Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 141 - 141
1 May 2016
Frisch N Wessell N Van Holsbeeck M Silverton C
Full Access

Introduction

The use of metal-on-metal (MOM) and modular total hip arthroplasty (THA) is associated with potentially serious complications including elevated serum metal ion levels, pseudotumor, cardiomyopathy and neurologic abnormalities. The primary aim of this analysis was to identify any associations between the presence of pseudotumor, serum metal ion levels, and specific dual modular implant components.

Methods

We evaluated prospectively collected data from 49 patients, mean age 58.4 years, who underwent implantation of modular THA from 01/2004-01/2010. The collected data spanned a 5–11 year period from the time of index procedure. Serum metal ion levels, including titanium, cobalt and chromium, were collected in 2012 and 2015. Hip ultrasounds were performed on each patient by a trained musculoskeletal radiologist for evaluation of the presence of soft-tissue pseudotumor. Univariate nonparametric tests were used to compare the two groups: Fisher's exact tests for categorical variables, and Wilcoxon two-group tests for continuous variables. For the purposes of analysis, values that were below the level of detection (LOD) were set to half the LOD. All analyses were performed using SAS 9.4 (SAS Institute Inc, Cary, NC, USA). Statistical significance is set at p<0.05.