Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 67 - 67
1 Sep 2012
Labek G Thaler M Agreiter M Williams A Krismer M Böhler N
Full Access

Introduction

Austin Moore cervicocephalic prostheses have been a therapeutical option for femoral neck fractures in patients with a reduced general condition for many years. Since treatments other than total hip arthroplasties have also been included in National arthroplasty registers during the last decade, adequate reference data for comparative analyses have recently become available.

Materials and Methods

Based on a standardised methodology, a comprehensive literature analysis of clinical literature and register reports was conducted. On the one hand, the datasets were examined with regard to validity and the occurrence of possible bias factors, on the other hand, the objective was to compile a summary of the data available. The main criterion is the indicator of Revision Rate. The definitions used with respect to revisions and the methodology of calculations are in line with the usual standards of international arthroplasty registers.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 184 - 184
1 May 2011
Labek G Pawelka W Janda W Liebensteiner M Williams A Agreiter M Krismer M
Full Access

Background: Implant fractures are complications that have a great impact on the patient’s quality of life after total hip arthroplasty. Nevertheless their occurrence is often considered as rare in clinical practice. We compared incidences of implant fractures in various datasets in order to calculate the risk of a fracture and assess the quality of these datasets for such evaluations.

Methods: In a structured literature analysis based on a standardised methodology the incidence of reoperations was evaluated comparing clinical studies published in Medline-listed journals and annual reports of National Arthroplasty Registers worldwide. Case reports and experimental studies were not considered.

Results: The majority of clinical studies are monocentre trials. The publications comprise a cumulative number of 72,571 stems with 234 stem fractures, 73,743 cups with 191 component fractures, and 16,381 ceramic heads with 44 fractures. A survey among the members of the American Association of Hip and Knee Surgeons covered 64,483 primary operations, hence including a similar number of primary cases as all monocentre studies together. This dataset involves 355 implant fractures.

By contrast, worldwide Register data refer to 733,000 primary operations, i.e. approximately 10 times as many as sample-based datasets.

In general, sample-based datasets present higher revision rates than register data. The deviations are high, with a maximum factor of 64 for hip stems. Whereas the AAHKS survey exhibits lower deviations than the monocentre trials, they are still too high for this data collection tool being considered as reliable and safe to provide valid data for general conclusions.

The incidence of implant fractures after total hip arthroplasty in pooled worldwide arthroplasty register datasets is 304 fractures per 100.000 implants. In other words, one out of 323 patients has to undergo revision surgery due to an implant fracture after THA in their lifetime.

Conclusion: For general assessments in the context of implant fractures, register data have the highest value. Clinical studies, which often focus on a particular implant, are of very limited value for global conclusions. Structured surveys produce more reliable data than clinical studies and are superior to monocentre trials. However, the AAHKS survey presents data that also show considerable differences to data from registers.

For the detection of rare, but severe complications like implant fractures sample-based studies achieve the goal of providing accurate figures only to a very limited extent, even if the samples are large. Here, too, comprehensive national arthroplasty registers are the most suitable tool to identify such incidents and calculate reliable figures.

Contrary to the prevalent opinion, implant fractures still are a relevant problem in arthroplasty.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 131 - 131
1 May 2011
Labek G Sekyra K Pawelka W Janda W Agreiter M Schlichtherle R Stöckl B Krismer M
Full Access

Background: Within the scope of the EU project EUPHORIC a methodology for direct comparison of different datasets was developed and applied on a sample of implants, among them the Oxford Unicompartmental Knee Arthroplasty (Oxford Uni). The aim was to identify potential bias factors inherent in the datasets and evaluate the outcome achieved with this implant.

Materials and Methods: A structured comparison was performed of data published on the revision rate of the Oxford Unicompartmental prosthesis. Both clinical follow-up studies published in Medline-listed journals and worldwide Register data were included. The data were stratified with regard to potential influence factors like the individual research groups or the geographical origin of the papers.

Results: A major proportion of the published data, between 50 and 75%, depending on the method of calculation, comes from studies including the developing institution in Oxford. The results published by this group deviates statistically significantly from the reference datasets from Register data or independent research groups. Data from the developing hospital show mean revision rates that are 4.4 times lower than those based on worldwide Register data, and 2.74 times lower than in independent studies. As opposed to this, independent studies on average publish data that are reproducible in Register data.

Conclusion: A conventional meta-analysis of clinical studies is significantly affected through the influence of the developing institution and is therefore subject to a bias. Neither through arthroplasty Register outcome data nor by other research groups that have disclosed outcome information on the Oxford Uni can the excellent results be reproduced that were published by the inventors.

Compared to other implants for unicompartmental knee arthroplasty in worldwide arthroplasty Registers, the Oxford Uni shows good results.

For the assessment of the outcome of implants, register data are to be rated superior and, in terms of reference data for the detection of potential bias factors in the clinical literature, can provide an essential contribution for scientific meta-analyses.