header advert
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 142 - 142
1 Mar 2013
Chen Y Kurosu S Lee Y Matsumoto H Koizumi Y Chiba A
Full Access

Introduction

Metal-on-metal (MOM) hip joints have regained a favor in arthroplasty since they own excellent wear resistance. In this study, wear tests by using a hip joint simulator were conducted with MOM bearings of specified 40 mm femoral heads. The influence of clearance on the wear behavior was investigated. Furthermore, an optimized radial clearance was estimated by lubricant film thickness and contact pressure analysis.

Materials and methods

Co-27Cr-5Mo-0.13N-0.05C (hereafter CCMN) alloy (mass %) was used. The ingots were vacuum induction melted, homogenized and hot forged successively. The microstructure shows equiaxed crystal grains with abundant annealing twins but no carbides.

Two groups of bearings, indicated as cr 1 and cr 2, were designed. The radial clearances for cr 1 and cr 2 were 37.9 and 148.7 μm, respectively. Wear tests were conducted in a hip joint simulator (INSTRON 8870) in Hanks' solution at 37±2°f. The force and 3-axile angle of movement were applied on the articulation according to ISO 14242-1 for 1.5 million cycles (Mc).

The contact pressures on the hip joints were also analyzed by using ABAQUS. The femoral heads were set 40 mm with radial clearances of 0–200 μm. Half models were set up and only the maximum force of 3 kN converted as pressure was applied as boundary condition.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 146 - 146
1 Mar 2013
Chiba A Kurosu S Koizumi Y Matsumoto H Lee Y
Full Access

Rapid manufacturing using laser beam and/or electron beam has been applied to fabrication of artificial hip and knee joints in quite recent years. In the electron beam melting (EBM) method, the high energy electron beam effectively melts the metal powder without creating flaws such as porosities or inclusions of oxide particles during building. Thus it is found that EBM technique for rapid manufacturing of artificial hip and knee joints processes a higher possibility as the next-generation methodology for fabrication of the medical devices such as hip and knee joints. In the present study, we focus on the EBM technique. The microstructures and mechanical properties of Co-29Cr-6Mo alloy with C and N additions, produced by using EBM method, were studied using X-ray diffraction, electron back scatter diffraction, transmission electron microscope (TEM), Vickers hardness tests, and tensile tests, focusing on the influences on the build direction and the various heat treatments after build. It is found that the microstructures for the as built specimens were changed from columnar (Fig. 1a) to eqiaxed grain structure (Fig. 1c) with average grain size of approximately 10–20 μm due to the heat treatment employing the reverse transformation from a lamellar (hcp + Cr2N) phase to an fcc phase. Our results will contribute to the development of biomedical Ni-free Co–Cr–Mo–N-C alloys, produced by EBM method, with refined grain size and good mechanical properties, without requiring any hot workings.

Fig. 1 Inverse pole figure (IPF) maps of microstructure of samples produced by EBM method, taken by EBSD. (a) as-built, (b) after aging treatment, (c) after reverse transformation heat treatment (RT-HT).