Purpose

The surgical treatment of scaphoid fractures consists of reduction of the fracture followed by stable internal fixation using a headless compression screw. Proper positioning of the screw remains technically challenging and therefore computer assisted surgery may have an advantage.

Navigation assisted surgery requires placement and registration of stable reference markers which is technically impossible in a small bone like the scaphoid. Custom made wrist-positioning devices with built-in reference markers have been developed for this purpose. The purpose of this study was to evaluate a different method of navigation assisted scaphoid fracture fixation. Temporary stabilisation with a pin of the scaphoid to the radius enables placement of the reference markers on the radius. Our hypothesis was that this method will achieve precise fracture fixation, superior to the standard free hand technique.

Methods

In 20 identical saw bone models with mobile scaphoids, the scaphoid was stabilised to the radius using one Kirschner wire (KW). An additional KW representing the fixating screw was placed either using the Mazor Renaissance Robotic System (MAZOR Surgical Technologies, Israel) or standard free hand technique. CT scans were performed prior to fixation and after fixation in order to plan the location of the KW and compare this planned location with the final result.