Vol. 104-B, Issue SUPP_13 | Pages 37 - 37
1 Dec 2022
Fleet C de Casson FB Urvoy M Chaoui J Johnson JA Athwal G
Full Access

Knowledge of the premorbid glenoid shape and the morphological changes the bone undergoes in patients with glenohumeral arthritis can improve surgical outcomes in total and reverse shoulder arthroplasty. Several studies have previously used scapular statistical shape models (SSMs) to predict premorbid glenoid shape and evaluate glenoid erosion properties. However, current literature suggests no studies have used scapular SSMs to examine the changes in glenoid surface area in patients with glenohumeral arthritis. Therefore, the purpose of this study was to compare the glenoid articular surface area between pathologic glenoid cavities from patients with glenohumeral arthritis and their predicted premorbid shape using a scapular SSM. Furthermore, this study compared pathologic glenoid surface area with that from virtually eroded glenoid models created without influence from internal bone remodelling activity and osteophyte formation. It was hypothesized that the pathologic glenoid cavities would exhibit the greatest glenoid surface area despite the eroded nature of the glenoid and the medialization, which in a vault shape, should logically result in less surface area.

Computer tomography (CT) scans from 20 patients exhibiting type A2 glenoid erosion according to the Walch classification [Walch et al., 1999] were obtained. A scapular SSM was used to predict the premorbid glenoid shape for each scapula. The scapula and humerus from each patient were automatically segmented and exported as 3D object files along with the scapular SSM from a pre-operative planning software. Each scapula and a copy of its corresponding SSM were aligned using the coracoid, lateral edge of the acromion, inferior glenoid tubercule, scapular notch, and the trigonum spinae. Points were then digitized on both the pathologic humeral and glenoid surfaces and were used in an iterative closest point (ICP) algorithm in MATLAB (MathWorks, Natick, MA, USA) to align the humerus with the glenoid surface. A Boolean subtraction was then performed between the scapular SSM and the humerus to create a virtual erosion in the scapular SSM that matched the erosion orientation of the pathologic glenoid. This led to the development of three distinct glenoid models for each patient: premorbid, pathologic, and virtually eroded (Fig. 1). The glenoid surface area from each model was then determined using 3-Matic (Materialise, Leuven, Belgium).

Figure 1. (A) Premorbid glenoid model, (B) pathologic glenoid model, and (C) virtually eroded glenoid model.

The average glenoid surface area for the pathologic scapular models was 70% greater compared to the premorbid glenoid models (P < 0 .001). Furthermore, the surface area of the virtual glenoid erosions was 6.4% lower on average compared to the premorbid glenoid surface area (P=0.361).

The larger surface area values observed in the pathologic glenoid cavities suggests that sufficient bone remodelling exists at the periphery of the glenoid bone in patients exhibiting A2 type glenohumeral arthritis. This is further supported by the large difference in glenoid surface area between the pathologic and virtually eroded glenoid cavities as the virtually eroded models only considered humeral anatomy when creating the erosion.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 85 - 85
1 Dec 2022
Fleet C McNeil D Trenholm JAI Johnson JA Athwal G
Full Access

Massive irreparable rotator cuff tears often lead to superior migration of the humeral head, which can markedly impair glenohumeral kinematics and function. Although treatments currently exist for treating such pathology, no clear choice exists for the middle-aged patient demographic. Therefore, a metallic subacromial implant was developed for the purpose of restoring normal glenohumeral kinematics and function. The objective of this study was to determine this implant's ability in restoring normal humeral head position. It was hypothesized that (1) the implant would restore near normal humeral head position and (2) the implant shape could be optimized to improve restoration of the normal humeral head position.

A titanium implant was designed and 3D printed. It consisted of four design variables that varied in both implant thickness (5mm and 8mm) and curvature of the humeral articulating surface (high constraint and low constraint. To assess these different designs, these implants were sequentially assessed in a cadaver-based biomechanical testing protocol. Eight cadaver specimens (64 ± 13 years old) were loaded at 0, 30, and 60 degrees of glenohumeral abduction using a previously developed shoulder simulator. An 80N load was equally distributed across all three deltoid heads while a 10N load was applied to each rotator cuff muscle. Testing states included a fully intact rotator cuff state, a posterosuperior massive rotator cuff tear state (cuff deficient state), and the four implant designs. An optical tracking system (Northern Digital, Ontario, Canada) was used to record the translation of the humeral head relative to the glenoid in both superior-inferior and anterior-posterior directions.

Superior-Inferior Translation

The creation of a posterosuperior massive rotator cuff tear resulted in significant superior translation of the humeral head relative to the intact cuff state (P=0.016). No significant differences were observed between each implant design and the intact cuff state as all implants decreased the superior migration of the humeral head that was observed in the cuff deficient state. On average, the 5mm low and high constraint implant models were most effective at restoring normal humeral head position to that of the intact cuff state (-1.3 ± 2.0mm, P=0.223; and −1.5 ± 2.3mm, P=0.928 respectively).

Anterior-Posterior Translation

No significant differences were observed across all test states for anterior-posterior translation of the humeral head. The cuff deficient on average resulted in posterior translation of the humeral head, however, this was not statistically significant (P=0.128). Both low and high constraint implant designs were found to be most effective at restoring humeral head position to that of the intact cuff state, on average resulting in a small anterior offset (5mm high constraint: 2.0 ± 4.7mm, P=1.000; 8mm high constraint: 1.6 ± 4.9mm, P=1.000).

The 5mm high constraint implant was most effective in restoring normal humeral head position in both the superior-inferior and anterior-posterior directions. The results from this study suggest the implant may be an effective treatment for restoring normal glenohumeral kinematics and function in patients with massive irreparable rotator cuff tears. Future studies are needed to address the mechanical efficiency related to arm abduction which is a significant issue related to patient outcomes.