with bilateral BHRs. Patients were divided into those with ARMD who either had undergone a revision for ARMD or had ARMD on imaging (n = 30), and those without ARMD (n = 155). Receiver operating characteristic analysis was used to determine the optimal thresholds of blood metal ion levels for identifying patients with ARMD. Results. The maximum level of cobalt or chromium ions in the blood was the parameter which produced the highest area under the curve (91.0%). The optimal threshold for distinguishing between patients with and without ARMD was 5.5 µg/l (83.3% sensitivity, 88.4% specificity, 58.1% positive and 96.5% negative predictive values). Similar results were obtained in a subgroup of 111 patients who all underwent cross-sectional imaging. Between 3.2% and 4.3% of patients with ARMD were missed if United Kingdom (7 µg/l) and United States (10 µg/l) authority thresholds were used respectively, compared with 2.7% if our implant specific threshold was used, though these differences did not reach statistical significance (p ≥ 0.248). Conclusion. Patients with bilateral BHRs who have blood metal ion levels below our implant specific threshold were at low-risk of having ARMD. Cite this article: Bone Joint J 2016;98-B:1455–62


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1020 - 1027
1 Aug 2017
Matharu GS Judge A Pandit HG Murray DW

Aims. To determine the outcomes following revision surgery of metal-on-metal hip arthroplasties (MoMHA) performed for adverse reactions to metal debris (ARMD), and to identify factors predictive of re-revision. Patients and Methods. We performed a retrospective observational study using National Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery for ARMD between 2008 and 2014. The outcomes studied following revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using competing-risk regression modelling. Results. Intra-operative complications occurred in 40 revisions (1.6%). The cumulative five-year patient survival rate was 95.9% (95% confidence intervals (CI) 92.3 to 97.8). Re-revision surgery was performed in 192 hips (7.6%). The cumulative five-year implant survival rate was 89.5% (95% CI 87.3 to 91.3). Predictors of re-revision were high body mass index at revision (subhazard ratio (SHR) 1.06 per kg/m. 2 . increase, 95% CI 1.02 to 1.09), modular component only revisions (head and liner with or without taper adapter; SHR 2.01, 95% CI 1.19 to 3.38), ceramic-on-ceramic revision bearings (SHR 1.86, 95% CI 1.23 to 2.80), and acetabular bone grafting (SHR 2.10, 95% CI 1.43 to 3.07). These four factors remained predictive of re-revision when the missing data were imputed. Conclusion. The short-term risk of re-revision following MoMHA revision surgery performed for ARMD was comparable with that reported in the NJR following all-cause non-MoMHA revision surgery. However, the factors predictive of re-revision included those which could be modified by the surgeon, suggesting that rates of failure following ARMD revision may be reduced further. Cite this article: Bone Joint J 2017;99-B:1020–7


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1060 - 1069
1 Oct 2023
Holleyman RJ Jameson SS Reed M Meek RMD Khanduja V Hamer A Judge A Board T

Aims

This study describes the variation in the annual volumes of revision hip arthroplasty (RHA) undertaken by consultant surgeons nationally, and the rate of accrual of RHA and corresponding primary hip arthroplasty (PHA) volume for new consultants entering practice.

Methods

National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man were received for 84,816 RHAs and 818,979 PHAs recorded between April 2011 and December 2019. RHA data comprised all revision procedures, including first-time revisions of PHA and any subsequent re-revisions recorded in public and private healthcare organizations. Annual procedure volumes undertaken by the responsible consultant surgeon in the 12 months prior to every index procedure were determined. We identified a cohort of ‘new’ HA consultants who commenced practice from 2012 and describe their rate of accrual of PHA and RHA experience.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims

To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure.

Patients and Methods

We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression.