header advert
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 29 - 29
1 Oct 2019
Tetreault MW Perry KI Pagnano MW Hanssen AD Abdel MP
Full Access

Introduction

Metaphyseal fixation during revision total knee arthroplasties (TKAs) is important, but potentially challenging with historical cone designs. Material and manufacturing innovations have improved the size and shape of cones available, and simplified requisite bone preparation. In a very large series, we assessed implant survivorship, radiographic results, and clinical outcomes of new porous 3-D printed titanium metaphyseal cones featuring a reamer-based system.

Methods

We reviewed 142 revision TKAs using 202 cones (134 tibial and 68 femoral) from 2015 to 2016. Sixty cases involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. Mean age was 66 years, with 54 % females. Mean BMI was 34 kg/m2. Patients had a mean of 2.4 prior surgeries and 48% had a history of periprosthetic infection. Mean follow-up was 2 years.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 35 - 35
1 Oct 2019
Bedard NA Tetreault MW Hanssen AD Lewallen DG Trousdale RT Berry DJ Abdel MP
Full Access

Introduction

Cementation of a new liner into an existing well-fixed acetabular component is common during revision total hip arthroplasties (THAs) for many indications, but most commonly for lack of a modern compatible crosslinked polyethylene liner. However, little is known about the long-term durability of this strategy. The purpose of this study was to evaluate the long-term implant survivorship, risk of complications, clinical outcomes, and radiographic results of cementing a new highly cross-linked polyethylene (HXLPE) liner into a well-fixed acetabular component.

Methods

We retrospectively identified 326 revision THAs where a non-constrained HXLPE liner was cemented into a well-fixed acetabular component. Mean age at revision THA was 63 years, with 50% being female. The most common indications for revision THA were wear and osteolysis (49%), aseptic femoral loosening (35%), and instability (8%). Mean follow-up was 10 years.