header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_I | Pages 79 - 79
1 Mar 2002
Weber F Grobbelaar C du Plessis T Cakic J Spirakis A Cappaert G
Full Access

Wear of ultra-high molecular weight polyethylene (UHMWP) acetabular cups is a well-known cause of osteolysis and loosening of the components. Improvement of the wear resistance of UHMWP could extend the clinical life of total hip arthroplasty (THA). Chemical cross-linking in acetylene with gamma radiation is a cheap and effective way of increasing wear resistance of UHMWP.

This study is a report on 263 patients (123 males and 140 females) on whom Dr Weber performed THA between 1977 and 1984, using the Pretoria (Grobbelaar) monobloc stainless steel hip with 30-mm metal head. There were 96 patients (107 prostheses) available for follow-up at a mean of 18.3 years, with 89 surviving prosthesis in 79 patients (83.2%).

We collected complete sets of radiographs of 54 patients (mean age 71.4 years) for a radiological survey in 1999. In 41 patients (76%) we found no wear. The mean age of these patients was 72 years. Wear was noted in the other 13 patients (24%), whose mean age was 75 years. The mean follow-up time was 16 years (8 to 23). The mean magnification in the radiological study was 18%. Mean wear for the total group was 1.29 mm and mean annual wear 0.17 mm.

A similar analysis performed on a group of 64 of Dr Grobbelaar’s patients at 15.5 year mean follow-up shows remarkable similarity, with mean wear of 0.172 mm for a group of 64 patients and annual wear 0.11 mm. Dr Oonishi of Japan, who has conducted the only other long-term follow-up, found similarly promising results.