Abstract. Background.
While
The purpose of this study was to precisely measure the 10-year polyethylene wear rate of primary total hips using Radiostereometric analysis (RSA) comparing
Introduction.
Introduction. In Total Hip Arthroplasty (THA), polyethylene wear reduction is key to implant longevity.
This study reports the ten-year polyethylene liner wear rates, incidence of osteolysis, clinical outcomes and complications of a three-arm, multicentre randomised controlled trial comparing Cobalt-Chrome (CoCr) and Oxidised Zirconium (OxZr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) versus highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing THA from four institutions were prospectively randomised into three groups. Group A received a CoCr femoral head and XLPE liner; Group B received an OxZr femoral head and XLPE liner; and Group C received an OxZr femoral head and UHMWPE liner. Blinded observers recorded predefined outcomes in 262 study patients at regular intervals for ten years following THA. At ten years follow-up, increased linear wear rates were recorded in group C compared to group A (0.133 ± 0.21 mm/yr vs 0.031 ± 0.07 mm/yr respectively, p<0.001) and group B (0.133 ± 0.21 mm/yr vs 0.022 ± 0.05 mm/yr respectively, p<0.001). Patients in group C were associated with increased risk of osteolysis and aseptic loosening requiring revision surgery compared with group A (7/133 vs 0/133 respectively, p=0.007) and group B (7/133 vs 0/135 respectively, p=0.007). There was a non-significant trend towards increased liner wear rates in group A compared to group B (0.031 ± 0.07 mm/yr vs 0.022 ± 0.05 mm/yr respectively, p=0.128). All three groups were statistically comparable preoperatively and at ten years follow-up when measuring normalised Western Ontario and McMaster Universities Osteoarthritis Index(p=0.410), short-form-36 (p = 0.465 mental, p = 0.713 physical), and pain scale scores (p=0.451). The use of UHMWPE was associated with progressively increased annual liner wear rates after THA. At ten years follow-up, this translated to UHMWPE leading to an increased incidence of osteolysis and aseptic loosening requiring revision THA, compared with XLPE. Femoral heads composed of OxZr were associated with a non-significant trend towards reduced wear rates compared to CoCr, but this did not translate to any differences in osteolysis, functional outcomes, or revision surgery between the two treatments groups.
The success of total knee replacement (TKR) surgery can be attributed to improvements in TKR design, instrumentation, and surgical technique. Over a decade ago oxidized zirconium (OxZr) femoral components were introduced as an alternative bearing surface to cobalt-chromium (CoCr), based on strong in-vitro evidence, to improve the longevity of TKR implants. Early reports have demonstrated the clinical success of this material however no long-term comparative studies have demonstrated the superiority of OxZr implants compared to a more traditional CoCr implant. This study aims to compare long-term survivorship and outcomes in OxZr and CoCr femoral components in a single total knee design. We reviewed our institutional database to identify all patients whom underwent a TKA with a posterior stabilized OxZr femoral component with a minimum of 10 years of follow-up. These were then matched to patients whom underwent a TKA with the identical design posterior stabilized CoCr femoral component during the same time period by gender, age and BMI. All patients had their patella resurfaced. All patients were prospectively evaluated preoperatively and postoperatively at 6 weeks, 3 months, 12 months, 2 years and every 1 to 2 years thereafter. Prospectively collected clinical outcome measures included, Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Short-Form 12 (SF-12) and Knee Society clinical rating scores (KSCRS). Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years allowing comparison between the two cohorts. Paired analysis was performed to determine if differences existed in patient reported outcomes.Purpose
Methods
This study reports the ten-year outcomes of a three-arm, multicentre randomised controlled trial comparing Cobalt-Chrome (CoCr) and Oxidised Zirconium (OxZr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) versus highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing THA from four institutions were prospectively randomised into three groups. Group A received a CoCr femoral head and XLPE liner; Group B received an OxZr femoral head and XLPE liner; and Group C received an OxZr femoral head and UHMWPE liner. The outcomes of 262 study patients were analysed at ten years follow-up. At ten years, increased linear wear rates were recorded in group C compared to group A (0.133 ± 0.21 mm/yr vs 0.031 ± 0.07 mm/yr respectively, p<0.001) and group B (0.133 ± 0.21 mm/yr vs 0.022 ± 0.05 mm/yr respectively, p<0.001). Patients in group C had increased risk of osteolysis and aseptic loosening requiring revision surgery compared with group A (7/133 vs 0/133 respectively, p=0.007) and group B (7/133 vs 0/135 respectively, p=0.007). There was a non- significant trend towards increased liner wear rates in group A compared to group B (0.031 ± 0.07 mm/yr vs 0.022 ± 0.05 mm/yr respectively, p=0.128). All three groups were statistically comparable preoperatively and at ten years follow-up from a clinical score perspective. The use of UHMWPE was associated with progressively increased annual liner wear rates. At ten years follow-up, this translated to an increased incidence of osteolysis and aseptic loosening requiring revision, compared with XLPE. Femoral heads composed of OxZr were associated with a non-significant trend towards reduced wear rates compared to CoCr, but this did not translate to any differences in osteolysis, functional outcomes, or revision surgery between the two treatments groups.
One of the recent advances in the hard-on-hard hip arthroplasty is the development of a new material of diffusion hardened oxidised zirconium (DHOxZr). The DHOxZr material consists of a ceramic layer on the top surface which is supported by a thick oxygen diffusion hardened (DH) zone underneath. With the desired properties of metal substrate, ceramic surface and a gradient structure of the oxygen diffusion zone, the DHOxZr-on-DHOxZr bearing combination is expected to produce low wear and minimal metal ions. This can possibly address the concerns associated with metal hypersensitivity associated with metal on metal bearings and fracture risk associated with ceramics. The aim of this study was to evaluate the wear of DHOxZr-on-DHOxZr as a possible hard on hard bearing combination in hips. Three pairs of 50 mm DHOxZr prototype hip joint devices, each consisting of a DHOxZr modular head and a DHOxZr liner were wear tested in a ProSim hip joint simulator under standard testing conditions used by the Implant Development Centre (IDC), Smith & Nephew, Leamington Spa for 5 million cycles (Mc). The flexion/extension was 30° and 15°. The internal/external rotation was ± 10°. The force was Paul-type stance phase loading, with a maximum load of 3 kN and a standard ISO swing phase load of 0.3 kN. The test frequency was 1 Hz. Gravimetric analysis was carried out at 0, 0.5, 1, 2, 3, 4 & 5 million cycles. The lubricant was new born calf serum with 2 g/l sodium azide concentration diluted with de-ionised water to achieve average protein concentration of 20 g/l. Lubricant was changed every 0.25Mc during the first million cycles of the test and at every 0.33 Mc from 1 to 5Mc.INTRODUCTION
METHODS
Femoral components with an oxidized zirconium-niobium (OxZr) gradient ceramic surface (Oxinium, Smith & Nephew, Memphis, TN) were introduced as an alternative to cobalt-chromium (CoCr) alloy femoral components for the purpose of PE wear reduction in total knee replacements [1]. In the present study, the surface damage and clinical performance of both CoCr alloy and OxZr femoral components were investigated. By matching CoCr alloy and OxZr femoral components for clinical factors, as done by Heyse et al. [2], the surface damage on retrieved CoCr alloy and OxZr femoral component was assessed. Twenty-six retrieved cobalt-chromium (CoCr) alloy femoral components were matched with twenty-six retrieved oxidized zirconium (OxZr) femoral components for implantation period, body-mass index, patient gender, implant type (cruciate ligament retaining/substituting), and polyethylene insert thickness. Detailed surface profilometry was performed on retrieved femoral condyles in areas that had not been damaged by gouging [3] with the specific purpose of investigating the in vivo wear behaviour of undamaged OxZr surface. In addition, the cumulative survivorships were calculated for patients who had received CoCr alloy or OxZr femoral components from our orthopaedic database. In order to identify factors that affect the clinical performance of CoCr alloy and OxZr femoral components, the findings from the retrieval analysis and the survivorship analysis were combined. The Rp, Rpm, and Rpk-values for the retrieved CoCr alloy femoral components were found significantly higher than the Rp, Rpm, and Rpk-values for the retrieved OxZr femoral components (p ≤ 0.031). The roughness parameters values (Ra, Rq, Rz, Rp, Rpm, Rpk, Rv, and Rsk) for the retrieved CoCr alloy femoral components were found significantly higher than the values of the new, never implanted CoCr alloy femoral components (p ≥ 0.001). The surface roughness was higher on the medial condyles than the lateral condyles of the retrieved CoCr alloy femoral components; such a difference was not observed on the retrieved OxZr femoral components. The OxZr bearing surface appeared to protect the femoral components from abrasive wear in vivo. At 8.5-years follow up, the cumulative survivorship for the CoCr alloy femoral components (98%) was not found to be statistically significantly different (p = 0.343, Breslow test) from the OxZr femoral components (97.5%). Therefore, OxZr femoral components appeared to possess low wear characteristics and could be particularly suitable for younger, heavier patients to ensure long-term durability.
Although total knee arthroplasty (TKA) has been a reliable procedure providing durable pain relief, polyethylene (PE) wear remains a major limitation of the long-term success of TKA. One potential method of lowering PE wear in TKA is to use oxidized zirconium (OxZr)-bearing surface. Although wear simulating testing of an OxZr counter surface of femoral component produced less PE wear and fewer particles than did cobalt-chrome (Co-Cr) counter surface of femoral component [1–4], this finding has not been demonstrated in vivo to our knowledge. We measured in vivo PE wear by isolating and analyzing PE wear particles in synovial fluid from wellfunctioning TKA [5]. The purpose of the current study was to determine the size, shape, and amount of PE wear particles isolated from synovial fluid of patients who underwent a bilateral simultaneous TKA prosthesis, but different materials of femoral components. We performed a bilateral simultaneous TKA in 100 patients (200 knees) who received an OxZr femoral component in one knee and a Co-Cr femoral component in the other. Mean age was 55.6 (44–60) years. Synovial fluid was obtained from 28 patients (56 knees) who had undergone a simultaneous bilateral TKA under completely sterile conditions at one or two years after the operation. Randomization to an OxZr or Co-Cr femoral component was accomplished with use of a sealed study number envelope, which was opened in the operating room before the skin incision had been made. After the opening the randomization envelope, the first knee received prosthesis indicated by the envelope (OxZr or Co-Cr component) and the contralateral (second TKA) knee received the other prosthesis (OxZr or Co-Cr component). All operations were performed by one surgeon using the same design of total knee prosthesis: Genesis II (Smith and Nephew, Memphis, Tennessee). Only the material of the femoral component differed between two groups. The preoperative diagnosis was osteoarthritis in all patients. Preoperative and post operative KS and HSS knee scores, KS functional scores and UCLA activity scores were evaluated. The amount of polyethylene wear particles in the aspirated synovial fluid sample was analyzed by thermogravimetic analysis (TGA) using a TGA instrument (TGA/SDTA 84le model, Mettler Toledo CO., Greifensee, Switzerland). The weight of the sample solution was measured before and after removing the organic content by heating the sample solution. The sample solution was casted onto petri dishes. The petri dish was covered and kept in a dry oven at 60°C for 2 days. While the sample solution was kept in a dry oven for 2 days, a small hole was made on the cover of the petri dish to allow water to evaporate slowly for 2 days. After this procedure, the cover of petri dish was removed and TGA sample was dried at 60°C for another 2 days. After the sample was completely dried out, the dried sample was measured using analytical balance. TGA was used to determine the weight change profiles of polyethylene subject to heating under a nitrogen atmosphere. The nitrogen flow rate was kept constant at 50mL per minute. TGA data were taken at heating rate as 5°C per minute in the temperature range of 20° to 1000°C. The weight loss data were recorded as a function of time and temperature using special software in computer. When the temperature reached to the point of decomposition of the sample, the sample started to lose weight. By calculating the weight of the sample around the temperature which led to start to decomposition, real amount of polyethylene in the sample was measured. The size and shape of PE particles were examined using scanning electron microscopy (JSH-6360A model, Jeol Co., Tokyo, Japan). The samples were coated using a platinum sputtering machine for 20 sec. ANOVA, nonparametric chi square test, nonpaired t-test and Mann-Whitney U-test were used for statistical analyses. Differences of P<
0.05 were considered statistically significant. Mean preoperative KS (27.5 vs 27.2 points) scores, HSS (51.1 vs 51.2 points) knee scores, KS functional scores (55.4 vs 55.4 points) and UCLA activity scores (2.8 vs 2.8 point) were not significantly different between two groups. Mean postoperative KS (93 vs 92 points), HSS knee scores (90 vs 89 points), KS functional scores (78 vs 78 points), and UCLA activity scores (7.8 vs 7.8) were not significantly different. Mean weight of the polyethylene particles was 0.0219 g (SD, 0.0058) in the Co-Cr femoral component groups and it was 0.0214 g (SD, 0.005) in the OxZr group. This difference was not significant (P=0.711139, paired t-test). The size of particles was not different between the two groups. Also, shape of particles was not different between the two groups. Under the condition and the duration of this study in this specific group of patients, TKA with OxZr or Co-CR femoral knee component had excellent clinical and radiographic outcomes with no osteolysis. While the wear simulator test in vitro demonstrated significant decrease in PE particles in the knees with an OxZr femoral component, our study in vivo revealed that total particle weight, size, and shape of PE wear particles were similar in the knees with an OxZr femoral component and in those with a Co-Cr femoral component.
Large diameter femoral heads offer increased range of motion and reduced risk of dislocation. However, their use in total hip arthroplasty has historically been limited by their correlation with increased polyethylene wear. The improved wear resistance of highly crosslinked UHWMPE has led a number of clinicians to transition from implanting traditionally popular sizes (28mm and 32 mm) to implanting 36 mm heads. Desire to further increase stability and range of motion has spurred interest in even larger sizes (> 36 mm). While the long-term clinical ramifications are unknown, in-vivo measurements of highly crosslinked UHMWPE liners indicate increases in head diameter are associated with increased volumetric wear [1]. The goal of this study was to determine if this increase in wear could be negated by using femoral heads with a ceramic surface, such as oxidized Zr-2.5Nb (OxZr), rather than CoCrMo (CoCr). Specifically, wear of 10 Mrad crosslinked UHMWPE (XLPE) against 36 mm CoCr and 44 mm OxZr heads was compared. Ram-extruded GUR 1050 UHMWPE was crosslinked by gamma irradiation to 10 Mrad, remelted, and machined into acetabular liners. Liners were sterilized using vaporized hydrogen peroxide and tested against either 36 mm CoCr or 44 mm OxZr (OXINIUM(tm)) heads (n=3). All implants were manufactured by Smith & Nephew (Memphis, TN). Testing was conducted on a hip simulator (AMTI, Watertown, MA) as previously described [2]. The 4000N peak load (4 time body weight for a 102 kg/225 lb patient) and 1.15 Hz frequency used are based upon data obtained from an instrumented implant during fast walking/jogging and have previously been shown to generate measurable XLPE wear [2,3]. Lubricant was a serum (Alpha Calf Fraction, HyClone Laboratories, Logan, UT) solution that was replaced once per week [2]. Liners were weighed at least once every million cycles (Mcycle) over the duration of testing (∼ 5 Mcycle). Loaded soak controls were used to correct for fluid absorption. Single factor ANOVA was used to compare groups (a = 0.05).Introduction
Materials and Methods
Introduction. The development of new bearing surfaces for total joint replacement is constantly evolving.
INTRODUCTION. The development of new bearing surfaces for total joint replacement is constantly evolving.
Introduction. Wear-related osteolysis continues to be a concern in the long-term outcome and survivorship of total hip arthroplasty (THA) and there continues to be an emphasis on bearing materials that exhibit improved wear profiles.
The use of metal on polyethylene articulations was a key development in establishing total hip arthroplasty as a successful and reproducible treatment for end stage osteoarthritis. In order to ensure implant durability in relatively younger populations, there is a need for alternative, wear resistant bearing surfaces.
Traditional polyethylene oxidizes, wears and generates particles over time, which most probably contributes to increased risk of periprosthetic osteolysis. Even contemporary sterilization methods such as radiation and package in oxygen reduced or oxygen substituted environment do not eliminate oxidation over time. Thus, there is a need for alternative bearing in total hip replacement surgery and especially in patients with high activity and long life expectancy. All three major alternate bearings, ceramic-on-ceramic, metal-on-metal and highly crosslinked polyethylene produce major reductions in volumetric wear. The electron beam, melted highly cross-linked polyethylene has an in vivo penetration rate after the bedding in period, which is less than 8 microns per year. This is not substantially different from ceramic on ceramic or metal on metal. Therefore, the inherent risk of periprosthetic osteolysis with these alternate bearings is probably smaller than observed with conventional polyethylene. In the competition between different articulations highly cross-linked polyethylene has some advantages. The polyethylene is more adaptable than the hard bearing surfaces. This means that extended lip liners, offset liners, constrained liners and further special designs may be used. These options are not possible with any of the hard bearings. Another advantage with polyethylene is forgiveness. Impingement in hard-on-hard bearings may lead to serious complications such as chipping of the ceramics or metallosis in a metal on metal articulation. Impingement should also be avoided with use of polyethylene, but if it occurs, the consequences are often more benign at least in the short term perspective. Micro-separation results in less material damage with use of polyethylene than with the 2 other types of articulations. A few degrees of additional abduction above the geometrical limits for a particular socket is far less harmful if it is made of polyethylene compared to the situation in ceramic-on-ceramic or metal-on-metal bearings. Polyethylene is also more familiar to the majority of orthopaedic surgeons. In the operating room the cross-linked polyethylene is identical to those types of polyethylene, which have been used fore 3 to 4 decades. Finally the cost is a major factor (Harris 2004). The hard-on-hard bearings are substantially more expensive. The fracture incidence of ceramics components has decreased with improved manufacturing technology, but the risk of polyethylene fracture appears to be still smaller. On the other hand using highly cross-link polyethylene carries some risks. Particles generated from this new material are smaller with higher inflammatory response. Compared with joints including conventionally sterilised polyethylene the total particle production is, however, reduced with more than 85%, which has implications for the magnitude of the inflammatory response. The significance and importance of the irradiation and melting induced changes of the mechanical properties of the polyethylene is not known. Long term follow-up is needed to evaluate this issue. Charnley preferred small head sizes in total hip replacement because they resulted in transmittance of low frictional torque to the acetabular implant. Mueller advocated larger head sizes with improved joint stability and lower contact pressure. Large heads do, however, imply increased volumetric wear. Therefore, 32 mm heads were abandoned in the early 90ties in favour of 28 mm heads. Another consequence of using larger heads is that polyethylene liners are relatively thin. The highly cross-link polyethylene and the hard bearings can be used with bigger femoral heads, which increases the range of motion and the hip joint stability. Amorphous diamond coatings has been studied as an alternative bearing surface in the laboratory (Santavirta 2003). Such coatings may provide wear rates 104 to 105 times lower than conventional THR articulations, because of their extremely hard surface and low coefficient of friction without any corrosion paths (Santavirta et al. 1999 Lappalainen et al. 2003).