Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 1 - 1
1 May 2016
Giles J Amirthanayagam T Emery R Amis A Rodriguez-Y-Baena F
Full Access

Background

Total Shoulder Arthroplasty (TSA) has been shown to improve the function and pain of patients with severe degeneration. Recently, TSA has been of interest for younger patients with higher post-operative expectations; however, they are treated using traditional surgical approaches and techniques, which, although amenable to the elderly population, may not achieve acceptable results with this new demographic. Specifically, to achieve sufficient visualization, traditional TSA uses the highly invasive deltopectoral approach that detaches the subscapularis, which can significantly limit post-operative healing and function. To address these concerns, we have developed a novel surgical approach, and guidance and instrumentation system (for short-stemmed/stemless TSA) that minimize muscle disruption and aim to optimize implantation accuracy.

Development

Surgical Approach: A muscle splitting approach with a reduced incision size (∼6–8cm) was developed that markedly reduces muscle disruption, thus potentially improving healing and function. The split was placed between the infraspinatus and teres-minor (Fig.1) as this further reduces damage, provides an obvious dissection plane, and improves access to the retroverted articular surfaces. This approach, however, precludes the use of standard bone preparation methods/instruments that require clear visualization and en-face articular access. Therefore, a novel guidance technique and instrumentation paradigm was developed.

Minimally Invasive Surgical Guidance: 3D printed Patient Specific Guides (PSGs) have been developed for TSA; however, these are designed for traditional, highly invasive approaches providing unobstructed access to each articular surface separately. As the proposed approach does not offer this access, a novel PSG with two opposing contoured surfaces has been developed that can be inserted between the humeral and scapular articular surfaces and use the rotator cuff's passive tension to self-locate (Fig.2). During computer-aided pre-operative planning/PSG design, the two bones are placed into an optimized relative pose and the PSG is constructed between and around them. This ensures that when the physical PSG is inserted intra-operatively, the bones are locked into the preoperatively planned pose.

New Instrumentation Paradigm: With the constraints of this minimally invasive approach, a new paradigm for bone preparation/instrumentation was required which did not rely on en-face access. This new paradigm involves the ability to simultaneously create glenoid and humeral guide axes – the latter of which can guide humeral bone preparation and be a working channel for tools – by driving a short k-wire into the glenoid by passing through the humerus starting laterally (Fig.3). By preoperatively defining the pose produced by the inserted PSG as one that collinearly aligns the bones’ guide axes, the PSG and an attached c-arm drill guide facilitate this new lateral drilling technique. Subsequently, bone preparation is conducted using novel instruments (e.g. reamers and drills for creating holes radial to driver axis) powered using a trans-humeral driver and guided by the glenoid k-wire or humeral tunnel.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 27 - 27
1 Oct 2012
Strachan R Konala P Iranpour F Prime M Amirthanayagam T Amis A
Full Access

Anatomical referencing, component positioning, limb alignments and correction of mechanical axes are essential first steps in successful computer assisted navigation. However, apart from basic gap balancing and quantification of ranges of motion, routine navigation technique usually fails to use the full potential of the registered information. Enhanced dynamic assessment using an upgraded navigation system (Brainlab V. 2.2) is now capable of producing enhanced ‘range of motion’ analysis, ‘tracking curves’ and ‘contact point observations’.

‘Range of motion analysis’ was performed simultaneously for both tibio-femoral and patella-femoral joints. Other dynamic information including epicondylar axis motion, valgus and varus alignments, antero-posterior tibio-femoral shifts, as well as flexion and extension gaps were simultaneously stored as a series of ‘tracking curves’ throughout a full range of motion. Simultaneous tracking values for both tibiofemoral and patellofemoral motion was also obtained after performing registration of the prosthetic trochlea. However, there seems to be little point in carrying out such observations without fully assessing joint stability by applying controlled force to the prosthetic joint.

Therefore, in order to fully assess ‘potential envelopes of motion’, observations have been made using a set of standardised simple dynamic tests during insertion and after final positioning of trial components. Also, such tests have been carried out before and after any necessary ligament balancing. Firstly, the lower leg was placed in neutral alignment and the knee put through a flexion-extension cycle. Secondly the test was repeated but with the lower leg being placed into varus and internal rotation. The third test was performed with the lower leg in valgus and external rotation. Force applied was up to the point where resistance occurred without any gross elastic deformation of capsule or ligament in a manner typical of any surgeon assessing the stability of the construct. Also a passive technique of using gravity to ‘Drop-Test’ the limb into flexion and extension gave useful information regarding potential problems such as blocks to extension, over-stuffing of the extensor mechanism and tightness of the flexion gap. All the definitive tests were performed after temporary medial capsular closure.

Ten total knee arthroplasties have been studied using this technique with particular reference to the patterns of instability found before, during and after adjustments to component positioning and ligament balancing. Marked intra-operative variation in the stability characteristics of the trial implanted joints has been quantified before correction. These corrections have been analysed in terms of change in translations, rotations and contact points induced by any such adjustments to components and ligament. Certain major typical patterns of instability have begun to be identified including excessive rotational and translational movements. Instability to valgus and external rotational stress was found in two cases and to varus and internal rotational stress in one case before correction. In particular, surprising amounts of edge loading in mid-flexion under stress testing has been identified and corrective measures carried out. Reductions in paradoxical tibio-femoral antero-posterior motion were also observed. Global instability and conversely tightness were also observed in early stages of surgery. Adjustments to component sizes, rotations, tibial slope angles and insert thickness were found to be necessary to optimise range of motion and stability characterisitics on an almost case-by-case basis. Two cases were identified where use of more congruent or stabilised components was necessary. Observation of quite marked loss of contact between tibia and femur was seen on the lateral side of the knee in deep flexion in several cases. Patellar tracking was also being observed during such dynamic tests and in two cases staged partial lateral retinacular releases were carried out to centre patellar tracking on the prosthetic trochlea.

Although numbers in this case series are small, it has been possible to begin to observe, classify and quantify patterns of instability intra-operatively using simple stress tests. Such enhanced intra-operative information may in future make it possible to create algorithms for logical and precise adjustments to ligaments and components in order to optimise range of motion, contact areas and stability in TKR.