Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 38 - 38
1 Feb 2017
Rusk J Behzadi K
Full Access

Purpose

Current methods for inserting a press fit hemispherical metal-backed acetabular component within the acetabula are uncontrolled, relying on the surgeon to generate the necessary forces required for sufficient introduction. While previous studies have recorded impact forces of 2–3 kN necessary to seat an acetabular cup using visual observation[1], some researchers have observed users imparting as high as 8.9 kN of force[2]. The aim of this study is to quantify the forces required to generate optimal implant primary stability, as well as compare force delivery methods.

Method

The experiments were carried out using prepared bone substitute. A high frequency force sensor was rigidly mounted under the substitute to measure impact force and duration. An acetabular cup was inserted using successive reproducible impacts of varying magnitude (2.5 kg falling 17, 34, 43, 51, 68, or 85 mm). Impacts were repeated until the cup was no longer advancing. Each test recorded the number of impacts, maximum impact force, impact duration, and extraction force of the cup after insertion. The results were then compared against manual insertion (tapping) and high frequency vibratory insertion (50–500 Hz).