Advertisement for orthosearch.org.uk
Results 1 - 8 of 8
Results per page:
Bone & Joint Research
Vol. 12, Issue 3 | Pages 155 - 164
1 Mar 2023
McCarty CP Nazif MA Sangiorgio SN Ebramzadeh E Park S

Aims

Taper corrosion has been widely reported to be problematic for modular total hip arthroplasty implants. A simple and systematic method to evaluate taper damage with sufficient resolution is needed. We introduce a semiquantitative grading system for modular femoral tapers to characterize taper corrosion damage.

Methods

After examining a unique collection of retrieved cobalt-chromium (CoCr) taper sleeves (n = 465) using the widely-used Goldberg system, we developed an expanded six-point visual grading system intended to characterize the severity, visible material loss, and absence of direct component contact due to corrosion. Female taper sleeve damage was evaluated by three blinded observers using the Goldberg scoring system and the expanded system. A subset (n = 85) was then re-evaluated following destructive cleaning, using both scoring systems. Material loss for this subset was quantified using metrology and correlated with both scoring systems.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 19 - 19
1 Oct 2018
Campbell P Park S Lu Z Ebramzadeh E
Full Access

Introduction

In 2010, a widely used metal-on-metal hip implant was voluntarily recalled from the market by the manufacturer. Our lab performed detailed retrieval analysis of 548 of the explanted devices and 165 periprosthetic tissue specimens. We reported a high degree of variability in the multiple measurements that we performed on the components, including volumetric wear.

Other studies using semi-quantitative tissue grading methods to describe the histology of the periprosthetic tissues from metal-on-metal hips have looked for relationships between component wear and histological features. Grammatopoulos et al found higher ALVAL scores in 45 hip resurfacing patients with pseudotumors compared to 11 without pseudotumors and a moderately positive correlation between wear and the histological rankings for tissue necrosis and lymphocytic response.

We examined correlations between the component wear measurements from retrieval analysis and the tissue features in this large group of one metal-on-metal hip design.

Materials and Methods

Tissues were processed routinely into paraffin sections, stained with H&E and characterized with light microscopic semi-quantitative scores for ALVAL features and for the number of macrophages and lymphocytes, extent of metal debris and the estimated percentage of necrosis per slide. Spearman correlation analysis was conducted to evaluate any correlations between retrieval wear measurements and histological findings.


Bone & Joint Research
Vol. 6, Issue 2 | Pages 90 - 97
1 Feb 2017
Rajfer RA Kilic A Neviaser AS Schulte LM Hlaing SM Landeros J Ferrini MG Ebramzadeh E Park S

Objectives

We investigated the effects on fracture healing of two up-regulators of inducible nitric oxide synthase (iNOS) in a rat model of an open femoral osteotomy: tadalafil, a phosphodiesterase inhibitor, and the recently reported nutraceutical, COMB-4 (consisting of L-citrulline, Paullinia cupana, ginger and muira puama), given orally for either 14 or 42 days.

Materials and Methods

Unilateral femoral osteotomies were created in 58 male rats and fixed with an intramedullary compression nail. Rats were treated daily either with vehicle, tadalafil or COMB-4. Biomechanical testing of the healed fracture was performed on day 42. The volume, mineral content and bone density of the callus were measured by quantitative CT on days 14 and 42. Expression of iNOS was measured by immunohistochemistry.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 58 - 58
1 Feb 2017
Campbell P Yuan N Ebramzadeh E
Full Access

Young osteoarthritic male patients have been considered the ideal candidates for Metal-on-Metal (MoM) hip resurfacing arthroplasty (HRA), based on generally good long term results. In contrast, hip resurfacing in young female patients has become controversial. Recently, one implant manufacturer withdrew 46mm and smaller components, citing poorer than expected 10 year outcomes in females with smaller HRAs. Whether this difference is related to gender or to component size is still debated. Possible reasons for higher failure rates reported in females include higher rates of hip dysplasia, poorer bone quality and the risk of higher wear in some smaller sized implants with low cup coverage angles.

We reviewed HRA revision specimens with the aim of comparing mode of failure, time to revision, femoral cement characteristics and acetabular bone attachment in specimens larger and smaller than 46mm and from male versus female patients.

Methods

The study included all of the MoM HRA devices in our collection. Of the 284 hip resurfacing devices with complete clinical information, 131 were from male and 153 from female patients. Femoral sizes ranged from 36 – 58mm, median and mode 46mm; median size in females was 44 and 50mm in males. Time to failure ranged from 1 to 178 months, median 24 mos. Seven designs were represented but the majority were Conserve Plus (n=105 WMT, USA) and BHR (n=78 Smith & Nephew, USA) which differ in cementing technique. 131 femoral components were sectioned and the width of the cement mantle and the amount of cement in the head were measured. Where available, the amount of bone attached to the cup porous surface (n=91), tissue ALVAL scores (n=75) and bearing wear depth (n=138) were included in the multivariate analysis.

Results

As a function of gender, there were no significant differences in time to revision, cement measurements or ALVAL scores. Wear depth was significantly higher in females (femoral 41um vs 21um; cup 50um vs 16um, p=0.05). As a function of size (46 and less = small), the <46mm group had a slightly shorter time to revision, 30 vs 38 months, p=0.04). Bone ingrowth ranged from 0 to 60% (Figure 1) and significantly less bone attachment was noted in both the smaller and larger components (p = 0.001). Other characteristics were similar in both groups. When wear-related failure modes (cup malposition, lysis, high ions) were compared, no differences between male and female or large vs small were found. The amount of cement in the femoral heads covered a wide range but femoral loosening or fracture rates were not different as a function of size or gender.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 65 - 65
1 May 2016
Campbell P Kung M Ebramzadeh E Van Der Straeten C DeSmet K
Full Access

Bone ingrowth fixation of large diameter, beaded cobalt chromium cups is generally considered to be reliable but this is typically judged radiographically. To date, implant retrieval data of attached bone has been limited. This study evaluated correlations between the pre-revision radiographic appearance and the measured amount of bone attachment on one design of porous coated cup.

Methods

Twenty-six monoblock, CoCr Birmingham Hip Resurfacing (BHR, Smith and Nephew, TN, USA) cups with macroscopic beads and hydroxyapatite coating were studied. Seventeen were revised for acetabular malposition with the remainder revised for femoral loosening (4), pain (1), infection (1), dislocation (1) or lysis (2). Median time to revision was 35 months (10 – 70 months). Ten patients were female; the median age of all patients was 54 years. The pre-revision radiographs were visually ranked for cup-bone integration as follows: 0 = none, 1 = < 50%, 2 = 50 – 75%, 3 = 76 – 95%, 4 = > 96% integration. Rankings were made for the superior and inferior aspects, without knowledge of the appearance of bone on the retrievals. The revised cups were photographed at an angle so the dome and the cup periphery were visualized. The area of bone in four equal segments in each of the superior and inferior aspects was measured with image analysis software. A probe was used to differentiate bone from soft tissue. Only bone that covered the beads was counted. Correlation coefficients were calculated for the radiographic and image analysis data.

Results

Radiographically, most cups were assessed as having more than 50% of bone attachment and 7 cups were ranked as having almost total integration with bone. Only 2 cups were assessed radiographically as fully loose. Measured total bone attachment ranged from none to 55%. Superior and inferior percent ingrowth were highly correlated (corr=0.68, p<0.001) but there was no correlation between percent bone and x-ray rank (inferior corr=0.01, p=0.96; superior corr=0.23, p=0.26). There was no correlation between cup malpositioning as a reason for revision and x-ray integration ranking (superior p=0.34; inferior p=0.80).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 391 - 391
1 Dec 2013
Lazennec JY Brusson A Ebramzadeh E Clarke I
Full Access

Unlike conventional radiographic methods, the newly introduced EOS system provides simultaneously-synchronized anteroposterior (AP) and true-lateral (LAT) x-ray images. EOS offers considerable potential for calculating parameters such as true femoral and acetabular angular positioning, impingement sites, and also for measuring wear in polyethylene cups. In this study we used THA wear-simulation fixtures to assess 3D-wear in polyethylene cups using EOS algorithms.

Material and methods

A validated phantom apparatus was used to simulate values of three-dimensional wear, controlled in the 3 directions (antero-posterior, medio-lateral, cranio-caudal) using micrometers. (Figure 1)

24 simulations of wear with controlled amplitudes and directions were imaged using the biplane EOS slot-scanning system. Wear amplitudes were between 0 and 3464 μm. Using dedicated software, wear was measured by a 2D/3D matching of 3D spheres onto the 2D frontal and lateral radiographs, allowing the determination of the 3D coordinates of both the cup and femoral head centers and thus the calculation of a 3D wear vector. (Figure 2)

Measured wear vector were compared to real wear vectors in terms of amplitude and direction.3D wear vectors were measured twice by 3 independent observers (for a total of 144 measurements) in order to evaluate intra- and inter-observer reliability.

Results

There was a strong correlation between the measured wear amplitude and the real wear amplitude (Pearson's r = 0,99). Mean error when comparing wear measurement amplitude with real wear amplitude was 356 μm (SD = 127 μm). None of the 144 measurements presented an error over 1 mm.

The accuracy of wear direction evaluation was highly correlated with wear amplitude (Spearman's rho = 0,98), the measurement of 3D wear direction presenting an accuracy better than 15° for wear amplitudes over 1,5 mm. Intra-observer errors for wear amplitude were between 138 μm and 221 μm depending on the observer. Inter-observer error for wear amplitude was 333 μm.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 30 - 30
1 Mar 2013
Yoon J Duff ML Johnson A Takamura K Ebramzadeh E Campbell P Amstutz HC
Full Access

It has been suggested that metal ion levels are indicative of in vivo bearing performance of metal-on-metal hip replacements. A cobalt or Chromium level of 7μg/L or higher is proposed to be indicative of a bearing malfunction and the need for clinical intervention. Component design, size, acetabular orientation, patient gender and activity level have been suggested as factors leading to accelerated wear and elevated metal ions. The contact patch to rim (CPR) distance is a calculation that describes the distance from the point where a theoretical joint reaction force intersects the cup to the acetabular rim for a patient in standing position, dependent on the coverage, size, and orientation of the acetabular component. It has been suggested that CPR distance determines the hip joint susceptibility to edge loading, and the risk for increased wear and high ion levels (Langton et al JBJS Br 91: 2009). This study examined the effects of patient activity, gender, and CPR distance on serum metal ion concentrations in a series of patients treated with one type of metal on metal hip resurfacing arthroplasty (MMHRA) performed by one surgeon.

182 patients (73 females and 109 males)with a unilateral Conserve Plus (WMT, TN USA) MMHRA and had who had provided blood for metal ion analysis data from December 2000 to June 2011 were retrospectively studied. Only measurements made more than 12 months after surgery were included in order to exclude hips that had yet to reach steady-state wear. For patients with multiple draws, the most recent qualifying draw was used. Activity level was assessed by the UCLA activity score. The mean age was 51.5 years (20.0 to 77.5 years). The mean follow-up time for the last blood draw was 70 months (range, 12 to 165). Serum cobalt (CoS) and chromium (CrS) levels were analyzed using inductively coupled plasma mass spectrometry in a specialized trace element lab. Using acetabular abduction and anteversion measured by EBRA, component size, and reported coverage angle of the acetabular component, the CPR distance was calculated as previously described. Multiple logistic regression was performed to identify significant relationships between high metal ion levels (7 μg/L or greater) and gender, activity and CPR distance.

The median CoS level for the entire cohort was 1.13 μg/L (range, 0.15 to 175.30), and the median CrS level was 1.49 μg/L (range, 0.06 to 88.70). The average CPR distance was 13.8 mm (range, 3.2 to 22.1). There was a significant association between low CPR values and CoS and CrS. There was a 37-fold increase in the risk of CoS >7μg/L (p=0.005) and 11-fold increase in the risk of CrS > 7μg/L (p=0.003) when CPR distance was 10 mm or less. No associations were shown for gender and UCLA activity scores.

CPR distance was found to be a reliable predictor of ion levels > 7μg/L and appears to be a useful indicator to evaluate the multi-factorial process of edge-loading and wear. Patients with a low CPR distance should be monitored for increased metal ion levels.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 150 - 150
1 Mar 2009
Billi F Aust S Ebramzadeh E
Full Access

TKR backside wear studies have concluded that, compared to rough trays, polished trays decrease total amount of backside wear by 80% to 87%. However, size and volumetric concentration of sub-micron-sized polyethylene particles are critical factors for macrophage-mediated osteolysis. We assessed the size and morphology of polyethylene wear debris from TKR backside wear simulations comparing polyethylene fretting against polished and blasted metal surfaces. A 3-station fretting wear simulator reproduced loads and motions typical of the backside of fixed-bearing inserts of TKRs. 5-million cyclic experiments combined low (50μm) or high (200μm) linear motion with +3o rotational motion. Load profile was double-peak Paul curve (peak 10MPa). Eight 3-station experiments measured polyethylene wear against blasted or polished metal surfaces of Ti6Al4V or CoCr. Polyethylene particles were isolated from serum following gradient separation and filtration on 0.01μm polycarbonate filters. Using SEM analysis, average 200 particles per sample were characterized with Meta-morph™ image analysis software. Concentration of submicron particles in the debris from rough surfaces was 31–32% under 50μm motion, 28–30% under 200μm. Surprisingly, this concentration from polished surfaces was substantially greater: 69–78% (50μm), and 57–63% (200μm). However, total poly wear against rough surfaces was 0.45–1.63mm3/

Mcycles, and 0–0.35mm3/Mcycles against polished. Taking this into account, the volume of submicron particles from polished surfaces is less than 0.1mm3/ Mcycles and from rough surfaces between 0.1 and 0.45mm3/Mcycles.

In conclusion, although polished metal trays produce up to five times less wear than blasted surfaces, they may also lead to an increase in the osteolytic potential of the polyethylene debris.