Numerous papers present in-vivo knee kinematics data following total knee arthroplasty (TKA) from fluoroscopic testing. Comparing data is challenging given the large number of factors that potentially affect the reported kinematics. This paper aims at understanding the effect of following three different factors: implant geometry, performed activity and analysis method. A total of 30 patients who underwent TKA were included in this study. This group was subdivided in three equal groups: each group receiving a different type of posterior stabilized total knee prosthesis. During single-plane fluoroscopic analysis, each patient performed three activities: open chain flexion extension, closed chain squatting and chair-rising. The 2D fluoroscopic data were subsequently converted to 3D implant positions and used to evaluate the tibiofemoral contact points and landmark-based kinematic parameters. Significantly different anteroposterior translations and internal-external rotations were observed between the considered implants. In the lateral compartment, these differences only appeared after post-cam engagement. Comparing the activities, a significant more posterior position was observed for both the medial and lateral compartment in the closed chain activities during mid-flexion. A strong and significant correlation was found between the contact-points and landmarks-based analyses method. However, large individual variations were also observed, yielding a difference of up to 25% in anteroposterior position between both methods. In conclusion, all three evaluated factors significantly affect the obtained tibiofemoral kinematics. The individual implant design significantly affects the anteroposterior tibiofemoral position, internal-external rotation and timing of post-cam engagement. Both kinematics and post-cam engagement additionally depend on the activity investigated, with a more posterior position and associated higher patella lever arm for the closed chain activities. Attention should also be paid to the considered analysis method and associated kinematics definition: analyzing the tibiofemoral contact points potentially yields significantly different results compared to a landmark-based approach.
Intra-articular cartilage pressure distribution in the knee joint is critical in the understanding of osteoarthritis. Combining personalized statistical modeling of the morphological characteristics with discrete element modeling enables patient-specific predictions of the pressure on the tibial plateau. However, modeling of the meniscus during gait is complicated by the dynamic nature of the structure. Nevertheless, the position of the meniscus has a substantial impact on intra-articular stress distribution. Therefore, the focus of this presentation will be on how modeling of meniscal movement during knee flexion improves insight in general meniscal kinematics for the use in tibiofemoral stress distribution calculations.
7–20 % of the patients with a total knee arthroplasty (TKA) are dissatisfied without an indication for revision. Therapeutic options for this patient population with mostly a lack of quadriceps strength are limited. The purpose of this study is to evaluate the effect of six weeks low load resistance training with blood flow restriction (BFR) on the clinical outcome in these unhappy TKA patients. Thirty-one unhappy TKA patients (of the scheduled fifty patients) without mechanical failure were included in this prospective study since 2022. The patients participate in a supervised resistance training combined with BFR, two times a week during nine weeks. Patients were evaluated by the Knee Osteoarthritis Outcome Score (KOOS), Knee Society Score: satisfaction (KSSs) and the Pain Catastrophizing Scale (PCS). Functionality was tested using the Six Minute Walk Test (6MWT) and the 30-Second Chair Stand Test (30CST). Follow-up took place at six weeks, three months and six months after the start. Six weeks training with BFR provided statistically significant improvements in all the KOOS subscales compared to the baseline, especially for symptoms (55.1 (±15.4) versus 48.0 (±16.5); p<0.001), activities in daily living (50.3 (±21.1) versus 43.7 (±17.2); p<0.00) and quality of life (24.6 (±18.5) versus 17.3 (±13.0); p<0.001). The PCS reduced from 27.4 (±11.0) to 23.2 (±11.4) at six weeks (p<0.01), whereas the KSSs increased from 11.8 (±6.5) to 14.9 (±7.6) (p=0.021). Both the 6MWT and the 30CST improved statistically significant from respectively 319.7 (±15.0) to 341.6m (±106.5) (p<0.01) and 8.6 (±3.9) to 9.3 times (±4.5) (p<0.01). Blood flow restriction appears to enhance the quality of life and functional performance of unhappy TKA patients. Based on these preliminary results, BFR seems to be a promising and valuable alternative for these TKA patients with limited therapeutic options.
Several emerging reports suggest an important involvement of the hindfoot alignment in the outcome of knee osteotomy. At present, studies lack a comprehensive overview. Therefore, we aimed to systematically review all biomechanical and clinical studies investigating the role of the hindfoot alignment in the setting of osteotomies around the knee. A systematic literature search was conducted on multiple databases combining “knee osteotomy” and “hindfoot/ankle alignment” search terms. Articles were screened and included according to the PRISMA guidelines. A quality assessment was conducted using the Quality Appraisal for Cadaveric Studies (QUACS) - and modified methodologic index for non-randomized studies (MINORS) scales. Three cadaveric, fourteen retrospective cohort and two case-control studies were eligible for review. Biomechanical hindfoot characteristics were positively affected (n=4), except in rigid subtalar joint (n=1) or talar tilt (n=1) deformity. Patient symptoms and/or radiographic alignment at the level of the hindfoot did also improve after knee osteotomy (n=13), except in case of a small pre-operative lateral distal tibia- and hip knee ankle (HKA) angulation or in case of a large HKA correction (>14.5°). Additionally, a pre-existent hindfoot deformity (>15.9°) was associated with undercorrection of lower limb alignment following knee osteotomy. The mean QUACS score was 61.3% (range: 46–69%) and mean MINORS score was 9.2 out of 16 (range 6–12) for non-comparative and 16.5 out of 24 (range 15–18) for comparative studies. Osteotomies performed to correct knee deformity have also an impact on biomechanical and clinical outcomes of the hindfoot. In general, these are reported to be beneficial, but several parameters were identified that are associated with newly onset – or deterioration of hindfoot symptoms following knee osteotomy. Further prospective studies are warranted to assess how diagnostic and therapeutic algorithms based on the identified criteria could be implemented to optimize the overall outcome of knee osteotomy. Remark: Aline Van Oevelen and Arne Burssens contributed equally to this work
The goal was to evaluate tibiofemoral knee joint kinematics during stair descent, by simulating the full stair descent motion in vitro. The knee joint kinematics were evaluated for two types of knee implants: bi-cruciate retaining and bi-cruciate stabilized. It was hypothesized that the bi-cruciate retaining implant better approximates native kinematics. The in vitro study included 20 specimens which were tested during a full stair descent with physiological muscle forces in a dynamic knee rig. Laxity envelopes were measured by applying external loading conditions in varus/valgus and internal/external direction.Aims
Methods
Robot systems have been successfully introduced to improve the accuracy and reduce severe iatrogenic soft tissue damage in knee arthroplasty. Unfortunately to perform complete a complete bone cut, the cutting tool has to slightly pass the edge of the bone. In the posterior zones were retractor protection is impossible this will lead to contact between the cutting tool and the soft tissue envelope. Therefore, complete soft tissue preservation cannot be guaranteed with the current commercial systems. This study presents an alternative robotic controlled cutting technique to perform the bone resections during TKA by milling a slot with a long slender high-speed milling tool. The system is composed by a long milling tool driven by a high-speed motor and a protector covering the end of the cutter. The protector is rigidly connected to the motor by the support structure next to the mill, which moves behind the mill in the slot created by the cutter. The protector at the end of the cutter has four functions: providing mechanical support for the mill, preventing soft tissue to come into contact with the cutter, sensing the edge of the bone to accurately follow the shape of the bone and releasing the attached soft tissue. The edge of the bone is sensed by force feedback and with the help of a probing motion the adaptive algorithm enables the protector to follow the edge of the bone closely by compensating for small segmentation and registration errors. A pilot test to evaluate the concept was performed on three fresh frozen knees. The flatness of the resection, the iatrogenic soft tissue damage, the cutting time and the efficiency of the bone contour following algorithm was measured.Introduction
Methods
Under- or oversizing of either component of a total knee implant can lead to early component loosening, instability, soft tissue irritation or overstuffing of joint gaps. All of these complications may cause postoperative persistent pain or stiffness. While survival of primary TKA's is excellent, recent studies show that patient satisfaction is worse. Up to 20% of the patients are not satisfied with the outcome as and residual pain is still a frequent occurrence. The goal of this study was therefore to evaluate if the sizing of the femoral component, as measured on a 3D-reconstructed projection, is related to patient reported outcome measures. From our prospectively collected TKA outcome database, all patients with a preoperative CT and a postoperative X-ray of their operated knee were included in this study. Of these 43 patients, 26 (60,5%) were women and 17 (39,5%) were men. The mean age (+/−SD) was 74,6 +/− 9 years. CT scans were acquired. All patients underwent TKA surgery in a single institution by one surgical team using the same bi- cruciate substituting total knee (Journey II BCS, Smith&Nephew, Memphis, USA). Using a recently released X-ray module in Mimics (Materialise NV, Leuven, Belgium), this module allows to align the post-operative bi-planar x-rays with the 3D- reconstructed pre-operative distal femur and to determine the 3D position of the bone and implant models using the CAD- file of the implant. This new technique was validated at our department and was found to have a sub-degree, sub-millimeter accuracy. Eleven zones of interest were defined. On the medial and the lateral condyle, the extension, mid-flexion and deep flexion facet were determined. Corresponding trochlear zones were defined and two zones were defined to evaluate the mediolateral width. In order to compare different sizes, elastic deforming mesh matching algorithms were implemented to transfer the selected surfaces from one implant to another. The orthogonal distances from the implant to the nearest bone were calculated. Positive values represent a protruding (oversized) femoral component, negative values an undersized femoral component. The figure shows the marked zones on the femoral implant. The KOOS subscores and KSS Satisfaction subscore were evaluated.Background
Methods
Kinematic patterns in total knee arthroplasty (TKA) can vary considerably from the native knee. No study has shown a relation between a given kinematic pattern and patient satisfaction yet. The purpose of this study was to test whether the kinematical pattern, and more specifically the anteroposterior translation during (1) open kinetic chain flexion-extension, (2) closed kinetic chain chair rising and (3) squatting, is related to the level of patient satisfaction after TKA.Background
Questions
Hindfoot disorders are complex 3D deformities. Current literature has assessed their influence on the full leg alignment, but the superposition of the hindfoot on plain radiographs resulted in different measurement errors. Therefore, the aim of this study is to assess the hindfoot alignment on Weight-Bearing CT (WBCT) and its influence on the radiographic Hip-Knee-Ankle (HKA) angle. A retrospective analysis was performed on a study population of 109 patients (mean age of 53 years ± 14,49) with a varus or valgus hindfoot deformity. The hindfoot angle (HA) was measured on the WBCT while the HKA angle, and the anatomical tibia axis angle towards the vertical (TAX) were analysed on the Full Leg radiographs. The mean HA in the valgus hindfoot group was 9,19°±7.94, in the varus hindfoot group −7,29°±6.09. The mean TAX was 3,32°±2.17 in the group with a valgus hindfoot and 1,89°±2.63 in the group with a varus hindfoot, which showed to be statistically different (p<0.05). The mean HKA Angle was −1,35°±2.73 in the valgus hindfoot group and 0,4°±2.89 in the varus hindfoot group, which showed to be statistically different (p<0.05). This study demonstrates a higher varus in both the HKA and TAX in valgus hindfoot and a higher tibia valgus in varus hindfoot. This contradicts the previous assumption that a varus hindfoot is associated with a varus knee or vice versa. In clinical practice, these findings contribute to a better understanding of deformity corrections of both the hindfoot and the knee.
An adult acquired flatfoot deformity (AAFD) is a complex 3D deformity. Surgical correction consists of a medial calcaneal osteotomy (MCO) but shows limitations due to the current 2D assessment. Therefore, the aim is to determine the influence of an MCO on the longitudinal foot arch assessed by 2D and 3D weightbearing CT (WBCT). Seventeen patients with a mean age of 44,5 years (range 18–66 yrs) were retrospectively included. MCO was indicated in a stage II AAFD (N=15) and a post-traumatic valgus deformity (N=2). Pre- and post-operative imaging was obtained from a WBCT. The height of the longitudinal foot arch was measured as the distance from the navicular tuberositas to the floor (Navicular Height, NH) on 2D CT images (NH2D) and computed on 3D CT data (NH3D). Additionally, 3D assessment could compute the degree of exorotation (α) of the navicular bone towards the vertical axis. The mean pre-operative NH2D and NH3D were respectively 29.57mm ± 7.59 and 28.34mm ± 6.51. These showed to be statistically different from the mean post-operative NH2D and NH3D, respectively 31.62mm ± 6.69 and 31.67mm ± 6.47 (p < 0,001). A statistical difference was also found when comparing the mean degree of exorotation in pre- and post-operative, respectively: αpre=14.08° ± 4,92 and the αpost=19,88° ± 3.50 (p < 0,001). This study demonstrates a significant correction of the longitudinal foot arch after a MCO. The novelty is attributed to the accurate degree of rotation assessment using WBCT. This information could be assistive to optimise a pre-operative planning.
Patients with a hindfoot deformity impose a particular challenge when performing a total knee arthroplasty (TKA). This could be attributed to the lack of insights concerning the outcome towards the hindfoot alignment. Our objective was to perform a systematic review of the literature to investigate the influence of TKA on hindfoot alignment and vice-versa. In accordance with the Methodological Index For Non-Randomized Studies (MINORS) statement standards, we performed a systematic review. Electronic databases Pubmed, EMBASE, Web of Science, Google Scholar and Cochrane Library were searched to identify capable studies studying the influence between TKA and hindfoot malalignment. We indentified four prospective cohort studies, seven retrospective cohort studies and one case-control study. All twelve articles addressed the influence of TKA on hindfoot alignment. Seven out of nine studies which noticed an improvement of hindfoot alignment after TKA, found a significant improvement (p<0.05). Aditionally three of these studies reported a significant improvement only in valgus hindfeet (p<0.05). On the topic of hindfoot alignment influencing TKA, we identified two studies. These studies reported an impact of hindfoot alignment on the weightbearing and described that 87% of hindfeet remained in valgus alignment after TKA. Available data suggests that alignment in valgus hindfeet can improve after TKA, though long term results are not present. Contrary to last, improvement of hindfoot alignment is not expected in varus hindfeet. Furthermore hindfoot alignment deformity may cause a reduction of the long term survival of the knee prosthesis and therefore should be taken in to account.
A calcaneal medial osteotomy (CMO) is a surgical procedure frequently performed to correct a valgus alignment of the hindfoot. However currently little is known on its accurate influence on hindfoot alignment (HA). To assess the influence of a CMO on HA in both 2D and 3D measurements using weightbearing CT (WBCT).Background
Aim
Thorough understanding and feedback of the post-operative implant position relative to the pre-operative anatomy is missing in today's clinical practice. However, three dimensional insights in the local under or oversizing of the implant can provide important feedback to the surgeon. For the knee for instance, to identify a shift in the sagittal joint line that potentially links to mid-flexion instability or to identify zones at risk for soft tissue impingement. Despite a proven inferior outcome, clinical post-operative implant evaluation remains primarily based on bi-planar, static 2D x-rays rather than 3D imaging. Along with the cost, a possible reason is the increased radiation dose and/or metal artifact scatter in computed tomography (CT) and/or magnetic resonance imaging (MRI). These detrimental effects are now avoided by using recently released x-ray processing software. This technique uses standard-of-care post-operative x-rays in combination with a pre-operative CT and 3D file of the implant to determine the implant position relative to the pre-operative situation. The accuracy of this new technique is evaluated in this paper using patient cases. Therefore, the obtained implant position is benchmarked against post-operative CT scans. Retrospectively, 19 patients were selected who underwent total knee arthroplasty and received pre- and post-operative CT of their diseased knee. The CT scans were performed with a pixel size of 0.39 mm and slice spacing of 0.60 mm (Somatom, Siemens, München, Germany). All patients underwent TKA surgery using the same bi-cruciate substituting total knee (Journey II, Smith&Nephew, Memphis, USA). Following surgery, standard bi-planar standing x-rays of the operated knee was additionally performed as standard of care. To evaluate the implant position relative to the pre-operative situation, the 3D implants are first positioned on the post-operative CT slices. Using Mimics (Materialise NV, Leuven, Belgium), the pre-operative bone was subsequently automatically matched onto the post-operative scan to identify the implant location relative to the reconstructed pre-operative bone. This has been independently repeated by three observers to assess the inter-observer variability. Second, the post-operative bi-planar x-rays are combined with the reconstructed pre-operative bone and 3D file of the implant. This combination is performed using the 2D-to-3D conversion integrated in the recently launched X-ray module of Mimics. This module uses a contour based registration method to determine the implant and bone position using the post-operative x-rays. For both reconstruction methods, the implant position has been evaluated in six degrees of freedom using an automated Matlab routine; resulting in three translations and three rotations.INTRODUCTION
MATERIALS & METHODS
To assess and compare the effect of new orthopedic surgical procedures, in vitro evaluation remains critical during the pre-clinical validation. Focusing on reconstruction surgery, the ability to restore normal kinematics and stability is thereby of primary importance. Therefore, several simulators have been developed to study the kinematics and create controlled boundary conditions. To simultaneously capture the kinematics in six degrees of freedom as outlined by Grood & Suntay, markers are often rigidly connected to the moving bone segments. The position of these markers can subsequently be tracked while their position relative to the bones is determined using computed tomography (CT) of the test specimen with the markers attached. Although this method serves as golden standard, it clearly lacks real-time feedback. Therefore, this paper presents the validation of a newly developed real-time framework to assess knee kinematics at the time of testing. A total of five cadaveric fresh frozen lower limb specimens have been used to quantitatively assess the difference between the golden standard, CT based, method and the newly developed real-time method. A schematic of the data flow for both methods. Prior to testing, both methods require a CT scan of the full lower limb. During the tests, the proximal femur and distal tibia are necessarily resected to fit the knees in the test setup, thus also removing the anatomical landmarks needed to evaluate their mechanical axis. Subsequently, a set of three passive markers are rigidly attached to the femur and tibia, referred to as M3F and M3T respectively. For the CT based method, the marker positions are captured during the tests and a second CT scan is eventually performed to link the marker positions to the knee anatomy. Using in-house developed software, this allowed to offline evaluate the knee kinematics in six degrees of freedom by combining both CT datasets with the tracked marker positions. For the newly developed real-time method, a calibration procedure is first performed. This calibration aims to link the position of the 3D reconstructed bone and landmarks with the attached markers. A set of bone surface points is therefore registered. These surface points are obtained by tracking the position of a pen while touching the bone surface. The pen's position is thereby tracked by three rigidly attached markers, denoted M3P. The position of the pen tip is subsequently calculated from the known pen geometry. The iterative closest point (ICP) algorithm is then used to match the 3D reconstructed bone to the registered surface points. Two types of 3D reconstructions have therefore been considered. First, the original reconstructions were used, obtained from the CT data. Second, a modified reconstruction was used. This modification accounted for the finite radius (r = 1.0 mm) of the registration pen, by shifting the surface nodes 1.0 mm along the direction of the outer surface normal. During the tests, the positions of the femur and tibia markers are tracked and streamed in real-time to an in-house developed, Matlab based software framework (MathWorks Inc., Natick, Massachussets, USA). This software framework simultaneously calculates the bone positions and knee kinematics in six degrees of freedom, displaying this information to the surgeons and operators. To assess the accuracy, all knee specimens have been subjected to passive flexion-extension movement ranging from 0 to 120 degrees of flexion. For each degree of freedom, the average root mean square (RMS) difference between both measurement methods has been evaluated during this movement. In addition, the distribution of the registered surface points has been assessed along the principal directions of the uniformly meshed 3D reconstructions (average mesh size of 1.0 mm).INTRODUCTION
MATERIALS & METHODS
The goal of this study was (1) to investigate the relationships between the bony contours of the knee and the popliteus tendon before and after TKA and (2) to analyse the influence of implant sizing. Our hypothesis was that an apparently well-sized prosthesis, will modify the position or the tracking of popliteus tendon. 4 fresh frozen cadavers were selected. The popliteus tendon was injected with contrast dye and a CT-scan was performed from full extension to full flexion with increments of 20°. Afterwards a total knee arthroplasty (TKA) was performed. Each cadaver received either a normal-sized, oversized, undersized or mobile bearing prosthesis. After TKA the limb was scanned again using the same protocol as before. 3D-reconstructions were created using Materialise Mimics software. These 3D-models were then imported into custom made Matlab software to measure and compare the deviation of the popliteus tendon before and after TKA.Background
Methods
A new knee simulator has been developed at Ghent University. This simulator provides the unique opportunity of evaluating the knee kinematics during activities of daily living. The simulator therefore controls the position of the ankle in the sagittal plane while keeping the hip at a fixed position. This approach provides full kinematic freedom to the knee. To evaluate and validate the performance of the simulator, the development of and comparison with a numerical simulation model is discussed in this paper. Both a two and three dimensional simulation model have been developed using the AnyBody Modelling System (AMS). In the two dimensional model, the knee joint is represented by a hinge. Similarly, the ankle and hip joint are represented by a hinge joint and a variable amplitude quadriceps and hamstrings force is applied. In line with this simulation model, a hinge model was created that could be mounted in the UGent knee simulator to evaluate the performance of the simulated model. The hinge model thereby performs a cyclic motion under varying simulated muscle loads while recording the ankle reaction forces. In addition to the two dimensional model, a three dimensional model has been developed. More specifically, a model is built of a sawbone leg holding a posterior stabilised single radius total knee implant. The physical sawbone model contains simplified medial and lateral collateral ligaments. In line with the boundary conditions of the UGent knee simulator, the simulated hip contains a single rotational degree of freedom and the ankle holds four degrees of freedom (three rotations, single translation). In the simulations, the knee is modelled using the force-dependent kinematics (FDK) method built in the AMS. This leaves the knee with six degrees of freedom that are controlled by the ligament tension in combination with the applied quadriceps load and shape of the implant. The physical sawbone model goes through five cycles in the UGent simulator using while recording the kinematics of the femur and tibia using a set of markers rigidly attached to the femur and tibia bone. The position of the implant with respect to the markers was evaluated by CT-scanning the sawbone model.Background
Methods
Biomechanical joint contact pressure distribution measurements have proven to be a very valuable tool in orthopaedic research to investigate the influence of surgical techniques such as total knee arthroplasty (TKA) on the human knee joint. Quantification of the in vitro tibiofemoral and patellofemoral contact pressure distribution before and after the intervention are an important measure to evaluate the impact of the surgery. The K scan pressure sensor from Tekscan (South Boston USA) is a commonly reported device for these in vitro pressure measurements. Despite the large interest in the sensor, the effective measurement accuracy for in vitro biomechanical joint contact measurement still remains a big question and therefore the reliability of these measurements should be questioned. Reliable contact pressure measurements can only be done if the sensor behaviour is fully understood. Therefore, a tailored multi-axial testing machine has been designed to profoundly investigate and characterise the sensor behaviour. This test setup is unique through its ability to apply a predefined tangential force or sliding velocity to the sensor's interface next to a normal force. Dynamic effects occurring in knee joint motion can thus be simulated while evaluating the effect on the contact pressure measurements.Background
Methods
Better functional outcomes, lower pain and better stability have been reported with knee designs which restore physiological knee kinematics. Also the ability of the TKA design to properly restore the physiological femoral rollback during knee flexion, has shown to be correlated with better restoration of the flexor/extensor mechanism, which is fundamental to the function of the human knee. The purpose of the study is to compare the kinematics of three different TKA designs, by evaluating knee motion during Activities of Daily Living. The second goal is to see if there is a correlation between the TKA kinematics and the patient reported outcomes. Ten patients of each design, who are at least 6 months after their Total Knee Replacement, will be included in this study. Seven satisfied and 3 dissatisfied patients will be selected for each design. In this study 5 different movements will be analysed: flexion/extension; Sitting on and rising from a chair, Stair climbing, descending stairs, Flexion and extension open chain and squatting. These movements will be captured with a fluoroscope. The 2D images that are obtained, will be matched with the 3D implants. This 3D image will be processed with custom-made software to be able to analyse the movement. Tibio-femoral contact points of the medial and lateral condyles, tibio-femoral axial rotation, determination of the pivot-point will be analysed and described. After this analysis, a correlation between the kinematics and the KOOS and KSS will be investigated.Introduction
Methods
In-vitro testing of knee joints remains vital in the understanding of knee surgery and arthroplasty. However, based on the design philosophy of the original Oxford knee simulator, this in-vitro testing has mainly focused on squatting motion. As the activities of daily living might drastically differ from this type of motion, both from a kinematic and kinetic point of view, a new knee simulator is required that allows studying more random motion patterns. This paper describes a novel knee simulator that overcomes the limitations of traditional Oxford simulators, providing both kinematic and kinetic freedom with respect to the applied boundary conditions. This novel test simulator keeps the hip at a fixed position, only providing a single rotational degree of freedom (DOF) in the sagittal plane. In addition, the ankle holds four DOF, including all rotational DOF and the translation along the medio-lateral axis. Combining these boundary conditions leaves five independent DOF to the knee; the knee flexion angle is actively controlled through the positioning of the ankle joint in the antero-posterior and proximal-distal direction. The specimens' quadriceps muscle is actively controlled, the medial and lateral hamstrings are passively loaded. To validate the performance of this simulator, two fresh frozen specimens have been tested during normal squatting and cycling. Their kinematic patterns have been compared to relevant literature data.Background
Methods
During TKA surgery, the usual goal is to achieve equal balancing between the lateral and medial side, which can be achieved by ligament releases or “pie crusting”. However little is known regarding a relationship between the balancing forces on the medial and lateral plateaus during TKA surgery, and the varus and valgus and rotational laxities when the TKA components are inserted. It seems preferable that the laxity after TKA is the same as for the normal intact knee. Hence the first aim of this study was to compare the laxity envelope of a native knee, with the same knee after TKA surgery. The second aim was to examine the relationship between the Varus-Valgus (VV) laxity and the contact forces on the tibial plateau. A special rig that reproduced surgical conditions and fit onto an operating table was designed (Figure 1) (Verstraete et al. 2015). The rig allows application of a constant varus/valgus moment, and an internal-external (IE) torque. A series of heel push tests under these loading conditions were performed on 12 non-arthritic half semibodies hip-to-toe cadaveric specimens. Five were used for method development. To measure laxities, the flexion angle, the VV and the IE angle were measured using a navigation system. After testing the native knee, a TKA was performed using the Journey II BCS implant, the navigation assuring correct alignments. Soft tissue balancing was achieved by measuring compressive forces on the lateral and medial condyles with an instrumented tibial trial (Orthosensor, Dania Beach, Florida). At completion of the procedure, the laxity tests were repeated for VV and IE rotation and the contact forces on the tibial plateau were recorded, for the full range of flexion.INTRODUCTION
METHODS