header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 54 - 54
1 May 2021
Debuka E Wilson G Philpott M Thorpe P Narayan B
Full Access

Introduction

IM (Intra Medullary) nail fixation is the standard treatment for diaphyseal femur fractures and also for certain types of proximal and distal femur fractures. Despite the advances in the tribology for the same, cases of failed IM nail fixation continue to be encountered routinely in clinical practice. Common causes are poor alignment or reduction, insufficient fixation and eventual implant fatigue and failure. This study was devised to study such patients presenting to our practice and develop a predictive model for eventual failure.

Materials and Methods

57 patients who presented with failure of IM nail fixation (± infection) between Jan 2011 – Jun 2020 were included in the study and hospital records and imaging reviewed. Those fixed with any other kinds of metalwork were excluded. Classification for failure of IM nails – Type 1: Failure with loss of contact of lag screw threads in the head due to backing out and then rotational instability, Type 2A: Failure of the nail at the nail and lag screw junction, Type 2B: Failure of the screws at the nail lag screw junction, Type 3: Loosening at the distal locking sites with or without infection. X-rays reviewed and causes/site of failure noted.