Radial head implant over-lengthening, a common cause of capitellar wear and clinical failure, is difficult to diagnose using radiographs of the injured elbow. The purpose of this study was to determine if a novel measurement technique based on contralateral elbow radiographs, termed the RACER method, could be used to accurately estimate the magnitude of radial head implant over-lengthening. Part I of this study examined the side-to-side consistency of radiographic landmarks used in the measurement technique. Part II of this study validated the technique using simulated radial head implant over-lengthening in a cadaveric model. Part I: A side-to-side comparison of elbow joint dimensions was performed in 50 patients (100 radiographs). Part II: Radial head prostheses of varying lengths (0,+2mm,+4mm,+6mm,+8mm) were implanted in 4 paired cadaveric specimens (8 elbows). Radiographs were obtained and measurements were performed by 2 examiners blinded to implant size to determine if contralateral radiographs could diagnose and provide a valid estimate of the magnitude of implant over-lengthening. Intra and inter-rater reliability was determined.Purpose
Method