header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 98 - 98
1 Jun 2012
Ichinohe S Kamei Y Tokunaga S Suzuki M
Full Access

Purpose

Many TKA instruments were developed in these days. Distal femoral cutting guide using intra-medullary system were divided into two methods, from anterior or medial. Many companies employed anterior cutting guide, however these guides have a disadvantage of wide skin and quadriceps incision. Only Zimmer provided medial cut guide which performed short skin and quadriceps incision. However, reference point (medial femoral condyle) will be a risk of imprecise cutting for a medial condyle defect cases. We tried L-shaped new distal femoral cutting guide, reference point will be both femoral condyle and cutting from antero-medial side. The purpose of this study was to prove usefulness of the new guide.

Materials and Methods

Twenty-nine knees were employed in this study. All knees were treated with Optetrak knee system (Exactec). Surgical methods were as follows, mid line skin incision, short para-patellar deep incision, no patellar resurfacing, PS type implant and cement fixation were employed. 13 knees were used original anterior cutting guide (O group) and 16 knees were used new antero-medial cut guide (N group). Study items were length of skin incision, length of Quadriceps incision, surgical time, JOA score, and component tilting angles (implant position were compared to femoral axis with AP and lateral view of roentgenograms).