Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_22 | Pages 24 - 24
1 Dec 2017
Ferry T Johan A Boucher F Chateau J Hristo S Daoud F Braun E Triffault-Fillit C Perpoint T Laurent F Alain-Ali M Chidiac C Valour F
Full Access

Aim

A two-stage surgical strategy (debridement-negative pressure therapy (NPT) and flap coverage) with prolonged antimicrobial therapy is usually proposed in pressure ulcer-related pelvic osteomyelitis but has not been widely evaluated.

Method

Adult patients with pressure ulcer-related pelvic osteomyelitis treated by a two-stage surgical strategy were included in a retrospective cohort study. Determinants of superinfection (i.e., additional microbiological findings at reconstruction) and treatment failure were assessed using binary logistic regression and Kaplan-Meier curve analysis.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 272 - 272
1 Dec 2013
Connor E Boucher F Wuestemann T Crawford R
Full Access

Introduction

The Exeter cemented polished tapered stem design was introduced into clinical practice in the early 1970's. [i] Design and cement visco-elastic properties define clinical results [ii]; a recent study by Carrington et al. reported the Exeter stem has 100% survivorship at 7 years. [iii] Exeter stems with offsets 37.5–56 mm have length 150 mm (shoulder to tip). Shorter stems, lengths 95–125 mm, exist in offsets 30–35.5 mm. The Australian National Joint Replacement Registry recently published that at 7 years the shorter stems are performing as well as longer stems on the registry [iv]. Clinical observation indicates in some cases of shorter, narrower femora that fully seating a 150 mm stem's rasp in the canal can be difficult, which may affect procedural efficiency. This study investigates the comparative risk of rasp distal contact for the Exeter 150 mm stem or a 125 mm stem.

Materials and Methods

Rasps for 37.5, 44, 50 mm offset, No.1, 150 mm length stems (Exeter, Stryker Orthopaedics, Mahwah NJ) were compared with shortened length models using SOMA™ (Stryker Orthopaedics Modeling and Analytics technology). 637 patients' CT scanned femora were filtered for appropriate offset and size by measuring femoral-head to femoral-axis distance and midsection cancellous bone width (AP view). These femora were analyzed for distal contact (rasp to cortices) for 150 mm and 125 mm models (Figure 1). The widths of the rasp's distal tip and the cancellous bone boundary were compared to assess contact for each femur in the AP and ML views; the rasp was aligned along an ideal axis and flexed in order to pass through the femoral neck (ML view only).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 280 - 280
1 Mar 2013
Nevelos J Boucher F Parvizi J Nessler JP Kolisek FR
Full Access

A novel cementless tapered wedge femoral hip implant has been designed at a reduced length and with a geometry optimized to better fit a wide array of bone types (Accolade II, Stryker, Mahwah, USA). In this study, finite element analysis (FEA) is used to compare the initial stability of the new proposed hip stem to predicate tapered wedge stem designs. A fit analysis was also conducted. The novel stem was compared to a predicate standard tapered stem and a shortened version of that same predicate stem.

Methods

The novel shortened tapered wedge stem geometry was designed based on a morphological study of 556 CT scans. We then selected 10 discrete femoral geometries of interest from the CT database, including champagne fluted and stove pipe femurs. The novel and the predicate stems were virtually implanted in the bones in ABAQUS CAE. A total of thirty FEA models were meshed with 4 nodes linear tetrahedral elements. Bone/implant interface properties was simulated with contact surface and a friction coefficient of 0.35. Initial stability of each stem/bone assembly was calculated using stair-climbing loading conditions. The overall initial stability of the HA coated surface was evaluated by comparing the mean rotational, vertical, gap-opening and total micromotion at the proximal bone/implant interface of the novel and predicate stem designs.

To characterize the fit of the stem designs we analyzed the ratio of a distal (60 mm below lesser trochanter) and a proximal (10 mm above lesser trochanter) cross section. A constant implantation height of 20 mm above the lesser trochanter was used. The fit of the stems was classified as Type 1 (proximal and distal engagement), Type 2 (proximal engagement only) and Type 3 (distal engagement only).

Results

The mean % micromotion of the HA coated surface greater than 50 mm was lowest at 40.2% (SD 11.5%) for the novel tapered wedge stem compared to the clinically successful predicate stem design (Accolade TMAZ, Stryker, Mahwah, USA) at 44.9% (SD 13.2%) and its shortened version at 48.5% (SD 9.0%) as shown in Figure 1. Improved initial stability of the new stem was also confirmed for rotational, vertical and gap-opening micromotion. However, there was no statistically significant difference.

The novel tapered stem design showed a well balanced proximal to distal ratio throughout the complete size range. The novel tapered stem design showed a reduced percentage of distal engagements (2.8%) compared to the predicate standard stem (17.2%). In the 40 to 60 year old male group the distal engagement for the standard stem increases (28.2%), whereas the distal engagements for the novel stem remains unchanged (1.3%).


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 523 - 523
1 Nov 2011
Hamadouche M Zaoui A El Hage S Moindreau M Boucher F Mathieu M Courpied J
Full Access

Purpose of the study: The purpose of this prospective study was to evaluated the risk of fracture of 22.2 mm Delta ceramic heads.

Material and methods: A preclinical study was performed on twenty 22.2 mm Delta ceramic femoral heads with a medium neck with 20 22.2 mm Delta ceramic femoral heads with a short neck. A V40TM cone was used in all cases with a 5 40 angle. In vitro tests consisted in the assessment of the fracture force under static pressure before and after stress tests, and with a static force shock test (Charpy model) simulating a microseparation during subluxation phenomena. Between April 2007 and April 200, a consecutive series of 55 composite Delta heads were used in 55 patients undergoing cemented total hip arthroplasty (THA). A polyethylene cup was sterilised under vacuum at 3 Mrads with a post-radiation temper (Duration®), and a stainless steel femoral piece with a highly polished surface and a V40TM Morse cone (Legend®). This series issued from a randomised prospective study designed to compare wear with 22.2mm Delta ceramic heads with stainless steel heads with the same diameter. Inclusion criteria were age < 75 years, degenerative disease on naive hip and patient residing in France. Mean patient age was 59.2±6.9 years (range 44–70).

Results: The resistance of the 22.2 mm heads was significantly less during the static tests before and after the stress tests. Nevertheless, the mean resistance was higher than the FDA recommendations of 46K. The tests simulating a microseparation showed a significantly superior resistance for the 22.2 mm heads. All patients had from 1 to 2 years follow-up (usual delay for 80% of in vivo ceramic fractures). There were no cases of femoral head fracture in this series. There were no cases of early wear at this same follow-up and no case of femoral or acetabular osteolysis.

Discussion and Conclusion: The results of this study indicate that the resistance of the 22.2mm ceramic Delta heads is very much superior to the recommendations for in vitro tests. In this series, the risk of fracture in vivo remained nil to two years follow-up. The pertinence of this ceramic implant for decreasing polyethylene wear in vivo is under evaluation.