Vasukutty NL King A Uglow MG
Full Access

Originally used for correction of angular malalignment, 2 hole plate epiphyseodesis has recently gained popularity in paediatric orthopaedic practice for the correction of leg length discrepancy. In this study we aim to assess the efficiency of guided growth plates in correcting leg length discrepancy

Thirty-three children treated for leg length discrepancy with guided growth plates (“8-Plate”, Orthofix, Inc and “I-Plate”, Orthopediatrics) in a tertiary referral centre were retrospectively analysed. Medial and Lateral plates were inserted for symmetrical growth reduction and patients were followed up with clinical and radiological assessment. Thirty patients had distal femoral epiphyseodesis and three had proximal tibial epiphyseodesis. Leg lengths and individual bone lengths were measured from pre and post – operative radiographs. The angle between the screws was measured from radiographs taken intra operatively and at the time of final follow up to assess screw divergence with growth. Efficiency was calculated as the ratio of growth inhibition achieved to the projected discrepancy at maturity if left untreated.

At a mean follow up of 17 months (4–30 m) leg length discrepancy improved from a mean of 30 mm (50–15mm) to 13 mm (2.5–39mm) (p < 0.01). The angle between screws increased from 6 degrees to 26 degrees over the follow up period. Efficiency was found to be 66%. There were 5 patients with angular deformity who needed plate removal and 2 patients developed superficial infection that responded to oral antibiotics.

Epiphyseodesis using guided growth plates is an effective way to correct leg length discrepancy as it is a reversible procedure. Patients undergoing this treatment should be kept under close follow up to prevent development of angular malalignment. Inserting the screws in a divergent fashion at the outset may increase the effectiveness of this procedure.