Esposito C Roques A Tuke M Walsh WK Walter WL
Full Access

Introduction

Two types of ceramic materials currently used in total hip replacements are third generation hot isostatic pressed (HIPed) alumina ceramic (commercially known as BIOLOX®forte, CeramTec) and fourth generation alumina matrix composite ceramic consisting of 75% alumina, 24% zirconia, and 1% mixed oxides (BIOLOX®delta, CeramTec). Delta ceramic hip components are being used worldwide, but very few studies have analyzed retrieved delta bearings. The aim of this study is to compare edge loading ‘stripe’ wear on retrieved femoral heads from delta-on-delta, delta-on-forte and forte-on-forte ceramic bearings revised within 2 years in vivo.

Material and Methods

Ceramic bearings revised at one center from 1998 to 2010 were collected (61 bearings). Eleven delta heads revised between 1–33 months were compared to 24 forte femoral heads with less than 24 months in vivo (Figure 1). The surface topography of the femoral heads was measured using a RedLux AHP (Artificial Hip Profiler, RedLux Ltd, Southampton, UK). Three representative samples were examined with a FEI Quanta 200 Scanning Electron Microscope (SEM).