The mean pre and post-operative lumbar lordosis was 34.58 and 53.48 respectively. The mean sagittal rotation was 6.5 degrees at 5 year follow-up, while the mean translation was 0.83 mm. The mean AVM, MVM and PVM were 0.59 mm, −3.96 mm and 3.69 mm respectively at 5 year follow-up.
The following measurements were performed on the replaced motion segment using a lateral radiograph:
The anterior-posterior (AP) dimension of the end plates. Amount of subsidence. The distance between the TDA and the posterior and anterior borders of the vertebra bodies (to represent the extent of uncoverage of the endplate by the TDA). The AP dimension of the TDA metal endplate. The ratio between the actual and radiographic AP length of the metal endplate was calculated and utilized as the correction factor for the error of magnification on all other radiographic measurements.
At L4–L5 the mean subsidence was 1.48 mm (L4) and 0.56 mm (L5). Posterior uncoverage of L4 and L5 vertebrae were 4.81 and 2.22 mm, respectively. Subsidence of more than 1 mm was present in all cases where the posterior uncoverage of the end plate with the TDA was more than 2 mm (odds ratio: 5.7). Subsidence was non – progressive in all cases. An anatomic mismatch exists between L5 and S1 endplates in the AP dimension; in more than half the patients S1 is shorter than L5.