Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 213 - 213
1 Jan 2013
Ajuied A Norris M Wong F Clements J Back D Davies A
Full Access

Introduction

The advent of double bundle ACL (Anterior Cruciate Ligament) reconstruction had been hailed as potentially allowing for more anatomically and physiologically functioning graft, however until recently there had been little evidence of enhanced functional outcomes.

The aim of this study is to explore whether the dimensions of hamstring two strand single bundle grafts, are predictive of the combined four strand single bundle graft that results from combining the single bundle grafts, as well as the impact of double bundle grafts upon the available healing and attachment area within the bony tunnels.

Methods

Grafts of all likely two strand single bundle graft sizes, measured to the nearest 0.5mm in diameter using unslotted sizing block, were prepared using porcine flexor tendons,. These two strand single bundles were then systematically combined, and re-measured.

By geometrical calculation, the sum of the circumferences of the two, two strand double bundle grafts were compared to the combined four strand single bundle graft formed by combining the two smaller bundles.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 382 - 383
1 Oct 2006
Clements J Gheduzzi S Webb J Schmotzer H Learmonth I Miles A
Full Access

Introduction: Immediate postoperative stability of cementless hip stems is one of the key factors for the long-term success of total hip replacement. The ability to discriminate between stable and unstable stems in the laboratory constitutes a desirable tool for the industry, as it would allow the identification of unsuitable stem designs prior to clinical trials. The use of composite femora for stability investigations is wide spread [1,2] even though their use in this application is yet to be validated. This study is aimed at establishing whether Sawbones composite femora are suitable for the assessment of migration and micromotion of a cementless hip stem. The stability of two SL Plus stems (Precision Implants, CH) implanted into Sawbone was compared to that of two SL Plus stems implanted into cadaveric femora. Ethical approval was obtained for the harvest and use of cadaveric material.

Methods: Stability was assessed in terms of micromotion and migration. Micromotion was defined as the recoverable movement of the implant relative to the bone under cyclic loading. Migration was defined as the non-recoverable movement of the implant with respect to the surrounding bone. Movement of the implant with respect to the surrounding bone was monitored at two locations on the lateral side of the stem by means of two custom made transducers based on the concept described by Berzins et al [3]. Each femur was tested in two different sinusoidal loading configurations: single leg stance (SLS-11° of adduction and 7° of flexion) [4] loaded up to 400N and stair climbing (SC-11° of adduction and 32° of flexion) loaded up to 300N. The effect of the abductor muscles was included in the model [5]. Each test consisted of 200 loading cycles applied at 50 Hz. The captured data was post-processed by a MATLAB routine and converted into translations and rotations of the stem with respect to the bone.

Results: The proximal part of the implant was subject to the highest amplitudes of micromotion in both loading configurations independent of the host. During SLS the largest micromotion was measured in the direction of the axis of the femur, this amplitude was in the order of 20 μm for the stems implanted in sawbones and varied between 13 and 39 μm for the stems implanted in cadaveric femora. The migration of the implants was minimal both in SLS and SC for both hosts with values measured in the sawbones model nearly on order of magnitude smaller than the cadaveric. In the case of SLS the prevalent movement consisted of a translation along the axis of the bone, while during SC the rotations became prevalent.

Discussion: This study has demonstrated that Sawbones provide an effective model to establish micromotion with oscillation patterns and orders of magnitiude similar to cadaveric bone. However the migration is much more dependent on the quality of fit and the internal geometry of the femur and therefore more caution should be placed on interpreting migration data from Sawbones models.