The assessment of leg length is essential for planning the correction of deformities and for the compensation of length discrepancy, especially after hip or knee arthroplasty. CT scan measures the “anatomical” lengths but does not evaluate the “functional” length experienced by the patients in standing position. Functional length integrates frontal orientation, flexion or hyperextension. EOS system provides simultaneously AP and lateral measures in standing position and thus provides anatomical and functional evaluations of the lower limb lengths. The objective of this study was to measure 2D and 3D anatomical and functional lengths, to verify whether these measures are different and to evaluate the parameters significantly influencing these potential differences 70 patients without previous surgery of the lower limbs (140 lower extremities) were evaluated on EOS images obtained in bipodal standing position according to a previously described protocol. We used the following definitions: anatomical femoral length between the center of the femoral head (A) and center of the trochlea (B) anatomical tibial length between the center tibial spine (intercondylar eminence) (C) and the center of the ankle joint (D) functional length is AD global anatomical length is AB + CD Other parameters measured are HKA, HKS, femoral and tibial mechanical angles (FMA, TMA), angles of flexion or hyperextension of the knee, femoral and tibial torsion, femoro-tibial torsion in the knee, and cumulative torsional index (CTI). All 2D et3D measures were evaluated and compared for their repeatability.Introduction
Material and Methods
The gold standard for knee surgery is the restoration of the so-called «neutral mechanical alignment ». Recent literature as pointed out the patients with «constitutional varus »; in these cases, restoring neutral alignment could be abnormal and even undesirable. The same situation can be observed in patients with «constitutional valgus alignment ». To date, these outliers cases have only been explored focusing on the lower limb; the influence of the pelvic morphotype has not been studied. Intuitively, the pelvic width could be a significant factor. The EOS low dose imaging technique provides full body standing X-rays to evaluate the global anatomy of the patient. This work explores the influence of the pelvic parameters on the frontal knee alignment. – We included 170 patients (340 lower extremities). 2 operators performed measurements once per patient on AP X-rays. The classical anatomical parameters were: Femoral mechanical angle (FMA) Tibial mechanical angle (TMA) Hip knee shaft angle (HKS) Hip knee ankle angle (HKA) Femoral and tibial lengths The morphotype was evaluated by: the distances between the center of two femoral heads (FHD), between knees (KD) and between ankles (AD) the medial neck-shaft angle (MNSA) the femoral offset The horizontal distance between the limb mechanical axis (line passing from center of the femoral head to the center of the ankle) and the center of the knee was called the intrinsic mechanical axis deviation (IMAD) (fig 1). The horizontal distance between the pelvic mechanical axis (line from the center of the sacral plate to the center of the ankle) and the center of the knee was called the global mechanical axis deviation (GMAD) (fig 2). Inter-Operator Reliability was calculated with Intra-class Correlation Coefficient (ICC) and Inter-Reader Agreement was assessed with Bland-Altman test. A relationship between IMAD and GMAD to the other parameters was assessed using Pearson's correlation coefficient.Introduction
Material and methods
Coronal misalignment of the lower limbs is closely related to the onset and progression of osteoarthritis. In cases of severe genu varus or valgus, evaluating this alignment can assist in choosing specific surgical strategies. Furthermore, restoring satisfactory alignment after total knee replacement promotes longevity of the implant and better functional results. Knee coronal alignment is typically evaluated with the Hip-Knee-Ankle (HKA) angle. It is generally measured on standing AP long-leg radiographs (LLR). However, patient positioning influences the accuracy of this 2D measurement. A new 3D method to measure coronal lower limb alignment using low-dose EOS images has recently been developed and validated. The goal of this study was to evaluate the relevance of this technique when determining knee coronal alignment in a referral population, and more specifically to evaluate how the HKA angle measured with this 3D method differs from conventional 2D methods. 70 patients (140 lower extremities) were studied for 2D and 3D lower limb alignment measurements. Each patient received AP monoplane and biplane acquisition of their entire lower extremities on the EOS system according the classical protocols for LLR. For each patient, the HKA angle was measured on this AP X-ray with a 2D viewer. The biplane acquisition was used to perform stereoradiographic 3D modeling. Valgus angulation was considered positive, varus angulation negative. Student's T-test was used to determine if there was a bias in the HKA angle measurement between these two methods and to assess the effect of flexion/hyperextension, femoral rotation and tibial rotation on the 2D measurements. One operator did measurements 2 times.Introduction
Materials and methods
Accurate evaluation of femoral offset is difficult with conventional anteroposterior (AP) X-rays. Routine CT imaging is costly and exposes patients to a significant dose of radiation. The EOS® imaging system is an innovative slot-scanning radiography system that makes possible the acquisition of simultaneous and orthogonal AP and lateral images of the patient in standing position. These 2-dimensional (2D) images are equivalent to standard plane X-rays. Three-dimension (3D) reconstructions are obtained from these paired images according to a validated protocol. This prospective study explores for the first time the value of the EOS® imaging system for comparing measurements of femoral offset obtained from 2D images and 3D reconstructions. Following our standard protocol, we included a series of 100 patients with unilateral total hip arthroplasty (THA). The 2D offset was measured on the AP view with the same protocol as for standard X-rays. The 3D offset was calculated from the reconstructions based on the orthogonal AP and lateral views. Reproducibility and repeatability studies were conducted for each measurement. We compared the 2D and 3D offsets for both hips (with and without THA).Introduction
Materials and Methods
The viscoelastic lumbar disk prosthesis ESP is an innovative one-piece deformable but cohesive interbody spacer; it provides 6 full degrees of freedom about the 3 axes including shock absorption. The prosthesis geometry allows limited rotation and translation with resistance to motion (elastic return property) aimed at avoiding overload of the posterior facets. The rotation center can vary freely during motion. It thus differs substantially from current prostheses. This study reports the results of a prospective series of 120 patients who are representative of the current use of the ESP implant since 2006. The surgeries were performed by 2 senior surgeons. There were 73 women and 47 men in this group. The average age was 42 (27–60). The average body mass index was 24.2 kg/m2 (18–33). The implantation was single level in 89% of cases. 134 ESP prostheses were analyzed. Clinical data and X-rays were collected at the preoperative time and at 3, 6, 12, 24, and 36 months post-op. The functional results were measured using VAS, GHQ 28, ODI, SF-36, (physical component PCS and mental component MCS. The analysis was performed by a single observer who was independent from the selection of patients and from the surgical procedure.Introduction
Material and methods
The position and orientation of the lower extremities are fundamental for planning and follow-up imaging after arthroplasty and lower extremity osteotomy. But no studies have reported the reproducibility of measurements over time in the same patient, and experience shows variability of the results depending on the protocols for patient positioning. This study explores the reproducibility of measurements in the lower extremity with the patients in “comfortable standing position” by the EOS® imaging system. Two whole-body acquisitions were performed in each of 40 patients who were evaluated for a spine pathology. The average interval between acquisitions was 15 months (4–35 months). Patients did not have severe spine pathology and did not undergo any surgery between acquisitions. The “comfortable standing position” is achieved without imposing on the patient any specific position of the lower limbs and pelvis. All the measurements were performed and compared in both 2- and 3-dimensional images. Distances between the centers of the femoral heads and between the centers of the knees and ankles were measured from the front. The profile is shown by the flexion angle between the axis of the femur (center of the femoral head and the top of the line Blumensaat) and the axis of the tibia.Introduction
Materials and Methods
Femoral stem anteversion after total hip arthroplasty (THA) has always been assessed using CT scan in supine position. In this study, we evaluated the anteversion of the femoral prosthesis neck in functional standing position using EOS® technology with repeatability and reproducibility of the measurements. The data obtained were compared with conventional anatomic measurements. We measured the anteversion of the femoral prosthesis neck in 45 consecutive patients who had THA performed in nine hospitals. All measurements were obtained using the EOS® imaging system with patients in comfortable standing position. The orientation of the final vector representing the femoral neck was measured on 3-dimensional reconstructions. The anatomic femoral anteversion was calculated as in a transverse plane relative to the scanner and to the plane of the reconstructed bicondylar femoral segment (femoral prosthesis neck against the femoral condyles). Functional femoral anteversion (FFA) was measured in the horizontal plane relative to the frontal plane of the patient through the center of two femoral heads. FFA embodies true anteversion of the femoral prosthesis neck relative to the pelvis, representing the combined lower extremity anteversion.Introduction
Materials and Methods
Recent literature points out the potential interest of standing and sitting X-rays for the evaluation of THA patients. The accuracy of the anterior pelvic plane measures is questionable due to the variations in the quality of lateral standing and sitting X-rays. The EOS® (EOS imaging, Paris, France) is an innovative slot-scanning radiograph system allowing the acquisition of radiograph images while the patient is in weightbearing position with less irradiation than standard imagers. This study reports the “functionnal” positions of a 150 THA cohort, including the lateral orientation of the cups. The following parameters were measured: sacral slope (SS), pelvic tilt (PT), pelvic incidence (PI) and anterior pelvic plane (APP) sagittal inclination (ASI), frontal inclination (AFI) and planar anteversion (ANT). Irradiation doses were calculated in standing and sitting acquisitions. Variations of sagittal orientation of the cup were measured on lateral standing and sitting images. Descriptive and multivariate analysis were performed for the different parameters studied.Background
Methods
Hip-spine relationships should be better investigated in THP as lumbo-sacral orientation in the sagittal plane plays a critical role in the function of the hip joints. Lateral X-rays showing spine and hips together in standing, sitting or squatting positions characterize the adaptations of the sagittal balance and the functionnal interactions between hips and spine. Acetabular cup implantation has to be planned for frontal inclination, axial anteversion, and sagittal orientation. The later refers to the sacro-acetabular angle, key-point in the spine – hip relationships, and that is redefined by the surgeon at the time of implantation. Usual standard CT-sections are biased for evaluating acetabular anteversion. The conventional CT procedure does not refer to the pelvic bony frame and. the measured anteversion is a projected angle on a transverse plane, depending on the pelvic adaptation in lying position. This measured angle is often considered as anatomical anteversion, leading to some confusion. Therefore this angle is only a “functional” supine anteversion, reflecting the anterior opening angle of the acetabulum in a specific position. According to the sagittal orientation of the pelvis, the true functional acetabular orientation can virtually be assessed in various postures from adjusted CT-scan sections. The EOS™ low irradiation 2D-3D X-ray scanner is an innovative technology already used for global evaluation of the spine. This technology allows simultaneously “full body” frontal and lateral X-rays with the patient in standing, sitting or squatting positions; a tridimensionnal patient specific bone recontruction can be performed and the cup anteversion can be directly assessed according to the position. We investigated the lumbo-pelvic parameters influencing the tridimensionnal orientation of the acetabulum. We compared the data obtained for real postural situations using the EOS™ system and the measures from plane X Rays and classical CT scan cuts replicating standing, and sitting positions.368 patients with cementless THP were involved in a prospective follow-up protocol. Sacral slope and pelvic tilt, incidence angle, acetabular frontal and sagittal inclination were evaluated on AP and lateral standard XRays. Functionnal anteversion of the cup has been measured using a previously described protocol with CTscan cuts oriented according to standing and sitting sacral slope. The mean difference between CTscan and EOS™ system was 4,4° with comparable accuracy and reproductibility. Sacral slope decrease in sitting position was linked to anteversion increase (38,8° SD 5,4°). Sacral slope increase in standing position was linked to lower ante-version (31,7° SD 5,6°). The anatomical acetabular anteversion, the frontal inclination, and the sagittal inclination were functional parameter which significantly varied between the standing, sitting, and lying positions. We noticed that the acetabular parameters in lying position highly correlated to the one in standing position, while poorly correlated with the one in sitting position. The difference between the lying and the sitting positions was about 10°, 25°, and 15° for the cup anteversion (CA) and the frontal and sagittal inclinations (FI,SI) respectively. The poor correlation between the lying and sitting positions suggests that the usual CT scan protocol is biased and not fully appropriate for investigating the cases of posterior THP dislocation and subluxation, which happen in sitting position. On the contrary, a strong correlation was observed between lying and standing measurements with all the acetabular parameters (CA,FI, SI), suggesting that the classical CT assessment of the cup anteversion remains an interesting source of information in case of anterior THP Each patient is characterized by a morphological parameter, the incidence angle. High incidence angle is linked to low acetabular anteversion, increasing the instability risk and anterior impingement in sitting and squatting position; higher anteversion angles are observed in low incidence angle patients, leading to more internal rotation of the hip in any position. Lumbo-sacral orientation in the sagittal plane influences the tridimensionnal orientation of the acetabulum, especially for anteversion. Aging of the hip-spine complex is linked to progressive pelvic posterior extension. Impingement phenomenons, orientation of stripe wear zones and some instability situations can be interpreted according to those data. This study points out the opportunity to adjust the CT scan sections to the sacral slope in functional position for properly investigating the orientation of the acetabular cup, mainly in case of posterior dislocation. In addition, the mobility of the lumbo-sacral junction could be a crucial parameter in the mechanical functioning and the stability of a THP due to its impact on sacral slope and pelvic tilt. Therefore we also recommend doing dynamic lateral radiographs of the lumbo sacral junction in standing and sitting position for planning a THP implantation in order to detect stiff lumbosacral junction or sagittal pelvic malposition.
Current total disc prostheses are 2- or 3-pieces devices, including 1 or 2 bearing surfaces, and providing 3 or 5 degrees of freedom but with no, or very little, resistance. The ESP® is a one-piece deformable implant made of silicon and polycarbonate polyurethane elastomer securely fixed to titanium endplates. It allows limited rotation and translation with elastic return. This cushion without fixed rotation center achieves 6 degrees of freedom including shock absorption. An earlier attempt to use elastomers (Acroflex®) failed clinically due to the polymer. This highlights the need for accurate in-vitro fatigue testing and clinical evaluations. In-vitro fatigue testing with more than 40 millions cycles were performed on different samples for compression, flexion-extension bending, lateral bending, torsion and shear. A prospective trial was initiated in 2004 for L3L4, L4L5 and L5S1 levels. Total disc replacements have been performed in 153 lumbar levels through extra-peritoneal mini-invasive anterior approach. After in-vitro testing, microscopic examination showed that the polymer core remained unchanged without evidence of cracking or other degradation. Gravimetric analysis revealed insignificant changes in weight. The geometrical characteristics and the cohesion of the implants remained stable. After 3 years clinical experience, there was no device related complication, except one early revision for a post-traumatic implant migration. VAS and ODI scores improvements were equivalent to other published series. In-vitro fatigue testing and short term results of the innovative ESP® prosthesis demonstrate the reliability of the concept. The results are equivalent to other series with conventional implants.
Mechanical failure in total hip arthroplasty is usually due to aseptic loosening related to wear particles as seen with polyethylene bearing. Alumina has been proposed for avoiding wear problems. In vitro and mid-term clinical studies showed tribological advantages but early acetabular fixation issues. Since alumina on alumina bearing is currently used with new fixation techniques, updated evaluations of the ancient series are informative regarding the long-term tolerance of alumina in vivo. In this paper, we investigated 104 consecutive lumina on alumina cemented total hip arthroplasties (CER-VAER-OSTEAL, Roissy, France) implanted 20 years ago in 81 patients (from 1979 to 1983). Alumina femoral head was 32 mm in diameter. Alumina acetabular socket and titanium femoral stem were cemented. The clinical evaluation used Postel Merle d’Aubigné score. Radiological wear and appearance of osteolysis or loosening were noted for establishing actuarial curves. When accessible, histological samples from revision procedures were analyzed. Six infected cases were not taken into account later. The average follow-up was 11 years, reaching 18 years in 38 cases. Twenty-three hip were revised for changing 23 acetabular sockets, 12 femoral heads, and 1 femoral stem. We noted 1 femoral head fracture, 24 definite ace-tabular loosenings, 12 probable acetabular loosenings, and 3 definite femoral loosenings. Radiological acetabular osteolysis was present in 4 cases, always limited to De Lee zone 1, and associated with loosening. Radiological wear was below eye detection. Peri-prosthetic tissue showed non-specific histological reaction to cement particles. Survival rate at 20 years was 61.4% in term of revision (57.1% and 95.2% concerning acetabular and femoral defininte loosening). Beside the high rate of cemented fixation failure of the socket, loosened and non loosend cases showed an excellent tolerance of alumina on alumina bearing in the long-term, with minimal wear and osteolysis. This may also have protected the femoral component from complications.