The definition of a robust and reproducible reference frame is the first step in order to perform a consistent biomechanical test [1]. The aims of the current study was to define an anatomical reference frame for pelvis which can also be applied to a hemi-pelvis. A robust alignment method was sought so as to replicate the anatomical pose during The anatomical reference frame was derived from the anterior pelvic plane [2] and was adapted for The alignment procedure represented an improved version of a previous procedure [3]. The hemi-pelvis has been positioned on blocks of plasticine close to the landmarks on a 5-screws adjustable plate. Three steps were performed: position of the landmarks at the same height; alignment of the PSIS and PT on a horizontal line parallel to x- direction; position of the setup in front of a 6 degrees of freedom manipulator so that both the manipulator and the 5-screws adjustable plate are in the same reference frame. The manipulator was used to move the specimen in the following steps: clamping and lifting up of the specimen; rotation around x-axis in the posterior direction by 45°; rotation around y-axis in the medial direction by 90°; rotation around x-axis (antero-posterior direction) until PT and ASIS lay in the same vertical plane. Five operators performed the alignment of a male and a female hemi-pelvis, three times each. The repeatability of the current procedure was good, with uncertainties below 1.0° within the same operator, and of less than ±1.5° between operators for the male hemipelvis, and ±2° for the female one. Thanks to the good results the reference frame and the alignment procedure may be adopted for