The constraint of total knee replacement (TKR) implants is not simply defined and many of the factors that influence it are not well understood. Variability in the constraint of different TKR implants designed for the same indication (e.g. cruciate-retaining, or posterior-stabilized) have been previously demonstrated, but these differences among implants have yet to be simply quantified. Furthermore, the relative importance of several variables on the implant constraint remains unknown. The purpose of this study was to quantify the differences in constraint that exist between different implant designs, and to examine the effects of axial load and flexion angle on the constraint of current cruciate-retaining (CR) TKR components. Four contemporary CR TKR designs underwent laxity testing using a multi-axis mechanical test machine. Implants were tested at flexion angles of 0°, 20°, 90° and maximum flexion and axial loads of 712 N (1 BW) and 1424 N (2 BW). Friction-free motion in all secondary degrees of freedom was allowed. Force-displacement curves were generated for each testing condition in both anterior-posterior (AP) and rotational tests. AP constraint (N/mm) and rotational constraint (Nm/deg) were then calculated.Background
Methods