Harman M Schmitt S Roessing S Banks S Scharf H Hodge WA
Full Access

There is renewed interest in unicondylar knee replacements (UKR) to meet the increasing demand for less invasive surgical procedures for knee arthroplasty. UKR survivorship exceeds 85% at 10 years, with unconstrained (round-on-flat) designs showing significantly better survivorship than conforming designs. However, round-on-flat articulation shave the potential for poor wear performance and more conforming, mobile-bearing UKR designs have been advocated. The purpose of this study was to evaluate the wear performance of unconstrained UKR polyethylene bearings retrieved at revision knee arthroplasty.

Forty-two UKR (eight designs) were retrieved from 26 female and 16 male patients. Patient age averaged 73+10 (45–89) years and time in-situ averaged 7+4 (1–19) years. Revision reasons included loosening (45%), progressive osteoarthritis (17%), polyethylene wear (14%), instability (5%), and other (19%). Retrospective radiographic review of radiolucent lines and component alignment was completed. Polyethylene damage (severity score, 0–3) and location were measured using optical microscopy and digital image analysis.

81% of the polyethylene inserts had a concave deformation located on the central or posterior third of the articular surface, consistent with damage due to femoral component articulation. Abrasive damage on 29 (69%) inserts was consistent with impingement between the polyethylene and extra-articular cement or bone. There was delamination in the central region of 7 (17%) inserts and on the extreme posterior rim of 3 (7%) inserts. Severity score averaged 2.0+1.2 for abrasion and 0.5+1.0 for delamination. Radiographic component position was correlated with abrasive polyethylene damage.

Despite initial tibiofemoral incongruity and concerns of high contact stress, round-on-flat UKR offers a durable knee arthroplasty. Delamination was infrequent and did not correlate with time in-situ. Rather, polyethylene cold flow increased the contact area during in-vivo function. Rigorous attention to cement technique and component position may reduce the incidence of abrasive damage on UKR polyethylene inserts.