Purpose

The foam metal backed Advance BioFoam Knee Arthroplasty components utilize a porous titanium coating on the underside of the tibial baseplate, intended to promote bone in-growth and provide a more robust bone-implant interface without cement. There is also a version of the Biofoam Advance that incorporates screwed fixation that allows for augmented fixation with up to four titanium screws; however, it is not clear that this augmentation is necessary. The purpose of this study was to employ radiostereometric analysis (RSA) to compare implant migration in a randomized controlled trial of this implant design with or without screw fixation.

Method

Fifty-one patients were randomized to receive a BioFoam total knee replacement (Wright Medical Technologies) with or without screw fixaiton. During surgery, eight tantalum markers, one millimetre in diameter, were inserted into the proximal tibia. Using a calibration box, stereo RSA radiographs were taken post-operatively and then again at six weeks and three, six and 12 months following surgery. Model Based RSA was used with 3D models of the tibial component to measure migration. Health status and functional outcome measures were recorded to quantify functional status of subjects before surgery and at each follow-up interval.