header advert
Results 1 - 9 of 9
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 66 - 66
1 Jan 2018
Muratoglu O Suhardi V Bichara D Freiberg A Rubash H Malchau H Oral E
Full Access

The gold standard for PJI treatment comprises the use of antibiotic-loaded bone cement spacers, which are limited in their load bearing capacity[1]. Thus, developing an antibiotic-eluting UHMWPE bearing surface can improve the mechanical properties of spacers and improve the quality of life of PJI patients. In this study, we incorporated vancomycin into UHMWPE to investigate its elution characteristics, mechanical properties and its efficacy against an acute PJI in an animal model.

Vancomycin hydrochloride was incorporated into UHMWPE (2 to 14%) by blending and consolidation. We studied drug elution with blocks in PBS and UV-Vis spectroscopy at 280 nm. We determined the tensile mechanical properties and impact strength [3]. We implanted osteochondral plugs in rabbits using either control UHMWPE, bone cement (40g) containing vancomycin (1g) and tobramycin (3.6g) or vancomycin-eluting UHMWPE (n=5) plugs in the patellofemoral groove of rabbits. All rabbits received a beaded titanium rod in the tibial canal. All groups received two doses of 5×107 cfu of bioluminescent S. aureus in the distal tibial canal prior to insertion of the rod and the articular space after closure of the joint capsule. No intravenous antibiotics were used. Bioluminescence signal was measured when the rabbits expired, or at 21-day post-op. Hardware, polyethylene implants, and joint tissues were sonicated to further quantify live bacteria via plate seeding.

Vancomycin elution increased with increasing drug loading. Vancomycin elution above MIC for 3 weeks and optimized mechanical properties were obtained at 6–7 wt% vancomycin loading in UHMWPE. In our lapine acute infection model using bioluminescent S. aureus, knees treated with UHMWPE without antibiotics and bone cement containing vancomycin and tobramycin had significantly higher bioluminescence compared to those treated with vancomycin-eluting UHMWPE.

These results suggest that an antibiotic-eluting UHMWPE spacer with acceptable properties as a bearing surface could be used to treat periprosthetic joint infection in lieu of bone cement spacers and this could allow safer load bearing and a higher quality of life for the patients during treatment. In addition, this presents a safer alternative in cases where the second stage surgery for the implantation of new components is hindered.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 69 - 69
1 Mar 2017
Muratoglu O Oral E Suhardi V Bichara D Rubash H Freiberg A Malchau H
Full Access

Introduction

Radiation cross-linking of ultrahigh molecular weight polyethylene (UHMWPE) has reduced the in vivo wear and osteolysis associated with bearing surface wear (1), significantly reducing revisions associated with this complication (2). Currently, one of the major and most morbid complications of joint arthroplasty is peri-prosthetic infection (3). In this presentation, we will present the guiding principles in using the UHMWPE bearing surface as a delivery device for therapeutic agents and specifically antibiotics. We will also demonstrate efficacy in a clinically relevant intra-articular model.

Materials and Methods

Medical grade UHMWPE was molded together with vancomycin at 2, 4, 6, 8, 10 and 14 wt%. Tensile mechanical testing and impact testing were performed to determine the effect of drug content on mechanical properties. Elution of the drug was performed in phosphate buffered saline (PBS) for up to 8 weeks and the detection of the drug in PBS was done by UV-Vis spectroscopy. A combination of vancomycin and rifampin in UHMWPE was developed to address chronic infection and layered construct containing 1 mm-thick drug-containing UHMWPE in the non-load bearing regions was developed for delivery. In a lapine (rabbit) intra-articular model (n=6 each), two plug of the layered UHMWPE construct were placed in the trochlear grove of the rabbit femoral surface and a porous titanium rod with a pre-grown biofilm of bioluminescent S. Aureus was implanted in the tibia. Bioluminescent imaging was employed to visualize and quantify the presence of the bacteria up to 3 weeks.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 25 - 25
1 Mar 2017
Muratoglu O Suhardi V Bichara D Bedair H Oral E
Full Access

Introduction

The use of narcotic medications to manage postoperative pain after TJA has been associated with impaired mobility, diminished capacity to engage in rehabilitation, and lower patient satisfaction [1]. In addition, side effects including constipation, dizziness, nausea, vomiting and urinary retention can prolong post-operative hospital stays. Intraarticular administration of local anesthetics such as bupivacaine – part of a multimodal postoperative pain management regimen – reduces pain and lowers patients' length of stay [2]. In addition to its anesthetic activity, bupivacaine also has antibacterial activity, particularly against gram-positive bacteria [3]. We have developed a bupivacaine-eluting ultrahigh molecular weight polyethylene (Bupi-PE) formulation; we hypothesized that elution of bupivacaine from polyethylene could have both anesthetic and antibacterial effects in vivo.

Methods


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 26 - 26
1 Mar 2017
Muratoglu O Suhardi V Bichara D Kwok S Freiberg A Rubash H Yun S Oral E
Full Access

Introduction

About 2% of primary total joint replacement arthroplasty (TJA) procedures become infected. Periprosthetic joint infection (PJI) is currently one of the main reasons requiring costly TJA revisions, posing a burden on patients, physicians and insurance companies.1 Currently used drug-eluting polymers such as bone cements offer limited drug release profiles, sometimes unable to completely clear out bacterial microorganisms within the joint space. For this study we determined the safety and efficacy of an antibiotic-eluting UHMWPE articular surface that delivered local antibiotics at optimal concentrations to treat PJI in a rabbit model.

Materials and Methods

Skeletally mature adult male New Zealand White rabbits received either two non-antibiotic eluting UHMWPE (CONTROL, n=5) or vancomycin-eluting UHMWPE (TEST, n=5) (3 mm in diameter and 6 mm length) in the patellofemoral groove (Fig. 1). All rabbits received a beaded titanium rod in the tibial canal (4 mm diameter and 12 mm length). Both groups received two doses of 5 × 107 cfu of bioluminescent S. aureus (Xen 29, PerkinElmer 119240) in 50 µL 0.9 % saline in the following sites: (1) distal tibial canal prior to insertion of the rod; (2) articular space after closure of the joint capsule (Fig. 1). None of the animals received any intravenous antibiotics for this study. Bioluminescence signal (photons/second) was measured when the rabbits expired, or at the study endpoint (day 21). The metal rods were stained with BacLight® Bacterial Live-Dead Stain and imaged using two-photon microscopy to detect live bacteria. Hardware, polyethylene implants and joint tissues were sonicated to further quantify live bacteria via plate seeding.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 43 - 43
1 Feb 2017
Muratoglu O Bichara D O'Brien C Doshi B Oral E
Full Access

Introduction

We have previously demonstrated that peroxide crosslinked vitamin E-blended UHMWPE maintains its clinically-required wear and mechanical properties [1]. This material can potentially be used as an irradiation-free bearing surface for TJA. However, using organic peroxides in medical devices requires a thorough examination of tissues in contact with the implant. For this study we crosslinked polyethylene using five times the needed concentration of peroxide (2,5-Dimethyl-2,5-di(t-butylperoxy)-hexyne-3 or P130), followed by implantation to determine implant biocompatibility, and pre and post implant peroxide residual contents.

Methods

The study was performed after institutional approval following ISO standard 10993–6. Study groups: not crosslinked (0.2 (1050) VE), crosslinked (0.2 VE (1050)/5% P130) and crosslinked-high temperature melted (HTM) (0.2 VE (1050)/5% P130). Materials were blended and consolidated, machined (2.5 diameter × 2.5 cm height), sterilized and implanted in the dorsum New Zealand white rabbits. Pre and post implantation FTIR was performed. Two samples were implanted in each rabbit; n=6 samples were included for each group. After 4 weeks, samples were explanted, analyzed using FTIR, and subcutaneous tissues processed for histological analysis.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 67 - 67
1 Feb 2017
Muratoglu O Chen W Suhardi V Bichara D
Full Access

Introduction

Periprosthetic joint infection (PJI) and particle-induced osteolysis are closely related to peri-implant local immunity and macrophage function. We previously demonstrated that titanium particles attenuate the immune response of macrophages caused by chronic inflammation [1]. In a separate study, we have determined that UHMWPE wear particles containing vitamin E (VE) induce less osteolysis compared to HXL UHMWPE wear particles in a murine calvarium model [2]. For this study we hypothesized that macrophages exposed to HXL UHMWPE particles containing VE would better maintain their ability to respond to S. aureus compared to HXL UHMWPE without VE.

Methods

A gamma-sterilized, HXL UHMWPE tibial bearing containing VE (E1, Biomet, “VE-PE”) and 100kGy irradiated and melted UHMWPE (“CISM 100”) were cryomilled to particles by Bioengineering Solutions (Oak Park, IL). In the first in vitro study, RAW 264.7 mouse macrophages were exposed (inverted co-culture) to either VE-PE particles or CISM100 particles and lipopolysaccharide (LPS) for 1–7 days. Macrophage viability was measured using a cell counting kit (CCK-8). Control group with no particles and a LPS group were also included. Terminal deoxynucleotidyl transferase dUTP nick end labeling (TUNEL) staining was performed to determine macrophage apoptosis rate in response to particle exposure over time. In the second study, macrophages were exposed to VE-PE or CISM100 particles for 48h, then exposed to LPS for 30 min. Subsequently, reactive oxygen species (ROS) generation and extracellular regulated protein kinase (ERK) phosphorylation were measured. In a third study, after exposure to particles for 48h, fatigued macrophages were co-cultured with bioluminescent S. aureus strain Xen29 for 3h and 6h. Bioluminescence signal was determined to measure the total amount of bacteria. Bacterial live/dead staining and optical density at 600 nm (OD 600) were also performed to determine S. aureus viability. Statistical analysis was performed using one-way or two-way ANOVA with a post hoc examination. *indicates p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 61 - 61
1 Jan 2016
Muratoglu O Bichara D Malchau E Hylleholt N Cakmak S
Full Access

Introduction

UHMWPE particle-induced osteolysis is one of the major causes of arthroplasty revisions. Recent in vitro findings have suggested that UHMWPE wear particles containing vitamin-E (VE) may have reduced functional biologic activity and decreased potential to cause osteolysis (Bladed C. L. et al, JBMR B 2012 and 2013). This is of significant importance since VE-stabilized cross-linked UHMWPEs were recently introduced for clinical use, and there is no in vivo data determining the effects of wear debris. In this study we hypothesized that particles from VE-stabilized, radiation cross-linked UHMWPE (VE-UHMWPE) would cause reduced levels of osteolysis in a murine calvarial bone model when compared to virgin gamma irradiated cross-linked UHMWPE.

Methodology

Study groups were the following: 1). Radiation cross-linked VE-UHMWPE (0.8% by weight) diffused after 100 kGy; 2). Radiation cross-linked virgin UHMWPE (virgin UHMWPE); 3). Sham controls. Particle generation and implantation: UHMWPE was sent to Bioengineering Solutions (Oak Park, IL) for particle generation. After IACUC approval, C57BL/6 mice (n=12 for each group) received equal amount of particulate debris (3mg) overlying the calvarium and were euthanized after 10 days. Micro-CT scans: High resolution micro-CT scans were performed using a set voltage of 70 kV and current of 70 µA. Topographical Grading Scale: Each calvarial bone was blindly scored using the following scale: 0=No osteolysis, defined as intact bone; 1=Minimal osteolysis, affecting 1/3 or less of the bone area; 2=Moderate osteolysis, affecting at least 2/3 of the bone area; 3=Severe osteolysis, defined as completely osteolytic bone. Histology: H&E and TRAP staining was done on tissue to confirm micro-CT findings and quantify osteoclasts. Statistical Analysis: Inter-rater analysis was done using Cohen's kappa analysis. An inter-rater coefficient >0.65 was considered as high inter-rater agreement. Comparison between groups was made using one-way ANOVA with post hoc Bonferroni correction for multiple comparisons. Correlations are reported as Spearman's rho. P-value<0.05 was considered statistically significant.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 255 - 255
1 Jul 2014
Bichara D Malchau E Sillesen N Cakmak S Muratoglu O
Full Access

Summary Statement

Vitamin E-UHMWPE particles have a reduced osteolysis potential in vivo when compared to virgin, highly cross-linked UHMWPE in a murine calvarial bone model.

Introduction

Ultra high-molecular weight polyethylene (UHMWPE) particle-induced osteolysis is one of the major causes of arthroplasty revisions. The lack of particle clearance from the joint inevitably leads to the upregulation of the inflammatory cascade, resulting in bone resorption and implant loosening. Recent in vitro findings (Bladed CL et al. ORS 2011 and J Biomed Mater Res B Appl Biomater, 2012) have suggested that UHMWPE wear particles containing vitamin-E (VE) may have reduced functional biologic activity and decreased potential to cause osteolysis. This is of significant importance since VE-stabilised cross-linked UHMWPEs were recently introduced for clinical use, and there is no in vivo data determining the effects of wear debris from this new generation of implants. In this study we hypothesised that particles from VE-stabilised, radiation cross-linked UHMWPE (VE-UHMWPE) would cause reduced levels of osteolysis in a murine calvarial bone model when compared to virgin gamma irradiated cross-linked UHMWPE.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 145 - 145
1 Dec 2013
Muratoglu O Bichara D Malchau E Hylleholt N Oral SCE
Full Access

Introduction

In vitro findings (Bladed CL et al. ORS 2011 and J Biomed Mater Res B Appl Biomater, 2012) have suggested that UHMWPE wear particles containing vitamin-E (VE) may have reduced functional biologic activity and decreased osteolytic potential. Currently, there is no in vivo data determining the effects of wear debris from this new generation of implants. In this study we hypothesized that particles from VE-stabilized, radiation cross-linked UHMWPE (VE-UHMWPE) would cause reduced levels of osteolysis in a murine calvarial bone model when compared to virgin gamma irradiated cross-linked UHMWPE.

Methods

Study groups: 1). Radiation cross-linked VE-UHMWPE, 0.8% by weight, diffused after 100 kGy; 2). Radiation cross-linked virgin UHMWPE (virgin UHMWPE); 3). Shams. Particle generation and implantation: UHMWPE was sent to Bioengineering Solutions for particle generation. After IACUC approval, C57BL/6 mice (n = 12 for each group) received 3 mg of particulate debris overlying the calvarium and euthanized after 10 days. Micro-CT scans: Performed using an X-Tek-HMX-ST-225 with 70 kV voltage and 70 μA current. Topographical Grading Scale: Each calvarial bone was blindly scored with the following scale: 0 = No osteolysis, defined as intact bone; 1 = Minimal osteolysis, affecting 1/3 or less of the bone area; 2 = Moderate osteolysis, affecting at least 2/3 of the bone area; 3 = Severe osteolysis, defined as completely osteolytic bone. Histology H&E and TRAP staining was performed. Statistical Analysis: Inter-rater analysis was performed using Cohen's kappa analysis. Inter-rater coefficient >0.65 was considered as high inter-rater agreement. Comparison between groups was made using one-way ANOVA with post hoc Bonferroni correction for multiple comparisons. Correlations are reported as Spearman's rho. A p-value<0.05 was considered statistically significant.