Vol. 90-B, Issue SUPP_I | Pages 63 - 63
1 Mar 2008
Lau J Stamatis E Parks B Schon L
Full Access

The Weil osteotomy has gained popularity for surgically treating patients with metatarsalgia, intractable plantar keratosis and/or metatarsophalangeal joint dislocation because of its simplicity and lack of complications. Different geometric configurations of the Weil osteotomy have been proposed to reduce plantar pressure. In a dynamic cadaver model, these different geometric configurations of the Weil osteotomy did not significantly alter plantar pressure. Metatarsal head resection was required to significantly reduce plantar pressure.

The purpose of this study was to evaluate the effect of different geometric configurations of the Weil osteotomy on the plantar pressures in a dynamic cadaver model.

Different geometric configurations of the Weil osteotomy have been proposed to decrease plantar pressure, but in a dynamic cadaver model, these modifications did not significantly alter plantar pressure. Metatarsal head resection was required to significantly reduce plantar pressure.

The plantar translation of the metatarsal head occurring with a more oblique Weil osteotomy compared to a standard Weil osteotomy did not significantly increase plantar pressure in a dynamic cadaver model. The addition of a 4 mm slice resection did not significantly reduce pressure. Metatarsal head resection was required to significantly reduce pressure (p=0.02).

Ten specimens (5 matched pairs of cadaver lower extremities) were tested. Each pair of specimens had an oblique Weil osteotomy performed on one side, and a standard (parallel) Weil osteotomy on the other. Then, a 4 mm slice resection, and metatarsal head resection were performed sequentially. The plantar pressures were measured with an F scan in-shoe sensor while cyclically loaded to 700 N at a frequency of 1 Hz in intact specimens, and after each intervention.

The different geometric configurations of the Weil osteotomy did not significantly alter plantar pressure; metatarsal head resection was required to significantly reduce pressure. The Weil osteotomy reliably reduces dislocated metatarsophalangeal joints. The angle of the osteotomy does not affect plantar pressure. Further study in a dynamic model is required to identify other factors, which affect plantar pressure.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 306 - 306
1 Mar 2004
Stamatis E Lau J Parks B Schon L
Full Access

Aims: To evaluate the effect of different geometric conþgurations of the Weil osteotomy on the plantar pressures in a dynamic in vitro cadaver model. Methods: Ten specimens consisting of 5 matched pairs of cadaver lower extremities were tested. Each pair of specimens had an oblique Weil osteotomy with 5 mm shift performed on one side, and a standard (parallel) Weil osteotomy with 5 mm shift on the other. Then, a 4 mm slice resection, and metatarsal head resection were performed sequentially on each specimen. The plantar pressures were measured while cyclically loaded to 700 N at a frequency of 1 Hz with a F scan in-shoe sensor in intact specimens, and after each intervention. Results: This is the þrst study to demonstrate that the plantar translation of the metatarsal head occurring with a more oblique Weil osteotomy compared to a standard (parallel) Weil osteotomy did not signiþcantly increase plantar pressure in a dynamic in vitro cadaver model. Furthermore, the addition of a 4 mm slice resection did not signiþcantly unload the metatarsal head. Metatarsal head resection was required to signiþcantly unload the metatarsal head (p=0.02). Conclusions: The different geometric conþgurations of the Weil osteotomy did not signiþcantly alter plantar pressures in a dynamic cadaver model. Metatarsal head resection was required to signiþcantly unload the metatarsal head. Future studies of the effect of metatarsal osteotomies on plantar pressure should include evaluation in a dynamic in vitrocadaver model to account for all factors, which determine the distribution of plantar pressure.