Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 260 - 260
1 Dec 2013
Cartner J Aldinger P Fessenden M Li C
Full Access

INTRODUCTION:

The use of tapered junctions in primary hip arthroplasty has excellent results. Large heads are being used to mitigate dislocation and optimize range of motion. The prevalence of larger heads, coupled with recent findings regarding corrosion artifacts at tapered surfaces, has spurred growing interest when considering revision rates. The purpose of this study was to determine if correlations exist between severity of corrosion artifacts and head size, head offset, time in vivo, orhead material in a 15 year retrieval database.

METHODS:

Retrieved hip arthroplasty devices with CoCrMo or oxidized zirconium (OxZr) heads were investigated for corrosion artifacts in this study. Female tapered surfaces were scored independently by a panel according to the Goldberg system for assessment of corrosion. Exclusion criteria included less than 1 week in vivo, ceramic taper, and modular proximal stem sleeves. Evaluation was performed on only stem/head taper junctions. SEM analyses and a newly developed method of vertical straightness deviation (VSD) were used on a subset of devices as an additional measure to quantify corrosion within the taper contact region by measuring depth of material loss.