Caraan N Windhager R Kuehn K
Full Access

BACKGROUND

This scientific work is a non-interventional, experimental and prospective comparative study of two very high-viscosity PMMA bone cements: DePuy CMW 2G and Palacos® fast R+G. Reference product: Palacos® R+G.

Fast-setting PMMA bone cements are used in the endoprothetics of the patella and knee (in Australia) and are also used to cement an artificial acetabulum (in the UK). Are there any differences regarding the characteristics of the two fast-setting PMMA bone cements?

MATERIALS AND METHODS

All cements were mixed as specified by the manufacturer and analysed on the following parameters: handling properties (mixing, waiting, working and hardening phase), powder/liquid-ratio, mechanical properties (ISO 5833:2002 and DIN 53435), fatigue strength (ISO 16402) and elution profile. All tests were done in an acclimatised laboratory with temperatures set at 23.5°C ± 0.5°C and a humidity of >40%. Of two batch numbers, 11 units of each bone cement were tested.