Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 37 - 37
1 Sep 2012
Kinzel V Scholes C Giuffrè B Coolican M Parker D
Full Access

Patient-matched instrumentation is advocated as the latest development in arthroplasty surgery. Custom-made cutting blocks created from preoperative MRI scans have been proposed to achieve perfect alignment of the lower limb in total knee arthroplasty (TKA). The aim of this study was to determine the efficacy of patient-specific cutting blocks by comparing them to navigation, the current gold standard.

25 TKA patients were recruited to undergo their surgery guided by Smith & Nephew Visionaire Patient-Matched cutting blocks. Continuous computer navigation was used during the surgery to evaluate the accuracy of the cutting blocks. The blocks were assessed for the fit to the articular surface, as well as alignment in the coronal and sagittal planes, sizing, and resection depth. Actual postoperative alignment was then assessed by detailed CT scans following the Perth protocol, comparing the results with intraoperative measurements.

All patient-matched cutting blocks were a good fit intra-operatively. Significant differences (p<0.05) in the resection depths of the distal femur and tibial plateau were observed between the cutting blocks and computer navigation for the medial compartment. Cutting block alignment of the femur and tibia in the coronal and sagittal planes also differed significantly (p<0.05) to navigation measurements. In addition, intraoperative assessment of sagittal femoral alignment differed to planned alignment by an average of 4.0 degrees (+/−2.3).

This study suggests the use of patient-matched cutting blocks is not accurate, particularly in the guidance of the sagittal alignment in total knee arthroplasty. Despite this technique creating well fitting cutting blocks, intraoperative monitoring, validated by postoperative CT scans, revealed an unacceptable degree of potential limb mal-alignment, resulting in increased outliers particularly when compared with standard computer navigation.