Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_8 | Pages 4 - 4
1 May 2021
Nicholas PRM Shields DW Mthethwa J Jamal B
Full Access

Gavril Ilizarov advocated a fine wire tension of between 900N and 1200N for circular frame construction. Wire tension can be achieved via a tensioning device or ‘Russian tensioning’ (a fixed wire lengthening around a bolt). There is limited information on the latter technique. This study explored the tensions achieved via Russian tensioning and reports the impact of a second wire on construct tension.

A single 160mm stainless-steel ring was constructed, then 1.8mm stainless steel wires were secured using a Russian fixation bolt and Russian tensioned with a 2nd bolt. The angle subtended by tensioning using the 2nd bolt was measured using a goniometer. Angles of 45°, 70° and 90° were repeated in triplicates, with wire tension measured using a calibrated tensiometer. A second, orthogonal wire was added and tensioned to the same angle. Tensions of both wires were remeasured and recorded. Unpaired t-tests were used to compare mean tensions. A value of p<0.05 was considered significant.

Tensioning at all angles was insufficient to achieve the target range of 900–1200N (range 99–110N). A second, orthogonal wire changed frame dynamics such that a 90° angle resulted in both wires achieving adequate tension (mean 1143N, SD 307N). Increases were significant across all tensioning angles (p=<0.002) however only biomechanically relevant for 90°.

Russian tensioning is insufficient with a single wire, however the addition of an orthogonal wire increases tension in both wires, reaching the target range at 90° deflection. Further study using wire tensioners is warranted, and also the impact of non-orthogonal wire constructs.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_19 | Pages 1 - 1
1 Nov 2017
Mthethwa J Keen J
Full Access

Enterococcus faecalis is a rare but recognized cause of prosthetic joint infection. It is notorious for formation of biofilm on prosthetic surfaces. We hypothesized that a ‘serum factor’ was responsible for transformation of E. faecalis from its planktonic form to a biofilm existence upon making contact with prostheses. Using a novel ‘proteomic approach’, we studied the protein expression profiles of this bacterium when grown on an artificial surface in a serum environment against a control.

E.faecalis 628 transconjugant formed by conjugation clinical strain (E55) and laboratory strain (JH2-2) was used to inoculate each of rabbit serum (RS) and Brain Heart Infusion (BHI) agar as a control and grown for 24 hours. Proteins were harvested for analysis in fractions including cell surface, membrane and cytosolic proteins. Recovered proteins were separated using 2-dimentional polyacrylamide gel electrophoresis (2D PAGE). Gels were stained and spots of interest harvested. These were analyzed using MALDI mass spectrometry followed by peptide mass fingerprinting using online database searches.

Two surface exclusion proteins Sea1 and PrgA were only expressed from the serum culture. These proteins are both encoded by genes very close to the gene for enterococcal aggregation substance PrgB, which plays an integral role in biofilm formation. PrgA and PrgB are both encoded by the prgQ operon and hence expressed simultaneously upon activation of the operon.

This tendency for serum only protein expression suggests the possibility of a pheromone-like activator in serum that could be a potential therapeutic target for management of biofilm associated E. faecalis prosthetic infections.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_31 | Pages 32 - 32
1 Aug 2013
Mthethwa J Hawkins A
Full Access

Despite widespread use, the benefit of knee arthroscopy for symptomatic osteoarthritis (OA) remains controversial. The theoretical benefit of removal of particulate debris and washout of inflammatory cytokines has not been supported by strong evidence. Arguments exist for its short term benefit in well selected patients. We sought to determine if arthroscopy provided any short term symptomatic relief in patients with a clinical diagnosis of OA in our unit.

A total of 20 patients were listed for routine arthroscopy over a one year period for OA. Mean age was 60 (range 48–74) years and 3 in 5 patients were female. 9 patients were listed by registrars, 6 by locum consultants and 5 by substantive consultants. One procedure was cancelled on the day of surgery due to lack of indication, with 19 knees proceeding to surgery. There was evidence of significant arthritis in 17 knees, 6 of which had associated degenerate meniscus tears. Two knees had meniscus tears without significant arthritis. All patients had washout and debridement and in addition, 8 partial menisectomies were carried out together with 3 loose body removals. Patients were followed up after an average of 12 weeks. The 2 patients (both male, mean age 52) with meniscus tears in the absence of significant arthritis fully recovered. Both had symptoms of true locking. 2 patients with loose bodies also fully recovered. With the exception of 2 patients with partial relief, the remainder had no symptomatic relief. One reported a worsening of symptoms and progressed to total knee arthroplasty.

These results suggest that arthroscopic debridement and washout plays a limited role in the short term symptomatic relief of OA. Surgeons should limit its use to younger patients with mild radiographic disease and true mechanical symptoms who are otherwise unsuitable for arthroplasty and not keen on undergoing joint conserving surgery.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_31 | Pages 33 - 33
1 Aug 2013
Mthethwa J Hawkins A
Full Access

Magnetic resonance imaging (MRI) is a useful diagnostic tool in evaluating meniscus pathology in the knee. Data from available literature suggests sensitivity and specificity rates around 90% when compared to the gold standard findings at knee arthroscopy. We sought to evaluate the sensitivity, specificity and precision rate (positive predictive value) of MRI at diagnosing meniscus tears within our unit.

A retrospective audit of a total of 79 MRI reports and arthroscopic findings spanning a one year period was carried out. There were 66 positive MRI reports and 13 negative reports. There were 6 false positives 4 false negatives when compared to arthroscopic findings. The sensitivity of MRI for detecting meniscus tears was 93.7% with 60 out of 64 tears detected. All 4 false negatives also had at least grade III osteoarthritic changes at arthroscopy. Specificity was rather low at 60% with MRI reporting 6 tears (false positives) out of 15 patients who had no tears found at arthroscopy. The positive predictive value (precision rate) of MRI detecting tears was 90.9%.

This data shows that MRI in our unit has a comparable high sensitivity to that in various literature making it a useful tool at ruling out disease with a negative result in the clinical setting. A more useful parameter in the clinical setting is its high precision rate when faced with a positive result. However, its specificity is much lower than that in most published data. A total of 6 tears on MRI turned out not to be on arthroscopy meaning patients could have been subjected to an avoidable invasive procedure in the absence of any other indication. This highlights the importance of obtaining reports from experienced musculoskeletal radiologists and the need for surgeons to review MRI images and match them to clinical information prior to subjecting patients to surgery.