This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man between 2003 and 2019. Patient identifiers were used to link records to national mortality data, and to NJR data to identify subsequent re-revision procedures. Multivariable Cox proportional hazard models with restricted cubic splines were used to define associations between volume and outcome.Aims
Methods
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
We compared the risks of re-revision and mortality between two-stage revision surgery and single-stage revision surgery among patients with infected primary knee arthroplasty. Patients with a periprosthetic joint infection (PJI) of their primary knee arthroplasty, initially revised with a single-stage or a two-stage procedure in England and Wales between 2003 and 2014, were identified from the National Joint Registry. We used Poisson regression with restricted cubic splines to compute hazard ratios (HR) at different postoperative periods. The total number of revisions and re-revisions undergone by patients was compared between the two strategies.Aims
Methods
Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes. Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.Aims
Methods
Surgical site infection (SSI) is a common complication of surgery
with an incidence of about 1% in the United Kingdom. Sutures can
lead to the development of a SSI, as micro-organisms can colonize
the suture as it is implanted. Triclosan-coated sutures, being antimicrobical,
were developed to reduce the rate of SSI. Our aim was to assess
whether triclosan-coated sutures cause a reduction in SSIs following
arthroplasty of the hip and knee. This two-arm, parallel, double-blinded study involved 2546 patients
undergoing elective total hip (THA) and total knee arthroplasty
(TKA) at three hospitals. A total of 1323 were quasi-randomized
to a standard suture group, and 1223 being quasi-randomized to the
triclosan-coated suture group. The primary endpoint was the rate
of SSI at 30 days postoperatively.Aims
Patients and Methods
This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation.Aims
Patients and Methods
The objective of this study was to assess all evidence comparing the Thompson monoblock hemiarthroplasty with modular unipolar implants for patients requiring hemiarthroplasty of the hip with respect to mortality and complications. A literature search was performed to identify all relevant literature. The population consisted of patients undergoing hemiarthroplasty of the hip for fracture. The intervention was hemiarthroplasty of the hip with a comparison between Thompson and modular unipolar prostheses. Pubmed, Embase, CINAHL, Web of Science, PROSPERO and the Cochrane Central Register of Controlled Trials. The study designs included were randomised controlled trials (RCTs), well designed case control studies and retrospective or prospective cohort studies. Studies available in any language, published at any time until September 2015 were considered. Studies were included if they contained mortality or complications.Objectives
Methods
We have increased the dose of tranexamic acid (TXA) in our enhanced total joint recovery protocol at our institution from 15 mg/kg to 30 mg/kg (maximum 2.5 g) as a single, intravenous (IV) dose. We report the clinical effect of this dosage change. We retrospectively compared two cohorts of consecutive patients undergoing total hip arthroplasty (THA) or total knee arthroplasty (TKA) surgery in our unit between 2008 and 2013. One group received IV TXA 15 mg/kg, maximum 1.2 g, and the other 30 mg/kg, maximum 2.5 g as a single pre-operative dose. The primary outcome for this study was the requirement for blood transfusion within 30 days of surgery. Secondary measures included length of hospital stay, critical care requirements, re-admission rate, medical complications and mortality rates.Objectives
Methods
A fracture of the hip is the most common serious orthopaedic
injury, and surgical site infection (SSI) is one of the most significant
complications, resulting in increased mortality, prolonged hospital
stay and often the need for further surgery. Our aim was to determine
whether high dose dual antibiotic impregnated bone cement decreases the
rate of infection. A quasi-randomised study of 848 patients with an intracapsular
fracture of the hip was conducted in one large teaching hospital
on two sites. All were treated with a hemiarthroplasty. A total
of 448 patients received low dose single-antibiotic impregnated
cement (control group) and 400 patients received high dose dual-antibiotic impregnated
cement (intervention group). The primary outcome measure was deep
SSI at one year after surgery.Aims
Patients and Methods
Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an ODEP rating despite over 50 years of clinical use, likely due to the paucity of implant survival data. Nationally, adherence to these guidelines is varied as there is debate as to which prosthesis optimises patient outcomes. This study design is a multi-centre, multi-surgeon, parallel, two arm, standard-of-care pragmatic randomised controlled trial. It will be embedded within the WHiTE Comprehensive Cohort Study (ISRCTN63982700). The main analysis is a two-way equivalence comparison between Hemi-Thompson and Hemi-Exeter polished taper with Unitrax head. Secondary outcomes will include radiological leg length discrepancy measured as per Bidwai and Willett, mortality, re-operation rate and indication for re-operation, length of index hospital stay and revision at four months. This study will be supplemented by the NHFD (National Hip Fracture Database) dataset.Background
Design
We wanted to investigate regional variations in the organisms
reported to be causing peri-prosthetic infections and to report
on prophylaxis regimens currently in use across England. Analysis of data routinely collected by Public Health England’s
(PHE) national surgical site infection database on elective primary
hip and knee arthroplasty procedures between April 2010 and March
2013 to investigate regional variations in causative organisms.
A separate national survey of 145 hospital Trusts (groups of hospitals
under local management) in England routinely performing primary
hip and/or knee arthroplasty was carried out by standard email questionnaire.Objectives
Methods
The National Institute for Health and Clinical
Excellence (NICE) has thus far relied on historical data and predominantly
industry-sponsored trials to provide evidence for venous thromboembolic
(VTE) prophylaxis in joint replacement patients. We argue that the
NICE guidelines may be reliant on assumptions that are in need of
revision. Following the publication of large scale, independent
observational studies showing little difference between low-molecular-weight
heparins and aspirin, and recent changes to the guidance provided
by other international bodies, should NICE reconsider their recommendations? Cite this article:
To review the current best surgical practice and detail a multi-disciplinary
approach that could further reduce joint replacement infection. Review of relevant literature indexed in PubMed.Objectives
Methods
We wished to determine the most accurate and reliable technique for insertion of tibial prostheses, with tibial resection guided by either intramedullary (IM) or extramedullary (EM) alignment jigs. 135 consecutive AGC cemented total knee replacements in 126 patients in a single unit were performed by, or directly supervised by, four consultant surgeons. Ethical approval and patient consent was obtained. Intramedullary alignment was used for the femoral cuts and patients were randomised at the time of operation to have either IM or EM guides for resection of the proximal tibia, cut with a zero degree posterior slope in both. The protocol only entered patients into the trial if their knees were suitable for use with both IM and EM tibial alignment although, in the event, no patients were excluded. Long leg radiographs (standing hip to ankle) were taken by a standardised method three months after the surgery. A blinded assessor, unaware of the alignment method used, evaluated acceptable films and measured tibial component alignment. The proportion of tibial prostheses aligned within two degrees of 90 was the endpoint of the study. Of the 135 knees 100 suitable x-rays were assessed. Correct tibial alignment was more likely in the IM group (85%) than the EM group (65%), p=0. 019. Though mean alignment was similar, variation (standard deviation) was less in the IM group (2. 0 vv 2. 2). In the AGC knee, intramedullary alignment guides are superior to extramedullary guides for alignment of the tibial prosthesis. We recommend the routine use of intramedullary tibial alignment.